| Literature DB >> 34938864 |
Mohammad Faramarzi1,2, Pegah Pedramfard3, Sareh Roosta1, Tayebeh Kazemi1,2, Hanie Aboughadare3.
Abstract
OBJECTIVE: We carried out this research to assess and compare post-stapedotomy hearing results of Matrix titanium prosthesis with a Teflon piston prosthesis, specifically the fluoroplastic (Teflon) Causse loop piston prosthesis in patients who suffered from otosclerosis.Entities:
Keywords: fluoroplastic Causse loop piston; matrix prosthesis; ossicular prosthesis; otosclerosis; stapes surgery
Year: 2021 PMID: 34938864 PMCID: PMC8665457 DOI: 10.1002/lio2.675
Source DB: PubMed Journal: Laryngoscope Investig Otolaryngol ISSN: 2378-8038
FIGURE 1Retrospective cohort flow diagram
Preoperative and postoperative mean BC, AC, and ABG (dB) at the frequencies of 0.5‐4 kHz
| Group | Preoperative | Postoperative | Gain |
| |
|---|---|---|---|---|---|
| Mean | Causse loop piston | 12.7 ± 5.5 | 9.3 ± 5.8 | 3.4 ± 6.3 | <0.001 |
| Matrix | 12.3 ± 5.5 | 9.5 ± 5.0 | 2.9 ± 5.4 | <0.001 | |
|
| 0.735 | 0.902 | 0.667 | ||
| Mean | Causse loop piston | 48.1 ± 8.3 | 23.9 ± 8.9 | 24.3 ± 8.9 | <0.001 |
| Matrix | 47.6 ± 9.4 | 23.7 ± 8.1 | 23.9 ± 9.8 | <0.001 | |
|
| 0.775 | 0.913 | 0.864 | ||
| Mean | Causse loop piston | 35.4 ± 7.6 | 14.5 ± 5.7 | 20.9 ± 7.6 | <0.001 |
| Matrix | 35.3 ± 8.5 | 14.2 ± 4.7 | 21.1 ± 8.7 | <0.001 | |
|
| 0.934 | 0.761 | 0.910 |
Within group comparison (preoperative and postoperative).
Frequencies of 0.5, 1, 2, and 4 kHz were encompassed.
Mean ± SD, ABG, air bone gap; AC, air conduction; BC, bone conduction.
Between group comparison.
Postoperative AC, BC, and ABG gain (dB) at each frequency
| Group | 250 (Hz) | 500 (Hz) | 1000 (Hz) | 2000 (Hz) | 4000 (Hz) | |
|---|---|---|---|---|---|---|
| BC gain (dB) | Causse loop piston | 1.7 ± 9.3 | 1.9 ± 7.4 | 4.9 ± 6.9 | 7.7 ± 9.0 | −0.9 ± 7.6 |
| Matrix | −0.2 ± 6.3 | 0.4 ± 6.4 | 4.0 ± 6.2 | 6.0 ± 7.8 | 1.0 ± 7.8 | |
|
| 0.444 | 0.327 | 0.613 | 0.437 | 0.260 | |
| AC gain (dB) | Causse loop piston | 25.3 ± 12.6 | 28.9 ± 12.9 | 29.7 ± 11.0 | 22.9 ± 10.3 | 15.7 ± 12.6 |
| Matrix | 29.3 ± 10.7 | 30.2 ± 10.7 | 28.6 ± 11.2 | 20.0 ± 10.8 | 16.9 ± 14.1 | |
|
| 0.088 | 0.482 | 0.909 | 0.249 | 0.649 | |
| ABG closure (dB) | Causse loop piston | 23.6 ± 13.2 | 26.9 ± 11.2 | 24.8 ± 9.7 | 15.2 ± 10.0 | 16.6 ± 10.8 |
| Matrix |
29.5 ± 12.1 | 29.8 ± 10.7 | 24.6 ± 10.9 | 14.0 ± 9.9 | 15.9 ± 12.2 | |
|
| 0.044 | 0.256 | 0.633 | 0.735 | 0.598 |
Mean ± SD, ABG, air‐bone gap; AC, air conduction; BC, bone conduction.
Between group comparison.
Mean ABG closure distribution at the frequencies of 0.5–4 kHz
| ABG closure | Causse loop piston group | Matrix group |
|---|---|---|
| ( | ( | |
| ≤10 | 13(29.5) | 11(25.0) |
| 11‐20 | 25(56.8) | 30(68.2) |
| 21‐30 | 6(13.6) | 3(6.8) |
| >30 | 0(0.0) | 0(0.0) |
N(%), ABG, air‐bone gap; frequencies of 0.5, 1, 2, and 4 kHz were encompassed in mean ABG.
