OBJECTIVES: This study was aimed to investigate the methods to reduce operator's radiation dose when taking intraoral radiographs with portable dental X-ray machines.Entities:
Keywords: Dental digital radiography; Portable dental X-ray; Radiation dosage; Radiation protection
Year: 2012 PMID: 23429415 PMCID: PMC3569401 DOI: 10.5395/rde.2012.37.3.160
Source DB: PubMed Journal: Restor Dent Endod ISSN: 2234-7658
Figure 1Human skull phantom DXTRR III.
Figure 2An 1,800 cc ionization chamber for low-level radiation measurement.
Figure 3Two kinds of portable hand-held X-ray machines (Left-DX3000; Right-Rextar).
Figure 4Operator's radiation dose was measured by an ionization chamber at the operator's hand level using the portable dental X-ray machine without a lead shield.
Figure 5Portable dental X-ray machine with a lead shield.
Figure 6Scattered radiation dose was measured by an ionization chamber at the operator's waist level.
Figure 7Radiation doses were measured at the hand level of operators using an ionization chamber covered with a lead sleeve.
Figure 8Operator's radiation dose at hand level of operator was measured using short and long cones.
Operator's radiation dose during the exposure of lower teeth with and without the backscatter shield
chest level, 130 cm height from floor; waist level, 70 cm height from floor.
*dose in µR means the leakage and scattered radiation dose and standard deviation measured by an 1,800 cc ionizing chamber.
Operator's radiation dose at the hand level during the exposure of lower teeth with and without the lead gloves
anterior teeth, lower anterior periapical radiography; posterior teeth, lower posterior periapical radiography.
*dose in µR means the average leakage and scattered radiation dose and standard deviation measured by an 1,800 cc ionizing chamber.
Operator's radiation dose at the hand level during the exposure of the lower teeth using 6 cm and 14 cm length cone
anterior teeth, lower anterior periapical radiography; posterior teeth, lower posterior periapical radiography.
*dose in µR means the average leakage and scattered radiation dose and standard deviation measured by 1,800 cc ionizing chamber.