| Literature DB >> 17823030 |
Charles R Bragdon1, Meridith E Greene, Andrew A Freiberg, William H Harris, Henrik Malchau.
Abstract
This study used radiostereometric analysis (RSA) to compare the femoral head penetration of 28- vs 36-mm-diameter femoral heads into highly cross-linked polyethylene in 2 groups of total hip arthroplasty patients. Thirty patients were enrolled in this RSA study using highly cross-linked polyethylene (Longevity, Zimmer Inc, Warsaw, Idaho) against either 28- or 36-mm-diameter cobalt chrome femoral heads. At 3-year follow-up, there was no significant difference in the total average femoral head penetration, including both creep and wear, using 3 methods of RSA measurement between the 2 groups. Importantly, after bedding-in, there was no further significant increase in the amount of femoral head penetration (ie, wear) with either head size between years 1 and 3. There were no radiographic signs of lysis or radiolucencies at a minimum 3-year follow-up.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17823030 DOI: 10.1016/j.arth.2007.03.009
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757