Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules.

Janett Liebich1, Jörg Licher, Christian Scherf, Eugen Kara, Nadine Koch, Claus Rödel, Ulla Ramm.   

Abstract

BACKGROUND: An electronic portal imaging device (EPID) is used to control for patient setup and positioning during fractionated radiotherapy. Due to the rising complexity and conformity of irradiation techniques, the demand for an accurate verification of the dose delivered to the patient has also increased. The purpose of this study was to investigate a simple guidance for dosimetry with an Elekta iViewGT™ EPID using commercial software modules.
MATERIAL AND METHODS: EPID measurements were performed using an Elekta iViewGT™ EPID on a linear accelerator with 6 MV x-ray beam. The EPID signal was studied for reproducibility, as well as characteristics as a function of dose, dose rate, and field size. A series of experiments, comparing the response of the flat panel imager and ionization chamber measurements of dose, determine the parameters for the calibration model. EPID measurements were also compared with calculations of the treatment planning system.
RESULTS: We found a stable response of the EPID signal over a period of 14 months. It showed nonlinearity depending on dose up to 6.8%. There were low oscillations up to 1.2% depending on dose rate. For all fields, the calibrated flat panel profiles match the measured and calculated dose profiles with maximum deviation of 2-3% for the in-field region. In the high gradient areas, higher differences up to 6% were found.
CONCLUSIONS: The gamma evaluation indicates good correlation between predicted and acquired EPID images. The EPID-based pretreatment IMRT verification method will help to improve the quality assurance procedure.

Entities:  

Mesh:

Year:  2011        PMID: 21544527     DOI: 10.1007/s00066-011-2176-z

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  27 in total

1.  IMRT verification with a camera-based electronic portal imaging system.

Authors:  M Partridge; J R Symonds-Tayler; P M Evans
Journal:  Phys Med Biol       Date:  2000-12       Impact factor: 3.609

2.  Calibration of an amorphous-silicon flat panel portal imager for exit-beam dosimetry.

Authors:  Josephine Chen; Cynthia F Chuang; Olivier Morin; Michèle Aubin; Jean Pouliot
Journal:  Med Phys       Date:  2006-03       Impact factor: 4.071

3.  Quality Control of Portal Imaging with PTW EPID QC PHANTOM.

Authors:  Csilla Pesznyák; Gábor Fekete; Arpád Mózes; Balázs Kiss; Réka Király; István Polgár; Pál Zaránd; Arpád Mayer
Journal:  Strahlenther Onkol       Date:  2009-02-18       Impact factor: 3.621

Review 4.  A literature review of electronic portal imaging for radiotherapy dosimetry.

Authors:  Wouter van Elmpt; Leah McDermott; Sebastiaan Nijsten; Markus Wendling; Philippe Lambin; Ben Mijnheer
Journal:  Radiother Oncol       Date:  2008-08-14       Impact factor: 6.280

5.  Overestimation of low-dose radiation in intensity-modulated radiotherapy with sliding-window technique.

Authors:  Hilke Vorwerk; Daniela Wagner; Björn Seitz; Hans Christiansen; Hendrik A Wolff; Clemens F Hess
Journal:  Strahlenther Onkol       Date:  2009-12       Impact factor: 3.621

6.  A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments.

Authors:  Markus Wendling; Leah N McDermott; Anton Mans; Jan-Jakob Sonke; Marcel van Herk; Ben J Mijnheer
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

7.  Dosimetric characteristics of a liquid-filled electronic portal imaging device.

Authors:  M Essers; B R Hoogervorst; M van Herk; H Lanson; B J Mijnheer
Journal:  Int J Radiat Oncol Biol Phys       Date:  1995-12-01       Impact factor: 7.038

8.  Transmission dosimetry with a liquid-filled electronic portal imaging device.

Authors:  M Essers; R Boellaard; M van Herk; H Lanson; B Mijnheer
Journal:  Int J Radiat Oncol Biol Phys       Date:  1996-03-01       Impact factor: 7.038

9.  Experimental investigation of the response of an amorphous silicon EPID to intensity modulated radiotherapy beams.

Authors:  Peter B Greer; Philip Vial; Lyn Oliver; Clive Baldock
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

10.  Comprehensive fluence model for absolute portal dose image prediction.

Authors:  K Chytyk; B M C McCurdy
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

View more
  3 in total

1.  Phase-specific cone beam computed tomography reduces reconstructed volume loss of moving phantom.

Authors:  H-L Chao; W-L Chen; C-C Hu; J-K Wu; C-J Wu; J C-H Cheng
Journal:  Strahlenther Onkol       Date:  2011-12-24       Impact factor: 3.621

2.  [Current requirements for image management in radiotherapy].

Authors:  V Steil; F Röhner; F Schneider; F Wenz; F Lohr; G Weisser
Journal:  Strahlenther Onkol       Date:  2012-03-30       Impact factor: 3.621

3.  Three-dimensional surface scanning for accurate patient positioning and monitoring during breast cancer radiotherapy.

Authors:  C Gaisberger; P Steininger; B Mitterlechner; S Huber; H Weichenberger; F Sedlmayer; H Deutschmann
Journal:  Strahlenther Onkol       Date:  2013-06-07       Impact factor: 3.621

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.