| Literature DB >> 28229059 |
Mohammad Faramarzi1, Nafiseh Gilanifar1, Sareh Roosta1.
Abstract
INTRODUCTION: Different types of prosthesis are available for stapes replacement. Because there has been no published report on the efficacy of the titanium soft-clip vs the fluoroplastic Causse loop Teflon piston, we compared short-term hearing results of both types of prosthesis in patients who underwent stapedotomy due to otosclerosis.Entities:
Keywords: Fluoroplastic (Teflon) Causse loop piston; Otosclerosis; Stapedotomy; Titanium soft-clip
Year: 2017 PMID: 28229059 PMCID: PMC5307301
Source DB: PubMed Journal: Iran J Otorhinolaryngol ISSN: 2251-7251
Fig 1Consort trial flow diagram
Pre-operative and post-operative mean BC, AC and ABG (dB) in frequencies of 0.5–3 kHz.
| Pre-operative | Post-operative | Gain | ||
|---|---|---|---|---|
| Causse loop piston | Mean BC (dB) | 15.0±8.5 | 14.1±10.2 | 0.9±9.8 |
| Mean AC (dB) | 49.0±10.6 | 31.9±14.6 | 17.1±16.1 | |
| Mean ABG (dB) | 34.1±9.2 | 17.8±7.6 | 16.3±11.0 | |
| Soft-clip | Mean BC (dB) | 15.4±5.5 | 12.1±7.5 | 3.3±5.7 |
| Mean AC (dB) | 46.1±7.7 | 30.3±11.1 | 15.9±12.7 | |
| Mean ABG (dB) | 30.8±7.9 | 18.1±6.4 | 12.6±9.8 |
Values are Mean±SD, AC: Air Conduction, BC: Bone Conduction, ABG: Air-Bone Gap
Post-operative AC, BC gain (dB) and ABG improvement (dB) in each frequency
|
|
|
|
|
|
| ||
|---|---|---|---|---|---|---|---|
| Causse loop piston | AC gain | 19.2 | 21.8 | 19.6 | 15.5 | 11.6 | 9.6 |
| Soft-clip | 15.9 | 17.6 | 18.3 | 14.3 | 13.3 | 14.3 | |
| P-value | 0.274 | 0.155 | 0.669 | 0.663 | 0.543 | 0.144 | |
| Causse loop piston | BC gain | -0.1 | -0.7 | 0.4 | 4.0 | -0.3 | -2.5 |
| Soft-clip | 4.1 | 3.0 | 3.0 | 4.1 | 2.8 | -0.2 | |
| P-value | 0.000 | 0.006 | 0.111 | 0.960 | 0.109 | 0.240 | |
| Causse loop piston | ABG improvement | 19.3 | 22.5 | 19.2 | 11.5 | 11.9 | 12.1 |
| Soft-clip | 11.7 | 14.6 | 15.2 | 10.2 | 10.4 | 14.6 | |
| P-value | 0.007 | 0.001 | 0.104 | 0.528 | 0.457 | 0.356 |
Values are Means, AC: Air Conduction, BC: Bone Conduction, ABG: Air-Bone Gap
Mean post-operative ABG distribution in frequencies of 0.5–3 kHz.
|
|
|
|
|---|---|---|
| ≤ 10 | 23(36.5) | 22(38.6) |
| 11–20 | 28(44.4) | 27(47.4) |
| 21–30 | 8(12.7) | 5(8.8) |
| > 30 | 4(6.3) | 3(5.3) |
Values are N(%), ABG: Air-Bone Gap
Pre-operative and post-operative SDS and SRT
|
|
|
| ||
|---|---|---|---|---|
| Causse loop piston | SDS (%) | 92.5± 8.0 | 94.0± 10.6 | 1.5 ± 12.8 |
| SRT (dB) | 50.6± 11.0 | 31.8± 14.6 | 18.8 ±17.4 | |
| Soft-clip | SDS (%) | 95.1± 4.0 | 96.0± 5.4 | 0.9 ±6.0 |
| SRT (dB) | 48.0± 9.6 | 29.6± 10.3 | 18.5 ±14.4 |
Values are Mean±SD, SDS: Speech Discrimination Scores, SRT: Speech Reception Thresholds
Literature review of post-operative hearing outcomes in stapes surgery regarding the type of prosthesis
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| Current study | Causse loop piston (63) | At least | 36.5 | 80.9 | Similar good results with both prostheses. Causse loop piston was moderately better at low frequencies |
| Mangham (2010)(7) | Platinum piston (144) | NR | 96 | 100 | No significant difference in ABG closure (better mean ABG in lower frequencies in Nitinol–Teflon group) |
| Fayad et al | SMart piston (306) | 5.6 months | 78.3 | 94.2 | Similar audiometric outcomes |
| Van Rompaey | Teflon (211) | 12 months | Overall | Overall | 5 group had no difference in ABG closure |
| Mangham | Teflon piston 0.5 mm (74) | 1 year | 85 | NR | Teflon piston achieved better result than Titanium clip piston |
| Schrotzlmair | Self -crimping Nitinol (Thermo Dummy) (21) | 70.7 days | 76.2 | 95.2 | ABG closure was better in self –crimping Nitinol especially in comparison with clip piston (a`Wengen) |
| Ying et al (2010)(12) | SMart (Teflon-based piston Nitinol) (190) | NR | NR | NR | 11% Revision rate was seen in SMart group and 4% in De La Cruz group |
| Huber et al (2008)(13) | Conventional (75) | At least | 43 | 92 | Similar in ABG closure within 20 dB. better ABG closure within 10 db for Nitinol SMart |
| Tange and Grolman | Titanium K-piston (63) | NR | 65 | 87.1 | Similar audiometric outcomes |
| Brown and Gantz (2007)(5) | Platinum wire piston (39) | 20 months | NR | NR | Similar audiometric outcomes |
| Massey et al | Kurz titanium K- piston (35) | 4 months | 71 | 97.1 | Similar audiometric outcomes |
| Zepeda-Lopez et al (2005)(15) | Schuknecht Teflon wire piston (70) | NR | 57.1 | NR | Fluoroplastic Teflon had better ABG closure in all frequencies |
NR: not reported