A technical solution to avoid partial scan artifacts in cardiac MDCT.

A N Primak1, Y Dong, O P Dzyubak, S M Jorgensen, C H McCollough, E L Ritman.   

Abstract

Quantitative evaluation of cardiac image data obtained using multidetector row computed tomography (CT) is compromised by partial scan reconstructions, which improve the temporal resolution but significantly increase image-to-image CT number variations for a fixed region of interest compared to full reconstruction images. The feasibility of a new approach to solve this problem is assessed. An anthropomorphic cardiac phantom and an anesthetized pig were scanned on a dual-source CT scanner using both full and partial scan acquisition modes under different conditions. Additional scans were conducted with the electrocardiogram (ECG) signal being in synchrony with the gantry rotation. In the animal study, a simple x-ray detector was used to generate a signal once per gantry rotation. This signal was then used to pace the pig's heart. Phantom studies demonstrated that partial scan artifacts are strongly dependent on the rotational symmetry of angular projections, which is determined by the object shape and composition and its position with respect to the isocenter. The degree of partial scan artifacts also depends on the location of the region of interest with respect to highly attenuating materials (bones, iodine, etc.) within the object. Single-source partial scan images (165 ms temporal resolution) were significantly less affected by partial scan artifacts compared to dual-source partial scan images (82 ms temporal resolution). When the ECG signal was in synchrony with the gantry rotation, the same cardiac phase always corresponded to the same positions of the x-ray tube(s) and, hence, the same scattering and beam hardening geometry. As a result, the range of image-to-image CT number variations for partial scan reconstruction images acquired in synchronized mode was decreased to that achieved using full reconstruction image data. The success of the new approach, which synchronizes the ECG signal with the position of the x-ray tube(s), was demonstrated both in the phantom and animal experiments.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18196800      PMCID: PMC2577228          DOI: 10.1118/1.2805476

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  34 in total

1.  Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography.

Authors:  Jiang Hsieh; John Londt; Melissa Vass; Jay Li; Xiangyang Tang; Darin Okerlund
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

Review 2.  Quantification of global left ventricular function: comparison of multidetector computed tomography and magnetic resonance imaging. a meta-analysis and review of the current literature.

Authors:  P A van der Vleuten; T P Willems; M J W Götte; R A Tio; M J W Greuter; F Zijlstra; M Oudkerk
Journal:  Acta Radiol       Date:  2006-12       Impact factor: 1.990

Review 3.  Assessment of myocardial microvascular function: new opportunities in fast computed tomography.

Authors:  Elena Daghini; Lilach O Lerman
Journal:  Trends Cardiovasc Med       Date:  2007-01       Impact factor: 6.677

Review 4.  Cardiac CT: coronary arteries and beyond.

Authors:  Andreas H Mahnken; Georg Mühlenbruch; Rolf W Günther; Joachim E Wildberger
Journal:  Eur Radiol       Date:  2006-10-26       Impact factor: 5.315

Review 5.  CT of cardiac function.

Authors:  Giancarlo Savino; Peter Zwerner; Christopher Herzog; Marco Politi; Lorenzo Bonomo; Philip Costello; U Joseph Schoepf
Journal:  J Thorac Imaging       Date:  2007-02       Impact factor: 3.000

Review 6.  Chasing the heart: new developments for cardiac CT.

Authors:  Thomas G Flohr; U Joseph Schoepf; Bernd M Ohnesorge
Journal:  J Thorac Imaging       Date:  2007-02       Impact factor: 3.000

Review 7.  Comprehensive cardiac CT study: evaluation of coronary arteries, left ventricular function, and myocardial perfusion--is it possible?

Authors:  Ricardo C Cury; Koen Nieman; Michael D Shapiro; Khurram Nasir; Roberto C Cury; Thomas J Brady
Journal:  J Nucl Cardiol       Date:  2007-04       Impact factor: 5.952

8.  Optimal short scan convolution reconstruction for fanbeam CT.

Authors:  D L Parker
Journal:  Med Phys       Date:  1982 Mar-Apr       Impact factor: 4.071

Review 9.  CT of coronary artery disease.

Authors:  Christopher Herzog; Stefan Zangos; Peter Zwerner; Philip Costello; Thomas J Vogl; U Joseph Schoepf
Journal:  J Thorac Imaging       Date:  2007-02       Impact factor: 3.000

10.  Accuracy of 16-row multidetector computed tomography for the assessment of coronary artery stenosis.

Authors:  Mario J Garcia; Jonathan Lessick; Martin H K Hoffmann
Journal:  JAMA       Date:  2006-07-26       Impact factor: 56.272

View more
  6 in total

1.  A strategy to decrease partial scan reconstruction artifacts in myocardial perfusion CT: phantom and in vivo evaluation.

Authors:  Juan C Ramirez-Giraldo; Lifeng Yu; Birgit Kantor; Erik L Ritman; Cynthia H McCollough
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  A new redundancy weighting scheme for nonstationary data for computed tomography.

Authors:  Katsuyuki Taguchi; Jochen Cammin
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

3.  Reduction of image noise in low tube current dynamic CT myocardial perfusion imaging using HYPR processing: a time-attenuation curve analysis.

Authors:  Michael A Speidel; Courtney L Bateman; Yinghua Tao; Amish N Raval; Timothy A Hacker; Scott B Reeder; Michael S Van Lysel
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

4.  Noise spatial nonuniformity and the impact of statistical image reconstruction in CT myocardial perfusion imaging.

Authors:  Pascal Theriault Lauzier; Jie Tang; Michael A Speidel; Guang-Hong Chen
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

5.  Early changes in myocardial microcirculation in asymptomatic hypercholesterolemic subjects: as detected by perfusion CT.

Authors:  Thomas R Behrenbeck; Cynthia H McCollough; Wayne L Miller; Eric E Williamson; Shuai Leng; Timothy L Kline; Erik L Ritman
Journal:  Ann Biomed Eng       Date:  2014-03       Impact factor: 3.934

6.  Qualitative and quantitative assessment of adenosine triphosphate stress whole-heart dynamic myocardial perfusion imaging using 256-slice computed tomography.

Authors:  Akira Kurata; Naoto Kawaguchi; Teruhito Kido; Katsuji Inoue; Jun Suzuki; Akiyoshi Ogimoto; Jun-ichi Funada; Jitsuo Higaki; Masao Miyagawa; Mani Vembar; Teruhito Mochizuki
Journal:  PLoS One       Date:  2013-12-23       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.