Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.

John R Feiner1, Mark D Rollins, Jeffrey W Sall, Helge Eilers, Paul Au, Philip E Bickler.   

Abstract

BACKGROUND: Carbon monoxide poisoning is a significant problem in most countries, and a reliable method of quick diagnosis would greatly improve patient care. Until the recent introduction of a multiwavelength "pulse CO-oximeter" (Masimo Rainbow SET(®) Radical-7), obtaining carboxyhemoglobin (COHb) levels in blood required blood sampling and laboratory analysis. In this study, we sought to determine whether hypoxemia, which can accompany carbon monoxide poisoning, interferes with the accurate detection of COHb.
METHODS: Twelve healthy, nonsmoking, adult volunteers were fitted with 2 standard pulse-oximeter finger probes and 2 Rainbow probes for COHb detection. A radial arterial catheter was placed for blood sampling during 3 interventions: (1) increasing hypoxemia in incremental steps with arterial oxygen saturations (SaO2) of 100% to 80%; (2) normoxia with incremental increases in %COHb to 12%; and (3) elevated COHb combined with hypoxemia with SaO2 of 100% to 80%. Pulse-oximeter (SpCO) readings were compared with simultaneous arterial blood values at the various increments of hypoxemia and carboxyhemoglobinemia (≈25 samples per subject). Pulse CO-oximeter performance was analyzed by calculating the mean bias (SpCO - %COHb), standard deviation of the bias (precision), and the root-mean-square error (A(rms)).
RESULTS: The Radical-7 accurately detected hypoxemia with both normal and elevated levels of COHb (bias mean ± SD: 0.44% ± 1.69% at %COHb <4%, and -0.29% ± 1.64% at %COHb ≥4%, P < 0.0001, and A(rms) 1.74% vs 1.67%). COHb was accurately detected during normoxia and moderate hypoxia (bias mean ± SD: -0.98 ± 2.6 at SaO2 ≥95%, and -0.7 ± 4.0 at SaO2 <95%, P = 0.60, and A(rms) 2.8% vs 4.0%), but when SaO2 decreased below approximately 85%, the pulse CO-oximeter always gave low signal quality errors and did not report SpCO values.
CONCLUSIONS: In healthy volunteers, the Radical-7 pulse CO-oximeter accurately detects hypoxemia with both low and elevated COHb levels, and accurately detects COHb, but only reads SpCO when SaO2 is more than approximately 85%.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23477959      PMCID: PMC4476500          DOI: 10.1213/ANE.0b013e31828610a0

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  31 in total

1.  Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study.

Authors:  Steven J Barker; Jeremy Curry; Daniel Redford; Scott Morgan
Journal:  Anesthesiology       Date:  2006-11       Impact factor: 7.892

2.  Agreement between methods of measurement with multiple observations per individual.

Authors:  J Martin Bland; Douglas G Altman
Journal:  J Biopharm Stat       Date:  2007       Impact factor: 1.051

3.  Finding needles in a haystack: a case series of carbon monoxide poisoning detected using new technology in the emergency department.

Authors:  Kerlen J Chee; Douglas Nilson; Robert Partridge; Ashley Hughes; Selim Suner; Andrew Sucov; Gregory Jay
Journal:  Clin Toxicol (Phila)       Date:  2008-06       Impact factor: 4.467

4.  Use of pulse co-oximetry as a screening and monitoring tool in mass carbon monoxide poisoning.

Authors:  Bryan E Bledsoe; Kevin Nowicki; James H Creel; Dale Carrison; Harry W Severance
Journal:  Prehosp Emerg Care       Date:  2010 Jan-Mar       Impact factor: 3.077

5.  Fire and ice: diagnosis of carbon monoxide poisoning in a remote environment.

Authors:  D M Crawford; N B Hampson
Journal:  Emerg Med J       Date:  2008-04       Impact factor: 2.740

6.  A new tool for the early diagnosis of carbon monoxide intoxication.

Authors:  A Piatkowski; D Ulrich; G Grieb; N Pallua
Journal:  Inhal Toxicol       Date:  2009-11       Impact factor: 2.724

