| Literature DB >> 18196396 |
George Anastopoulos1, Panagiotis G Ntagiopoulos, Dionisios Chissas, Athanasios Papaeliou, Antonios Asimakopoulos.
Abstract
UNLABELLED: The indications for intramedullary nailing have expanded to include most tibial shaft fractures. Nail design has improved since their first introduction, but distal locking remains a difficult part of the procedure, resulting in radiation exposure to the patient and the surgeon and increased operation time. To address these issues, we describe an alternative surgical technique using a newly designed distal targeting device that consists of a proximally mounted aiming arm, and we report the preliminary data from its use in all tibial shaft fractures amenable to surgery for a 2-year period. Sixty-three tibial shaft fractures were treated with this method. The mean duration of the distal locking was 6.5 minutes, and in all successful cases, radiation exposure for distal locking was two shots (one shot before targeting and another for the confirmation of proper screw insertion). Radiation exposure was on average 0.85 seconds (range, 0.4-1.2 seconds) and 1.4 mGy (range, 0.8-1.9 mGy). There were no major intraoperative complications related to the technique. The method has certain advantages and can reduce radiation exposure and operation time. Nonetheless, familiarity with the instrumentation is a prerequisite for accurate distal locking. LEVEL OF EVIDENCE: Level IV Therapeutic study.Entities:
Mesh:
Year: 2008 PMID: 18196396 PMCID: PMC2505307 DOI: 10.1007/s11999-007-0036-z
Source DB: PubMed Journal: Clin Orthop Relat Res ISSN: 0009-921X Impact factor: 4.176