Literature DB >> 17719394

Use of straight and curved 3-dimensional titanium miniplates for fracture fixation at the mandibular angle.

Juergen Zix1, Olivier Lieger, Tateyuki Iizuka.   

Abstract

PURPOSE: The aim of this follow-up study was to evaluate the clinical usefulness of a new type of 3-dimensional (3D) miniplate for open reduction and monocortical fixation of mandibular angle fractures. PATIENTS AND METHODS: In 20 consecutive patients, noncomminuted mandibular angle fractures were treated with open reduction and fixation using a 2 mm 3D miniplate system in a transoral approach. All patients were systematically monitored until 6 months postoperatively. Among the outcome parameters recorded were infection, hardware failure, wound dehiscence, and sensory disturbance of the inferior alveolar nerve.
RESULTS: The mean operation time from incision to wound closure was 65 minutes. Two patients had a mucosal wound dehiscence with no consequences. None developed an infection requiring a plate removal. All but 2 patients had normal sensory function 3 months after surgery. Plate fracture occurred in one patient in whom a preceding surgical removal of the third molar had been the reason for the mandibular fracture. In the absence of clinical symptoms, the patient declined plate removal. On final follow-up, fracture healing was considered clinically complete in all patients.
CONCLUSIONS: The 3D plating system described here is suitable for fixation of simple mandibular angle fractures and is an easy-to-use alternative to conventional miniplates. The system may be contraindicated in patients in whom insufficient interfragmentary bone contact causes minor stability of the fracture.

Entities:  

Mesh:

Year:  2007        PMID: 17719394     DOI: 10.1016/j.joms.2007.03.013

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  29 in total

1.  Efficacy of 3-Dimensional plates over Champys miniplates in mandibular anterior fractures.

Authors:  Dhananjay H Barde; Anupama Mudhol; Fareedi Mukram Ali; R S Madan; Sanjay Kar; Farheen Ustaad
Journal:  J Int Oral Health       Date:  2014-02-26

2.  Versatility of Titanium 3D Plate in Comparison with Conventional Titanium Miniplate Fixation for the Management of Mandibular Fracture.

Authors:  Kamal Malhotra; Ashish Sharma; Girish Giraddi; Ashish Kumar Shahi
Journal:  J Maxillofac Oral Surg       Date:  2012-03-29

3.  Treatment of mandibular angle fracture with a 2mm, 3 dimensional rectangular grid compression miniplates: A prospective clinical study.

Authors:  Samir Mansuri; Abdul Mujeeb Abdulkhayum; Giath Gazal; Mohammed Abid Zahir Hussain
Journal:  J Int Oral Health       Date:  2013-12-26

4.  Biomechanical comparison of four mandibular angle fracture fixation techniques.

Authors:  Jose Luis Muñante-Cardenas; Luis Augusto Passeri
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2014-11-20

5.  A Comparative Study of 3-Dimensional Titanium Versus 2-Dimensional Titanium Miniplates for Open Reduction and Fixation of Mandibular Parasymphysis Fracture.

Authors:  Yogesh Mittal; K George Varghese; S Mohan; N Jayakumar; Somil Chhag
Journal:  J Maxillofac Oral Surg       Date:  2015-03-27

6.  Treatment of Mandibular Angle Fractures with Single Three-Dimensional Locking Miniplates without Maxillomandibular Fixation: How Much Fixation Is Required?

Authors:  Sanjay Rastogi; Sam Paul; Sumedha Kukreja; Karun Aggarwal; Rupshikha Choudhury; Amit Bhugra; Niranjana Prasad Indra B; Moazzam Jawaid
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2017-03-29

7.  Comparison of 3D plate and locking plate in treatment of mandibular fracture-a clinical study.

Authors:  Sakshi Aggarwal; Manpreet Singh; Palash Modi; Esha Walia; Rachit Aggarwal
Journal:  Oral Maxillofac Surg       Date:  2017-08-07

8.  Use of 3-d plate in displaced angle fracture of mandible.

Authors:  Uma Shanker Pal; R K Singh; Satish Dhasmana; Somdipto Das; Sanjib K Das
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2013-01-15

9.  Complications of the use of trans-osseous wire osteosynthesis in the managementof compound, unfavorable and non-comminuted mandibular angle fractures.

Authors:  Charles E Anyanechi; Otasowie D Osunde; Birch D Saheeb
Journal:  Ghana Med J       Date:  2016-09

10.  Management of mandibular angle fractures using a 1.7 mm 3-dimensional strut plate.

Authors:  Varnika Pandey; Ongkila Bhutia; Shakil Ahmed Nagori; Ashu Seith; Ajoy Roychoudhury
Journal:  J Oral Biol Craniofac Res       Date:  2015-11-27
View more