The accuracy of a near-infrared spectroscopy cerebral oximetry device and its potential value for estimating jugular venous oxygen saturation.

Keita Ikeda1, David B MacLeod, Hilary P Grocott, Eugene W Moretti, Warwick Ames, Charles Vacchiano.   

Abstract

BACKGROUND: An intriguing potential clinical use of cerebral oximeter measurements (SctO2) is the ability to noninvasively estimate jugular bulb venous oxygen saturation (SjvO2). Our purpose in this study was to determine the accuracy of the FORE-SIGHT(®) (CAS Medical Systems, Branford, CT), which is calibrated to a weighted average of 70% (SjvO2) and 30% arterial saturation, for Food and Drug Administration pre-market approval 510(k) certification by adapting an industry standard protocol, ISO 9919:2005 (www.ISO.org) (used for pulse oximeters), and to evaluate the use of SctO2 and SpO2 measurements to noninvasively estimate jugular venous oxygen saturation (SnvO2).
METHODS: Paired blood gas samples from the radial artery and the jugular venous bulb were collected from 20 healthy volunteers undergoing progressive oxygen desaturation from 100% to 70%. The blood sample pairs were analyzed via co-oximetry and used to calculate the approximate mixed vascular cerebral blood oxygen saturation, or reference SctO2 values (refSctO2), during increasing hypoxia. These reference values were compared to bilateral FORE-SIGHT SctO2 values recorded simultaneously with the blood gas draws to determine its accuracy. Bilateral SctO2 and SpO2 measurements were then used to calculate SnvO2 values which were compared to SjvO2.
RESULTS: Two hundred forty-six arterial and 253 venous samples from 18 subjects were used in the analysis. The ipsilateral FORE-SIGHT SctO2 values showed a tolerance interval (TI) of [-10.72 to 10.90] and Lin concordance correlation coefficient (CCC) with standard error (SE) of 0.83 ± 0.073 with the refSctO2 values calculated using arterial and venous blood gases. The ipsilateral data had a CCC of 0.81 + 0.059 with TI of [-9.22 to 9.40] with overall bias of 0.09%, and amplitude of the root mean square of error after it was corrected with random effects analysis was 2.92%. The bias and variability values between the ipsilateral and the contralateral FORE-SIGHT SctO2 measurements varied from person to person. The SnvO2 calculated from the ipsilateral SctO2 and SpO2 data showed a CCC ± SE of 0.79 ± 0.088, TI = [-14.93 to 15.33], slope of 0.98, y-intercept of 1.14% with SjvO2 values with a bias of 0.20% and an Arms of 4.08%. The SnvO2 values calculated independently from contralateral forehead FORE-SIGHT SctO2 values were not as correlated with the SjvO2 values (contralateral side CCC + SE = 0.72 ± 0.118, TI = [-14.86 to 15.20], slope of 0.66, and y-intercept of 20.36%).
CONCLUSIONS: The FORE-SIGHT cerebral oximeter was able to estimate oxygen saturation within the tissues of the frontal lobe under conditions of normocapnia and varying degrees of hypoxia (with 95% confidence interval of [-5.60 to 5.78] with ipsilateral blood sample data). These findings from healthy volunteers also suggest that the use of the calculated SnvO2 derived from SctO2 and SpO2 values may be a reasonable noninvasive method of estimating SjvO2 and therefore global cerebral oxygen consumption in the clinical setting. Further laboratory and clinical research is required to define the clinical utility of near-infrared spectroscopy determination of SctO2 and SnvO2 in the operating room setting.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25313967      PMCID: PMC4237713          DOI: 10.1213/ANE.0000000000000463

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  43 in total

1.  Optical tomography in the presence of void regions

Authors: 
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2000-09       Impact factor: 2.129

2.  Precise control of end-tidal carbon dioxide levels using sequential rebreathing circuits.

Authors:  R B Somogyi; A E Vesely; D Preiss; E Prisman; G Volgyesi; T Azami; S Iscoe; J A Fisher; H Sasano
Journal:  Anaesth Intensive Care       Date:  2005-12       Impact factor: 1.669

3.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

4.  Characterization of the near infrared absorption spectra of cytochrome aa3 and haemoglobin for the non-invasive monitoring of cerebral oxygenation.

Authors:  S Wray; M Cope; D T Delpy; J S Wyatt; E O Reynolds
Journal:  Biochim Biophys Acta       Date:  1988-03-30

5.  Effects of skin pigmentation on pulse oximeter accuracy at low saturation.

Authors:  Philip E Bickler; John R Feiner; John W Severinghaus
Journal:  Anesthesiology       Date:  2005-04       Impact factor: 7.892

6.  Differences in the control of breathing between Andean highlanders and lowlanders after 10 days acclimatization at 3850 m.

