Pacemaker longevity: the world's longest-lasting VVI pacemaker.

David Katz1, Toshio Akiyama.   

Abstract

OBJECTIVE: To describe VVI-pacemaker longevity by model type at our institution and report on a long-lasting model and the longest-lasting pacemaker to be described in the literature.
BACKGROUND: Cardiac pacemakers are becoming increasingly common in the United States. Presently their batteries are expected to last up to 12 years. Pacemaker generator change is associated with increased cost to the health care system and is inconvenient for patients.
METHODS: After identifying a group of very long-lasting CPI Microlith 605 VVI pulse generators, we reviewed records on all patients who had either Guidant or Medtronic pulse generator explantation at our institution over a 10-year period. Average longevities were calculated for all VVI pacemakers, four common VVI models, and the CPI Microlith 605.
RESULTS: A total of 105 VVI-programmed pacemakers were identified. Their average longevity was 7.2 years. The two most common Medtronic VVI-programmed pacemakers explanted were the Thera (7.1 years) and Kappa (7.3 years). The two most common Guidant/CPI models were the Vigor (4.2 years) and Discovery (5.7 years). The CPI Microlith 605 (19.2 years) lasted more than 26 years in one patient before being explanted.
CONCLUSION: At a time when pacemakers are being used more frequently, pacemaker longevity may decrease as a result of the use of dual-chamber pacing systems. In our study, the CPI Microlith 605 had an average longevity more than twice that of all other VVI pacemakers. We also report on a pulse generator that lasted 26.3 years.

Entities:  

Mesh:

Year:  2007        PMID: 17617067      PMCID: PMC6931941          DOI: 10.1111/j.1542-474X.2007.00165.x

Source DB:  PubMed          Journal:  Ann Noninvasive Electrocardiol        ISSN: 1082-720X            Impact factor:   1.468


  10 in total

1.  Extension of generator longevity by use of high impedance ventricular leads.

Authors:  M Scherer; K Ezziddin; A Klesius; M Skupin; S Helms; A Moritz; H G Olbrich
Journal:  Pacing Clin Electrophysiol       Date:  2001-02       Impact factor: 1.976

2.  Runaway pulse generator malfunction resulting from undetected battery depletion.

Authors:  Pugazhendhi Vijayaraman; Kedarnath Vaidya; Soo G Kim; Jay N Gross; Kevin J Ferrick; Eugen Palma; Seymour Furman; John D Fisher
Journal:  Pacing Clin Electrophysiol       Date:  2002-02       Impact factor: 1.976

3.  The impact of pulse generator longevity on the long-term costs of cardiac pacing.

Authors:  A M Gillis
Journal:  Pacing Clin Electrophysiol       Date:  1996-10       Impact factor: 1.976

4.  A randomized study on the effects of pacemaker programming to a lower output on projected pulse generator longevity.

Authors:  A Schuchert; T Meinertz
Journal:  Pacing Clin Electrophysiol       Date:  2001-08       Impact factor: 1.976

5.  Longevity of dual chamber pacemakers: device and patient related determinants.

Authors:  M Kindermann; B Schwaab; M Berg; G Fröhlig
Journal:  Pacing Clin Electrophysiol       Date:  2001-05       Impact factor: 1.976

6.  Effect of modern pacing algorithms on generator longevity: a predictive analysis.

Authors:  Dan Gelvan; Eugene Crystal; Barbados Dokumaci; Yuval Goldshmid; I Eli Ovsyshcher
Journal:  Pacing Clin Electrophysiol       Date:  2003-09       Impact factor: 1.976

7.  Permanent pacemakers in centenarians.

Authors:  J L Cobler; T Akiyama; G W Murphy
Journal:  J Am Geriatr Soc       Date:  1989-08       Impact factor: 5.562

8.  Late complications following permanent pacemaker implantation or elective unit replacement.

Authors:  A A Harcombe; S A Newell; P F Ludman; T E Wistow; L D Sharples; P M Schofield; D L Stone; L M Shapiro; T Cole; M C Petch
Journal:  Heart       Date:  1998-09       Impact factor: 5.994

9.  The influence of high versus normal impedance ventricular leads on pacemaker generator longevity.

Authors:  Thomas Berger; Franz X Roithinger; Herwig Antretter; Herbert Hangler; Otmar Pachinger; Florian Hintringer
Journal:  Pacing Clin Electrophysiol       Date:  2003-11       Impact factor: 1.976

10.  Pacemakers and implantable cardioverter defibrillators: device longevity is more important than smaller size: the patient's viewpoint.

Authors:  David M Wild; John D Fisher; Soo G Kim; Kevin J Ferrick; Jay N Gross; Eugen C Palma
Journal:  Pacing Clin Electrophysiol       Date:  2004-11       Impact factor: 1.976

  10 in total
  2 in total

1.  Pacing and Sensing of Human Heart for over 31 Years with the Same Apparatus (Generator and Lead).

Authors:  Evangelos Papasteriadis; Panagiotis Margos
Journal:  Case Rep Cardiol       Date:  2015-10-26

2.  Single-Chamber Cardiac Pacemaker Implantation in a Donkey with Complete AV Block: A Long-Term Follow-Up.

Authors:  Markéta Sedlinská; Radovan Kabeš; Miroslav Novák; Filip Kološ; Pavlína Melková
Journal:  Animals (Basel)       Date:  2021-03-09       Impact factor: 2.752

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.