Incidence of post‐operative sensorineural hearing loss at different frequencies
| Group | 250 (Hz) | 500 (Hz) | 1000 (Hz) | 2000 (Hz) | 4000 (Hz) | 500‐4000 |
|---|---|---|---|---|---|---|
| Causse loop piston | 2(4.5) | 2(4.5) | 3(6.8) | 1(2.3) | 3(6.8) | 1(2.3) |
| Matrix | 1(2.3) | 1(2.3) | 0(0.0) | 2(4.5) | 2(4.5) | 1(2.3) |
|
| 1.000 | 1.000 | 0.241 | 1.000 | 1.000 | 1.000 |
Frequencies of 0.5, 1, 2, and 4 kHz were encompassed.
N(%).
Incidence of post‐operative sensorineural hearing loss at a frequency of 4 kHz
| Group | No. | 10‐20 (dB) | 21‐30 (dB) | >30 (dB) | Total (dB) |
|---|---|---|---|---|---|
| Causse loop piston | 44 | 3(6.8) | 0(0.0) | 0(0.0) | 3(6.8) |
| Matrix | 44 | 2(4.5) | 0(0.0) | 0(0.0) | 2(4.5) |
|
|
1.000 | ‐ | ‐ | 1.000 |
N(%).
Literature review of postoperative hearing outcomes in stapes surgery regarding the type of prosthesis
| Author (year) | Types of prosthesis (No) | Duration of follow‐up | ABG ≤10 dB (%) | ABG ≤20 dB (%) | Conclusion |
|---|---|---|---|---|---|
| Current study | Causse loop piston (44) | At least 6 months | 29.5 | 86.4 | Similar results with both prostheses. Matrix performed minimally better in ABG closure in 250 Hz. |
| Matrix (44) | 25.0 | 93.2 | |||
| Faramarzi et al | Causse loop piston (76) | At least 6 months | 57.5 | 92.5 | Similar results with both prostheses. Causse loop piston was moderately better in low frequencies. |
| Big easy piston (72) | 59.1 | 94.2 | |||
| Faramarzi et al | Causse loop piston (63) | At least 6 months | 36.5 | 80.9 | Similar results with both prostheses. |
| Titanium soft‐clip (57) | 38.6 | 86 | |||
| Schrötzlmair et al | Self‐crimping Nitinol | 70.7 days | 76.2 | 95.2 |
Better ABG closure with self‐crimping than with clip piston àWengen. |
| (Thermo dummy) (21) | |||||
| Titanium K‐piston (28) | 89.6 days | 53.6 | 89.3 | ||
|
Clip piston àWengen (13) | 163.1 days | 23.1 | 69.2 | ||
| Ying et al |
SMart (Teflon‐based piston Nitinol) (190) Manual‐crimp platinum; De la Cruz (145) | NRa | NR | NR |
Revision rate of 11% in the SMart group and 4% in the de la Cruz group. |
| Mangham | Platinum piston (144) | NR | 96 | 100 | No difference in ABG closure. Nitinol‐Teflon group had smaller mean ABG in lower frequencies. |
| Nitinol‐Teflon piston (44) | 92 | 100 | |||
| Fayad et al | SMart piston (306) | 5.6 months | 78.3 | 94.2 | No differences. |
| Richards' platinum piston (110) | 84.2 |
98 | |||
| Van Rompaey et al | Teflon (211) | 12 months | Overall | Overall | No differences. |
| Teflon wire (168) | 63.6 | 92.6 | |||
| Titanium (112) | |||||
| Clip piston (49) | |||||
| Smart (74) | |||||
| Huber et al | Conventional (75) | At least 12 months | 43 | 92 | No difference in ABG closure within 20 dB, but Nitinol was better in ABG closure within 10 dB. |
| Nitinol smart (75) | 71 | 94 | |||
| Mangham | Teflon piston 0.5 mm (74) | 1 year | 85 | NR | Teflon piston achieved better result than titanium clip piston. |
| Teflon piston 0.6 mm (74) | 91 | ||||
| Titanium clip piston (33) | 84 | ||||
| Tange and Grolman | Titanium K‐piston (63) | NR | 65 | 87.1 | No differences. |
| Clip piston àWengen (63) | 71 | 91 | |||
| Brown and Gantz | Platinum wire piston (39) | 20 months | NR | NR | No differences. |
| Nitinol piston (40) | 9 months | ||||
| Massey et al | Kurz titanium K‐piston (35) | 4 months | 71 | 97.1 | No differences. |
| Teflon platinum wire (183) | 86 | 97.8 | |||
| Zepeda‐López et al | Schuknecht Teflon wire piston (70) | NR | 57.1 | NR | Fluoroplastic Teflon was better in low frequencies and resulted in better ABG closure in all frequencies. |
| Fluoroplastic Teflon (76) | 93.4 |
ABG, air‐bone gap; NR, not reported.