Authors:  Marat Slessarev; Alexandra Mardimae; David Preiss; Alex Vesely; Dahlia Y Balaban; Richard Greene; James Duffin; Joseph A Fisher
Journal:  J Physiol       Date:  2010-03-15       Impact factor: 5.182

7.  Accuracy of a cerebral oximeter in healthy volunteers under conditions of isocapnic hypoxia.

Authors:  L C Henson; C Calalang; J A Temp; D S Ward
Journal:  Anesthesiology       Date:  1998-01       Impact factor: 7.892

8.  Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans.

Authors:  P W McCormick; M Stewart; M G Goetting; G Balakrishnan
Journal:  Stroke       Date:  1991-05       Impact factor: 7.914

9.  Development and validation of a cerebral oximeter capable of absolute accuracy.

Authors:  David B MacLeod; Keita Ikeda; Charles Vacchiano; Aaron Lobbestael; Joyce A Wahr; Andrew D Shaw
Journal:  J Cardiothorac Vasc Anesth       Date:  2012-08-09       Impact factor: 2.628

10.  The influence of arterial oxygenation on cerebral venous oxygen saturation during hyperventilation.

Authors:  B F Matta; A M Lam; T S Mayberg
Journal:  Can J Anaesth       Date:  1994-11       Impact factor: 5.063

View more
  17 in total

Review 1.  In vivo validation of cerebral near-infrared spectroscopy: a review.

Authors:  Amalie la Cour; Gorm Greisen; Simon Hyttel-Sorensen
Journal:  Neurophotonics       Date:  2018-11-27       Impact factor: 3.593

2.  Influence of continuously evolving transcatheter aortic valve implantation technology on cerebral oxygenation.

Authors:  Ward Eertmans; Cornelia Genbrugge; Tom Fret; Maud Beran; Kim Engelen; Herbert Gutermann; Margot Vander Laenen; Willem Boer; Bert Ferdinande; Frank Jans; Jo Dens; Cathy De Deyne
Journal:  J Clin Monit Comput       Date:  2016-12-26       Impact factor: 2.502

Review 3.  Neuroprotective Strategies in Repair and Replacement of the Aortic Arch.

Authors:  Frank Manetta; Clancy W Mullan; Michael A Catalano
Journal:  Int J Angiol       Date:  2018-05-27

4.  Can the interchangeability of somatic oximeters in cardiac surgery inform cerebral applications?

Authors:  Hessam H Kashani; Hilary P Grocott
Journal:  J Clin Monit Comput       Date:  2018-07-04       Impact factor: 2.502

5.  An observational study of the optimal placement of a cerebral oximeter probe to avoid the frontal sinus in children.

Authors:  Eun-Hee Kim; Kyu-Young Song; In-Kyung Song; Ji-Hyun Lee; Young-Eun Jang; Hee-Soo Kim; Seo-Hee Lee; Jin-Tae Kim
Journal:  J Clin Monit Comput       Date:  2017-12-11       Impact factor: 2.502

6.  Dysfunctional cerebral autoregulation is associated with delirium in critically ill adults.

Authors:  Kevin Fh Lee; Michael D Wood; David M Maslove; John G Muscedere; J Gordon Boyd
Journal:  J Cereb Blood Flow Metab       Date:  2018-10-08       Impact factor: 6.200

7.  Hydroyxurea improves cerebral oxygen saturation in children with sickle cell anemia.

Authors:  Kristine Karkoska; Charles T Quinn; Omar Niss; Amanda Pfeiffer; Min Dong; Alexander A Vinks; Patrick T McGann
Journal:  Am J Hematol       Date:  2021-02-19       Impact factor: 10.047

8.  Accuracy of pulse oximeters in detecting hypoxemia in patients with chronic thromboembolic pulmonary hypertension.

Authors:  Tomoki Kohyama; Kiyoshi Moriyama; Riichiro Kanai; Mariko Kotani; Kohji Uzawa; Toru Satoh; Tomoko Yorozu
Journal:  PLoS One       Date:  2015-05-15       Impact factor: 3.240

9.  Non-invasive assessment of cerebral oxygenation: A comparison of retinal and transcranial oximetry.

Authors:  Karel Van Keer; Jan Van Keer; João Barbosa Breda; Vahid Nassiri; Cathy De Deyne; Cornelia Genbrugge; Luís Abegão Pinto; Ingeborg Stalmans; Evelien Vandewalle
Journal:  PLoS One       Date:  2018-01-05       Impact factor: 3.240

10.  Effects of Hypothermic Cardiopulmonary Bypass on Internal Jugular Bulb Venous Oxygen Saturation, Cerebral Oxygen Saturation, and Bispectral Index in Pediatric Patients Undergoing Cardiac Surgery: A Prospective Study.

Authors:  Zhiyong Hu; Lili Xu; Zhirui Zhu; Robert Seal; Patrick M McQuillan
Journal:  Medicine (Baltimore)       Date:  2016-01       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.