A global calibration model for a-Si EPIDs used for transit dosimetry.

S M J J G Nijsten1, W J C van Elmpt, M Jacobs, B J Mijnheer, A L A J Dekker, P Lambin, A W H Minken.   

Abstract

Electronic portal imaging devices (EPIDs) are not only applied for patient setup verification and detection of organ motion but are also increasingly used for dosimetric verification. The aim of our work is to obtain accurate dose distributions from a commercially available amorphous silicon (a-Si) EPID for transit dosimetry applications. For that purpose, a global calibration model was developed, which includes a correction procedure for ghosting effects, field size dependence and energy dependence of the a-Si EPID response. In addition, the long-term stability and additional buildup material for this type of EPID were determined. Differences in EPID response due to photon energy spectrum changes have been measured for different absorber thicknesses and field sizes, yielding off-axis spectrum correction factors based on transmission measurements. Dose measurements performed with an ionization chamber in a water tank were used as reference data, and the accuracy of the dosimetric calibration model was determined for a large range of treatment conditions. Gamma values using 3% as dose-difference criterion and 3 mm as distance-to-agreement criterion were used for evaluation. The field size dependence of the response could be corrected by a single kernel, fulfilling the gamma evaluation criteria in case of virtual wedges and intensity modulated radiation therapy fields. Differences in energy spectrum response amounted up to 30%-40%, but could be reduced to less than 3% using our correction model. For different treatment fields and (in)homogeneous phantoms, transit dose distributions satisfied in almost all situations the gamma criteria. We have shown that a-Si EPIDs can be accurately calibrated for transit dosimetry purposes.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17985633     DOI: 10.1118/1.2776244

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  A generalized calibration procedure for in vivo transit dosimetry using siemens electronic portal imaging devices.

Authors:  Andrea Fidanzio; Francesca Greco; Laura Gargiulo; Savino Cilla; Domenico Sabatino; Massimo Cappiello; Cinzia Di Felice; Elisabetta Di Castro; Luigi Azario; Mariateresa Russo; Luciano Pompei; Guido D'Onofrio; Angelo Piermattei
Journal:  Med Biol Eng Comput       Date:  2010-11-04       Impact factor: 2.602

2.  The use of EPID-measured leaf sequence files for IMRT dose reconstruction in adaptive radiation therapy.

Authors:  Louis Lee; Weihua Mao; Lei Xing
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

3.  Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules.

Authors:  Janett Liebich; Jörg Licher; Christian Scherf; Eugen Kara; Nadine Koch; Claus Rödel; Ulla Ramm
Journal:  Strahlenther Onkol       Date:  2011-04-26       Impact factor: 3.621

4.  Radioluminescence characterization of in situ x-ray nanodosimeters: Potential real-time monitors and modulators of external beam radiation therapy.

Authors:  Jeffrey S Souris; Shih-Hsun Cheng; Charles Pelizzari; Nai-Tzu Chen; Patrick La Riviere; Chin-Tu Chen; Leu-Wei Lo
Journal:  Appl Phys Lett       Date:  2014-11-20       Impact factor: 3.791

5.  Transit dosimetry in dynamic IMRT with an a-Si EPID.

Authors:  Mahsheed Sabet; Pejman Rowshanfarzad; Fred W Menk; Peter B Greer
Journal:  Med Biol Eng Comput       Date:  2014-05-31       Impact factor: 2.602

6.  Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software.

Authors:  Mareike Held; Joey Cheung; Angelica Perez Andujar; François Husson; Olivier Morin
Journal:  Cureus       Date:  2018-02-02

7.  aSi-EPID transit signal calibration for dynamic beams: a needful step for the IMRT in vivo dosimetry.

Authors:  Francesca Greco; Angelo Piermattei; Luigi Azario; Lorenzo Placidi; Savino Cilla; Rocchina Caivano; Vincenzo Fusco; Andrea Fidanzio
Journal:  Med Biol Eng Comput       Date:  2013-07-09       Impact factor: 2.602

8.  Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry.

Authors:  Tae Seong Baek; Eun Ji Chung; Jaeman Son; Myonggeun Yoon
Journal:  Radiat Oncol       Date:  2014-12-04       Impact factor: 3.481

9.  Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry.

Authors:  Heeteak Chung; Jonathan Li; Sanjiv Samant
Journal:  J Appl Clin Med Phys       Date:  2011-04-08       Impact factor: 2.102

10.  Impact of backscattered radiation from the bunker structure on EPID dosimetry.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.