Literature DB >> 19384538

[Stapes surgery : first experiences with the new Soft-CliP piston].

C Brase1, J Zenk, J Wurm, B Schick, H Iro, J Hornung.   

Abstract

BACKGROUND: The first hearing results with a new stapes prosthesis with clip function (Soft-CliP(R) piston) are presented. PATIENTS AND METHODS: This new prosthesis was used in 15 patients (mean age 45.2 years; range 21-63 years) undergoing routine stapes surgery. Soft-CliP(R) piston prostheses with a shaft diameter of 0.4 mm and a length ranging from 4.25 mm to 5.5 mm were used. Postoperative audiological testing and measurement of the air-bone gap were performed after an average of 47.3 days and compared with the preoperative values.
RESULTS: The median observed postoperative air-bone gap (ABG) was 8.33 dB +/-4.16 dB. All patients had less than 20 dB ABG and in 53.3% of cases was less than 10 dB. The operating time showed a clear difference between the left (66.5 min +/-37.79 min) and right ears (47.2 min +/-11.08 min). DISCUSSION: This new prosthesis design greatly facilitates a very difficult step in stapes surgery, the prosthesis fixation to the incus. The first postoperative hearing results are very promising but long-term results in a larger group of patients are still pending.

Entities:  

Mesh:

Year:  2009        PMID: 19384538     DOI: 10.1007/s00106-009-1899-y

Source DB:  PubMed          Journal:  HNO        ISSN: 0017-6192            Impact factor:   1.284


  17 in total

1.  Stapes surgery: how precisely do different prostheses attach to the long process of the incus with different instruments and different surgeons?

Authors:  Pingling Kwok; Ugo Fisch; Jürgen Strutz; John May
Journal:  Otol Neurotol       Date:  2002-05       Impact factor: 2.311

2.  Viruses and otosclerosis: chance association or true causal link?

Authors:  Alfio Ferlito; Wolfgang Arnold; Alessandra Rinaldo; Hans P Niedermeyer; Alexis Bozorg Grayeli; Kenneth O Devaney; Michael J McKenna; Matti Anniko; Jack L Pulec; Brian F McCabe; Paul van den Broek; John J Shea
Journal:  Acta Otolaryngol       Date:  2003-08       Impact factor: 1.494

3.  A new crimping device for stapedectomy prostheses.

Authors:  Karl-Bernd Huettenbrink; Dirk Beutner
Journal:  Laryngoscope       Date:  2005-11       Impact factor: 3.325

4.  A new self-retaining titanium clip stapes prosthesis.

Authors:  Daniel F À Wengen
Journal:  Adv Otorhinolaryngol       Date:  2007

5.  Hearing results in otosclerosis surgery after partial stapedectomy, total stapedectomy and stapedotomy.

Authors:  P Persson; H Harder; B Magnuson
Journal:  Acta Otolaryngol       Date:  1997-01       Impact factor: 1.494

6.  First experience with a new stapes clip piston in stapedotomy.

Authors:  W Grolman; R A Tange
Journal:  Otol Neurotol       Date:  2005-07       Impact factor: 2.311

7.  The loose wire syndrome.

Authors:  T M McGee
Journal:  Laryngoscope       Date:  1981-09       Impact factor: 3.325

8.  [First experiences with a new nickel-titanium piston with a shape memory feature].

Authors:  J Hornung; J Zenk; B Schick; J Wurm; H Iro
Journal:  HNO       Date:  2007-02       Impact factor: 1.284

9.  Titanium CliP piston versus platinum-ribbon Teflon piston: piston and fenestra size affect air-bone gap.

Authors:  Charles A Mangham
Journal:  Otol Neurotol       Date:  2008-01       Impact factor: 2.311

10.  A personal history of stapedectomy.

Authors:  J J Shea
Journal:  Am J Otol       Date:  1998-09
View more
  1 in total

1.  [Stapes surgery for otosclerosis using a new titanium prosthesis with a superelastic nitinol clip: First experiences].

Authors:  J Zirkler; T Rahne; S K Plontke
Journal:  HNO       Date:  2016-02       Impact factor: 1.284