Literature DB >> 18022336

An analysis of the air-bone gap closure obtained by a crimping and a non-crimping titanium stapes prosthesis in otosclerosis.

Rinze A Tange1, Wilko Grolman.   

Abstract

OBJECTIVE: To evaluate the hearing results of stapes surgery with two different titanium stapes prostheses. A crimping and a non-crimping prosthesis.
METHODS: Retrospective analyses were carried out on the pre- and postoperative hearing results obtained after 126 primary stapedotomies with two different prostheses in patients with otosclerosis. Sixty-three patients had a stapedotomy with a K-piston Titanium (No. 1006107) type and 63 patients received a CliP-piston àWengen Titanium (No. 1006807) type after stapedotomy. The results were compared according to mean audiometrical parameters. A Chi-square Test analysis (a contingency table) was performed with odds ratio to analyze this retrospective case-control study.
RESULTS: The hearing results of both titanium stapes prostheses were comparable. There was no difference in the final hearing results between the two pistons studied. No statistically significant difference between the postoperative air-bone gap closures by the two titanium stapes prostheses studied could be demonstrated.
CONCLUSIONS: Because of the different properties of the two titanium piston studied (CliP-piston àWengen Titanium type, no crimping and K-piston Titanium crimping necessary) a difference in the closure of the air-bone gap could be suspected. This comparative study could not demonstrate a statistical better air-bone gap closure for one of the titanium pistons. Both stapes prostheses showed good results in reconstruction of the hearing loss due to otosclerosis.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18022336     DOI: 10.1016/j.anl.2007.04.007

Source DB:  PubMed          Journal:  Auris Nasus Larynx        ISSN: 0385-8146            Impact factor:   1.863


  9 in total

1.  Comparison between Fluoroplastic and Platinum/Titanium Piston in Stapedotomy: A Prospective, Randomized Clinical Study.

Authors:  Mohammad Faramarzi; Sareh Roosta; Niloofar Daneshian
Journal:  J Int Adv Otol       Date:  2020-08       Impact factor: 1.017

2.  Treatment of otosclerosis with a superelastic nitinol piston: first results.

Authors:  F Bast; S Weikert; T Schrom
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2011-02-09

3.  [Microsurgical middle ear operations : Variability of the audiometric results from frequency variations].

Authors:  L von Gierke; I Baumann; P K Plinkert; M Praetorius
Journal:  HNO       Date:  2011-07       Impact factor: 1.284

4.  [Stapes surgery : first experiences with the new Soft-CliP piston].

Authors:  C Brase; J Zenk; J Wurm; B Schick; H Iro; J Hornung
Journal:  HNO       Date:  2009-05       Impact factor: 1.284

5.  Our experience with the titanium soft clip piston stapedotomy.

Authors:  P P Singh; Arun Goyal
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2012-01-06

6.  Teflon Implants Versus Titanium Implants in Stapes Surgery.

Authors:  Mohan Bansal
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2015-12-14

7.  Retrospective analysis of the results of implanting Nitinol pistons with heat-crimping piston loops in stapes surgery.

Authors:  Joachim Albert Hornung; Christoph Brase; Alessandro Bozzato; Johannes Zenk; Bernhard Schick; Heinrich Iro
Journal:  Eur Arch Otorhinolaryngol       Date:  2010-01       Impact factor: 2.503

8.  Comparison of Fluoroplastic Causse Loop Piston and Titanium Soft-Clip in Stapedotomy.

Authors:  Mohammad Faramarzi; Nafiseh Gilanifar; Sareh Roosta
Journal:  Iran J Otorhinolaryngol       Date:  2017-01

9.  Evaluating hearing results in stapedotomy: Comparison of fluoroplastic Causse loop piston and Matrix prosthesis.

Authors:  Mohammad Faramarzi; Pegah Pedramfard; Sareh Roosta; Tayebeh Kazemi; Hanie Aboughadare
Journal:  Laryngoscope Investig Otolaryngol       Date:  2021-10-04
  9 in total