7.  Reliability of new pulse CO-oximeter in victims of carbon monoxide poisoning.

Authors:  M Coulange; A Barthelemy; F Hug; A L Thierry; L De Haro
Journal:  Undersea Hyperb Med       Date:  2008 Mar-Apr       Impact factor: 0.698

8.  Non-invasive pulse CO-oximetry screening in the emergency department identifies occult carbon monoxide toxicity.

Authors:  Selim Suner; Robert Partridge; Andrew Sucov; Jonathan Valente; Kerlen Chee; Ashley Hughes; Gregory Jay
Journal:  J Emerg Med       Date:  2008-01-28       Impact factor: 1.484

Review 9.  Misconceptions in reporting oxygen saturation.

Authors:  John Toffaletti; Willem G Zijlstra
Journal:  Anesth Analg       Date:  2007-12       Impact factor: 5.108

10.  Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender.

Authors:  John R Feiner; John W Severinghaus; Philip E Bickler
Journal:  Anesth Analg       Date:  2007-12       Impact factor: 5.108

View more
  9 in total

Review 1.  Carbon Monoxide and Cyanide Poisoning in the Burned Pregnant Patient: An Indication for Hyperbaric Oxygen Therapy.

Authors:  Derek M Culnan; Beretta Craft-Coffman; Genevieve H Bitz; Karel D Capek; Yiji Tu; William C Lineaweaver; Maggie J Kuhlmann-Capek
Journal:  Ann Plast Surg       Date:  2018-03       Impact factor: 1.539

Review 2.  Anesthesia-Related Carbon Monoxide Exposure: Toxicity and Potential Therapy.

Authors:  Richard J Levy
Journal:  Anesth Analg       Date:  2016-09       Impact factor: 5.108

3.  Measurement of blood-oxygen saturation using a photoacoustic technique in the rabbit hypoxemia model.

Authors:  Kiguna Sei; Masanori Fujita; Takeshi Hirasawa; Shinpei Okawa; Toshihiro Kushibiki; Hidenori Sasa; Kenichi Furuya; Miya Ishihara
Journal:  J Clin Monit Comput       Date:  2018-06-06       Impact factor: 2.502

Review 4.  Pulse oximetry.

Authors:  Amal Jubran
Journal:  Crit Care       Date:  2015-07-16       Impact factor: 9.097

5.  Accuracy of detection of carboxyhemoglobin and methemoglobin in human and bovine blood with an inexpensive, pocket-size infrared scanner.

Authors:  Margot P Bickler; Laura J Rhodes
Journal:  PLoS One       Date:  2018-03-07       Impact factor: 3.240

6.  A multicentre prospective observational study comparing arterial blood gas values to those obtained by pulse oximeters used in adult patients attending Australian and New Zealand hospitals.

Authors:  Janine Pilcher; Laura Ploen; Steve McKinstry; George Bardsley; Jimmy Chien; Lesley Howard; Sharon Lee; Lutz Beckert; Maureen Swanney; Mark Weatherall; Richard Beasley
Journal:  BMC Pulm Med       Date:  2020-01-09       Impact factor: 3.317

Review 7.  The Use of Pulse Oximetry in the Assessment of Acclimatization to High Altitude.

Authors:  Tobias Dünnwald; Roland Kienast; David Niederseer; Martin Burtscher
Journal:  Sensors (Basel)       Date:  2021-02-10       Impact factor: 3.576

Review 8.  Pulse oximetry: fundamentals and technology update.

Authors:  Meir Nitzan; Ayal Romem; Robert Koppel
Journal:  Med Devices (Auckl)       Date:  2014-07-08

9.  Methaemoglobin and COHb in patients with malaria.

Authors:  Thomas Hänscheid; Tom Gresnigt; Sascha Löhr; Arnaud Flamen; Thomas Zoller; José Melo-Cristino; Martin P Grobusch
Journal:  Malar J       Date:  2014-07-23       Impact factor: 2.979

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.