Literature DB >> 22830741

Noise spatial nonuniformity and the impact of statistical image reconstruction in CT myocardial perfusion imaging.

Pascal Theriault Lauzier1, Jie Tang, Michael A Speidel, Guang-Hong Chen.   

Abstract

PURPOSE: To achieve high temporal resolution in CT myocardial perfusion imaging (MPI), images are often reconstructed using filtered backprojection (FBP) algorithms from data acquired within a short-scan angular range. However, the variation in the central angle from one time frame to the next in gated short scans has been shown to create detrimental partial scan artifacts when performing quantitative MPI measurements. This study has two main purposes. (1) To demonstrate the existence of a distinct detrimental effect in short-scan FBP, i.e., the introduction of a nonuniform spatial image noise distribution; this nonuniformity can lead to unexpectedly high image noise and streaking artifacts, which may affect CT MPI quantification. (2) To demonstrate that statistical image reconstruction (SIR) algorithms can be a potential solution to address the nonuniform spatial noise distribution problem and can also lead to radiation dose reduction in the context of CT MPI.
METHODS: Projection datasets from a numerically simulated perfusion phantom and an in vivo animal myocardial perfusion CT scan were used in this study. In the numerical phantom, multiple realizations of Poisson noise were added to projection data at each time frame to investigate the spatial distribution of noise. Images from all datasets were reconstructed using both FBP and SIR reconstruction algorithms. To quantify the spatial distribution of noise, the mean and standard deviation were measured in several regions of interest (ROIs) and analyzed across time frames. In the in vivo study, two low-dose scans at tube currents of 25 and 50 mA were reconstructed using FBP and SIR. Quantitative perfusion metrics, namely, the normalized upslope (NUS), myocardial blood volume (MBV), and first moment transit time (FMT), were measured for two ROIs and compared to reference values obtained from a high-dose scan performed at 500 mA.
RESULTS: Images reconstructed using FBP showed a highly nonuniform spatial distribution of noise. This spatial nonuniformity led to large fluctuations in the temporal direction. In the numerical phantom study, the level of noise was shown to vary by as much as 87% within a given image, and as much as 110% between different time frames for a ROI far from isocenter. The spatially nonuniform noise pattern was shown to correlate with the source trajectory and the object structure. In contrast, images reconstructed using SIR showed a highly uniform spatial distribution of noise, leading to smaller unexpected noise fluctuations in the temporal direction when a short scan angular range was used. In the numerical phantom study, the noise varied by less than 37% within a given image, and by less than 20% between different time frames. Also, the noise standard deviation in SIR images was on average half of that of FBP images. In the in vivo studies, the deviation observed between quantitative perfusion metrics measured from low-dose scans and high-dose scans was mitigated when SIR was used instead of FBP to reconstruct images.
CONCLUSIONS: (1) Images reconstructed using FBP suffered from nonuniform spatial noise levels. This nonuniformity is another manifestation of the detrimental effects caused by short-scan reconstruction in CT MPI. (2) Images reconstructed using SIR had a much lower and more uniform noise level and thus can be used as a potential solution to address the FBP nonuniformity. (3) Given the improvement in the accuracy of the perfusion metrics when using SIR, it may be desirable to use a statistical reconstruction framework to perform low-dose dynamic CT MPI.

Mesh:

Year:  2012        PMID: 22830741      PMCID: PMC3382258          DOI: 10.1118/1.4722983

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  48 in total

1.  Image reconstruction with shift-variant filtration and its implication for noise and resolution properties in fan-beam computed tomography.

Authors:  Xiaochuan Pan; Lifeng Yu
Journal:  Med Phys       Date:  2003-04       Impact factor: 4.071

2.  Assessment of myocardial perfusion and viability from routine contrast-enhanced 16-detector-row computed tomography of the heart: preliminary results.

Authors:  Konstantin Nikolaou; Javier Sanz; Michael Poon; Bernd J Wintersperger; Bernd Ohnesorge; Teresa Rius; Zahi A Fayad; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2005-03-18       Impact factor: 5.315

3.  A technical solution to avoid partial scan artifacts in cardiac MDCT.

Authors:  A N Primak; Y Dong; O P Dzyubak; S M Jorgensen; C H McCollough; E L Ritman
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

4.  Partial scan artifact reduction (PSAR) for the assessment of cardiac perfusion in dynamic phase-correlated CT.

Authors:  Philip Stenner; Bernhard Schmidt; Herbert Bruder; Thomas Allmendinger; Ulrike Haberland; Thomas Flohr; Marc Kachelriess
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

5.  Multidetector computed tomography myocardial perfusion imaging during adenosine stress.

Authors:  Richard T George; Caterina Silva; Marco A S Cordeiro; Anthony DiPaula; Douglas R Thompson; William F McCarthy; Takashi Ichihara; Joao A C Lima; Albert C Lardo
Journal:  J Am Coll Cardiol       Date:  2006-06-21       Impact factor: 24.094

6.  Radiation dose reduction in time-resolved CT angiography using highly constrained back projection reconstruction.

Authors:  Mark Supanich; Yinghua Tao; Brian Nett; Kari Pulfer; Jiang Hsieh; Patrick Turski; Charles Mistretta; Howard Rowley; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2009-06-30       Impact factor: 3.609

7.  Dynamic perfusion CT: optimizing the temporal resolution for the calculation of perfusion CT parameters in stroke patients.

