Zakithi Lungile Mpumelelo Msimang1, Debbie van der Merwe1, Nkosingiphile Maphumulo1.
Abstract
There are well established dosimetry reference standards for broad beams; however, there are no reference standards that can be used for both broad and small fields. The variation of the equivalent square fields and field output factors in small static photon fields when using a synthetic diamond, an electron diode, and ionization chambers (pin point, semiflex, and liquid filled) was investigated over time. Data from this study were compared to the data from other hospitals in the country and standard data sets, i.e., the British Journal of Radiology Supplement No. 25 of 1996 (BJR25) and the Radiological Physics Centre (RPC) 2012 data. The results showed that reliance on one detector and one measurement session, could yield incorrect field output factors (FOFs) for small fields. At least one of the detectors should be a solid state type with published field output correction factors and at least three measurement sessions should be performed for each FOF data point. Comparing measured data with published datasets, like RPC, will assist in verifying data. BJR25 datasets should not be used for S clin ≤4 cm. Copyright:Entities:
Keywords: Accurate; equivalent square field; field output factor; reference dosimetry
Year: 2021 PMID: 34267489 PMCID: PMC8240910 DOI: 10.4103/jmp.JMP_93_20
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Specifications of the ionization chambers used during the study as provided by the manufacturer
| Ionisation chamber type | Central electrode | Nominal sensitive volume | Dimensions of sensitive volume | Wall material and thickness | Reference point |
|---|---|---|---|---|---|
| PTW 31016 | Aluminum | 0.016 cm3 | Radius 1.45 mm, length 2.9 mm | PMMA + graphite, 85 mg/cm2 | 2.4 mm from chamber tip, on chamber axis |
| PTW 31018 | Graphite | 0.0017 cm3 | Radius 1.25 mm, depth 0.35 mm | Polystyrene + graphite + varnish, 107 mg/cm2 | 0.975 mm from the entrance window, on chamber axis |
| PTW 31021 | Aluminum | 0.07 cm3 | Radius 2.4 mm, length 4.8 mm | PMMA + graphite, 84 mg/cm² | 3.45 mm from the chamber tip, on chamber axis |
Specifications of the solid-state detectors used during the study as provided by the manufacturer
| Solid-state detector description | Dimensions of a detectors | Nominal sensitive volume | Dimensions of sensitive volume | Reference point |
|---|---|---|---|---|
| PTW 60012 unshielded diode E with a p-type silicone diode | Diameter of 7 mm, length 45.5 mm | 0.0025 mm3 | Radius 0.56 mm and a depth of 2.5 µm | 0.6 mm from the detector tip |
| PTW 60019 microDiamond | Diameter of 7 mm, length 45.5 mm | 0.004 mm3 | Radius of 1.1 mm and thickness of 1 µm | 1 mm from the detector tip |
Sclin measured in three measurement sessions using PTW 60012, 60019, 31016, 31018, and 31021 detectors
| Set field size (cm) | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Session 1 | Session 2 | Session 3 | ||||||||||||
| PTW 60012 | PTW 60019 | PTW 31021 | Sth DEV | PTW 60012 | PTW 31018 | PTW 31021 | Sth DEV | PTW 60012 | PTW 60019 | PTW 31016 | PTW 31018 | PTW 31021 | Sth DEV | |
| 10 × 10 | 10.01 | 10.03 | 10.26 | 0.14 | 9.75 | 9.73 | 9.67 | 0.04 | 9.84 | 9.84 | 9.83 | 9.84 | 9.79 | 0.02 |
| 6 × 6 | 5.99 | 5.97 | 6.19 | 0.12 | 5.78 | 5.77 | 5.71 | 0.04 | 5.86 | 5.86 | 5.83 | 5.85 | 5.79 | 0.03 |
| 4 × 4 | 3.94 | 3.96 | 4.14 | 0.11 | 3.78 | 3.76 | 3.70 | 0.04 | 3.85 | 3.85 | 3.82 | 3.85 | 3.79 | 0.03 |
| 3 × 3 | 2.95 | 2.92 | 3.11 | 0.10 | 2.78 | 2.77 | 2.70 | 0.04 | 2.85 | 2.85 | 2.84 | 2.86 | 2.80 | 0.02 |
| 2 × 2 | 1.93 | 1.92 | 2.12 | 0.11 | 1.79 | 1.77 | 1.72 | 0.03 | 1.86 | 1.86 | 1.85 | 1.86 | 1.81 | 0.02 |
| 1 × 1 | 0.98 | 0.96 | 1.18 | 0.12 | 0.84 | 0.84 | 0.85 | 0.01 | 0.91 | 0.91 | 0.91 | 0.92 | 0.91 | 0.004 |
| 0.6 × 0.6 | 0.61 | 0.65 | 0.84 | 0.12 | 0.51 | 0.54 | 0.64 | 0.07 | 0.59 | 0.59 | 0.60 | 0.93 | 0.17 | |
Std DEV: Standard deviation
Field output factor derived from three measurement sessions using PTW 60012, 60019, 31016, 31018, and 31021 detectors, as a function of set field size
| Set field size (cm) | FOF | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Session 1 | Session 2 | Session 3 | |||||||||||
| PTW 60012 | PTW 60019 | PTW 31021 | Sth DEV | PTW 60012 | PTW 31018 | Sth DEV | PTW 60012 | PTW 60019 | PTW 31016 | PTW 31018 | PTW 31021 | Sth DEV | |
| 10 × 10 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | 1.000 | |||
| 6 × 6 | 0.917 | 0.918 | 0.917 | 0.001 | 0.918 | 0.919 | 0.001 | 0.918 | 0.918 | 0.919 | 0.914 | 0.918 | 0.002 |
| 4 × 4 | 0.862 | 0.862 | 0.862 | 0.0002 | 0.862 | 0.863 | 0.001 | 0.862 | 0.863 | 0.863 | 0.859 | 0.863 | 0.002 |
| 3 × 3 | 0.827 | 0.830 | 0.827 | 0.002 | 0.831 | 0.832 | 0.001 | 0.831 | 0.831 | 0.833 | 0.830 | 0.832 | 0.001 |
| 2 × 2 | 0.782 | 0.785 | 0.781 | 0.002 | 0.785 | 0.786 | 0.001 | 0.784 | 0.786 | 0.788 | 0.788 | 0.786 | 0.002 |
| 1 × 1 | 0.636 | 0.645 | 0.628 | 0.009 | 0.619 | 0.625 | 0.005 | 0.634 | 0.633 | 0.648 | 0.646 | 0.633 | 0.01 |
| 0.6 × 0.6 | 0.467 | 0.482 | 0.428 | 0.03 | 0.387 | 0.402 | 0.01 | 0.449 | 0.439 | 0.453 | 0.371 | 0.04 | |
FOF: Field output factor, Std DEV: Standard deviation
Figure 1Field output factor versus Sclin for the third session
Figure 2Field output factor versus equivalent square field (Sclin), from this study (a), other centres (D, E and E-M), and published data from British Journal of Radiology Supplement No. 25 of 1996[12] and Radiological Physics Centre (Siemens, Varian and Elekta)[13]