Radiation exposure with the NOMAD portable X-ray system.

A D Goren1, M Bonvento, J Biernacki, D C Colosi.   

Abstract

OBJECTIVES: A new hand-held battery-operated portable X-ray system was tested for possible leakage radiation through the existing heavy metal compounds surrounding the X-ray tube, backscatter radiation through the lead-filled acrylic shield attached at the end of the exit tube and patient exposure.
METHODS: Dose measurements were conducted using a DXTRR phantom and a water phantom. All measurements were recorded using calibrated thermoluminescent dosimetry (TLD), calibrated Unfors Model 583L dosemeter, and a calibrated Radcal MDH model 1015 dosemeter. The settings for all exposure were 60 kVp, 2.3 mA and 0.25 s using Kodak Insight (Class F) film.
RESULTS: All backscatter measurements, in front of the shield, behind the shield, at the finger of the operator, the operator's chest, eyes and gonads were significantly below the maximum permissible radiation leakage as per the United States Food and Drug Administration regulations (100 mR h(-1)). Our measurements indicate that the exposure would be well within the occupational maximum permissible dose for an occupationally exposed person. Film dose was consistent with the manufacturer's recommendations. As a result of our measurements, the State of New York Bureau of Environmental Radiation Protection granted us a variance to use the NOMAD on a case-by-case basis.
CONCLUSIONS: Our data have shown that the NOMAD presents risks that are no greater than with standard dental radiographic units to the patient or operator and the measured doses are well below recommended levels.

Entities:  

Mesh:

Year:  2008        PMID: 18239038     DOI: 10.1259/dmfr/33303181

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  10 in total

1.  Effect of the amount of battery charge on tube voltage in different hand-held dental x-ray systems.

Authors:  Eun-Kyung Kim
Journal:  Imaging Sci Dent       Date:  2012-03-22

2.  The effects of device position on the operator's radiation dose when using a handheld portable X-ray device.

Authors:  Jimmy Makdissi; Ravikiran R Pawar; Ben Johnson; Bun S Chong
Journal:  Dentomaxillofac Radiol       Date:  2016-01-14       Impact factor: 2.419

3.  Assessment of image quality and exposure parameters of an intraoral portable X-rays device.

Authors:  Elton G Zenóbio; Madelon Af Zenóbio; Carolina Db Azevedo; Maria do Socorro Nogueira; Cláudio D Almeida; Flávio R Manzi
Journal:  Dentomaxillofac Radiol       Date:  2018-12-03       Impact factor: 2.419

4.  Assessment of the radiological safety of a Genoray portable dental X-ray unit.

Authors:  S M Hosseini Pooya; L Hafezi; F Manafi; A R Talaeipour
Journal:  Dentomaxillofac Radiol       Date:  2014-10-24       Impact factor: 2.419

5.  Justification and good practice in using handheld portable dental X-ray equipment: a position paper prepared by the European Academy of DentoMaxilloFacial Radiology (EADMFR).

Authors:  W E R Berkhout; A Suomalainen; D Brüllmann; R Jacobs; K Horner; H C Stamatakis
Journal:  Dentomaxillofac Radiol       Date:  2015-02-24       Impact factor: 2.419

6.  Forensic odontology in the disaster victim identification process.

Authors:  P Pittayapat; R Jacobs; E De Valck; D Vandermeulen; G Willems
Journal:  J Forensic Odontostomatol       Date:  2012-07-01

7.  Evaluation of stray radiation to the operator for five hand-held dental X-ray devices.

Authors:  Richard Smith; Richard Tremblay; Graeme M Wardlaw
Journal:  Dentomaxillofac Radiol       Date:  2019-02-12       Impact factor: 2.419

8.  The way we were (and how we got here): fifty years of technology changes in dental and maxillofacial radiology.

Authors:  Roberto Molteni
Journal:  Dentomaxillofac Radiol       Date:  2020-06-11       Impact factor: 2.419

9.  Assessment of endodontically treated teeth by using different radiographic methods: an ex vivo comparison between CBCT and other radiographic techniques.

Authors:  Kemal Özgür Demiralp; Kıvanç Kamburoğlu; Kahraman Güngör; Selcen Yüksel; Gokcen Demiralp; Ozlem Uçok
Journal:  Imaging Sci Dent       Date:  2012-09-21

10.  The reduction methods of operator's radiation dose for portable dental X-ray machines.

Authors:  Jeong-Yeon Cho; Won-Jeong Han
Journal:  Restor Dent Endod       Date:  2012-08-29
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.