Authors:  Andreas Kämena; Florian Streitparth; Christian Grieser; Lukas Lehmkuhl; Basil Jamil; Katarzyna Wojtal; Jens Ricke; Maciej Pech
Journal:  Eur J Radiol       Date:  2007-03-23       Impact factor: 3.528

8.  Detection of hemodynamically significant coronary artery stenosis: incremental diagnostic value of dynamic CT-based myocardial perfusion imaging.

Authors:  Fabian Bamberg; Alexander Becker; Florian Schwarz; Roy P Marcus; Martin Greif; Franz von Ziegler; Ron Blankstein; Udo Hoffmann; Wieland H Sommer; Verena S Hoffmann; Thorsten R C Johnson; Hans-Christoph R Becker; Bernd J Wintersperger; Maximilian F Reiser; Konstantin Nikolaou
Journal:  Radiology       Date:  2011-09       Impact factor: 11.105

9.  Incremental value of adenosine-induced stress myocardial perfusion imaging with dual-source CT at cardiac CT angiography.

Authors:  Jose A Rocha-Filho; Ron Blankstein; Leonid D Shturman; Hiram G Bezerra; David R Okada; Ian S Rogers; Brian Ghoshhajra; Udo Hoffmann; Gudrun Feuchtner; Wilfred S Mamuya; Thomas J Brady; Ricardo C Cury
Journal:  Radiology       Date:  2010-02       Impact factor: 11.105

10.  Optimal medical therapy with or without percutaneous coronary intervention to reduce ischemic burden: results from the Clinical Outcomes Utilizing Revascularization and Aggressive Drug Evaluation (COURAGE) trial nuclear substudy.

Authors:  Leslee J Shaw; Daniel S Berman; David J Maron; G B John Mancini; Sean W Hayes; Pamela M Hartigan; William S Weintraub; Robert A O'Rourke; Marcin Dada; John A Spertus; Bernard R Chaitman; John Friedman; Piotr Slomka; Gary V Heller; Guido Germano; Gilbert Gosselin; Peter Berger; William J Kostuk; Ronald G Schwartz; Merill Knudtson; Emir Veledar; Eric R Bates; Benjamin McCallister; Koon K Teo; William E Boden
Journal:  Circulation       Date:  2008-02-11       Impact factor: 29.690

View more
  6 in total

1.  Characterization of statistical prior image constrained compressed sensing. I. Applications to time-resolved contrast-enhanced CT.

Authors:  Pascal Theriault Lauzier; Guang-Hong Chen
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

Review 2.  Cardiac CT for myocardial ischaemia detection and characterization--comparative analysis.

Authors:  A M Bucher; C N De Cecco; U J Schoepf; R Wang; F G Meinel; S R Binukrishnan; J V Spearman; T J Vogl; B Ruzsics
Journal:  Br J Radiol       Date:  2014-08-19       Impact factor: 3.039

3.  Low dose dynamic CT myocardial perfusion imaging using a statistical iterative reconstruction method.

Authors:  Yinghua Tao; Guang-Hong Chen; Timothy A Hacker; Amish N Raval; Michael S Van Lysel; Michael A Speidel
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

4.  Statistical model based iterative reconstruction (MBIR) in clinical CT systems: experimental assessment of noise performance.

Authors:  Ke Li; Jie Tang; Guang-Hong Chen
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

5.  Robust dynamic myocardial perfusion CT deconvolution for accurate residue function estimation via adaptive-weighted tensor total variation regularization: a preclinical study.

Authors:  Dong Zeng; Changfei Gong; Zhaoying Bian; Jing Huang; Xinyu Zhang; Hua Zhang; Lijun Lu; Shanzhou Niu; Zhang Zhang; Zhengrong Liang; Qianjin Feng; Wufan Chen; Jianhua Ma
Journal:  Phys Med Biol       Date:  2016-10-26       Impact factor: 3.609

Review 6.  Myocardial perfusion imaging with cardiac computed tomography: state of the art.

Authors:  Amit R Patel; Nicole M Bhave; Victor Mor-Avi
Journal:  J Cardiovasc Transl Res       Date:  2013-08-21       Impact factor: 4.132

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.