Artifacts in spine magnetic resonance imaging due to different intervertebral test spacers: an in vitro evaluation of magnesium versus titanium and carbon-fiber-reinforced polymers as biomaterials.

Thorsten Ernstberger1, Gottfried Buchhorn, Gabert Heidrich.   

Abstract

INTRODUCTION: Intervertebral spacers are made of different materials, which can affect the postfusion magnetic imaging (MRI) scans. Susceptibility artifacts especially for metallic implants can decrease the image quality. This study aimed to determine whether magnesium as a lightweight and biocompatible metal is suitable as a biomaterial for spinal implants based on its MRI artifacting behavior.
MATERIALS AND METHODS: To compare artifacting behaviors, we implanted into one porcine cadaveric spine different test spacers made of magnesium, titanium, and carbon-fiber-reinforced polymers (CFRP). All test spacers were scanned using two T1-TSE MRI sequences. The artifact dimensions were traced on all scans and statistically analyzed.
RESULTS: The total artifact volume and median artifact area of the titanium spacers were statistically significantly larger than magnesium spacers (p < 0.001), while magnesium and CFRP spacers produced almost identical artifacting behaviors (p > 0.05).
CONCLUSION: Our results suggest that spinal implants made with magnesium alloys will behave more like CFRP devices in MRI scans. Given its osseoconductive potential as a metal, implant alloys made with magnesium would combine the advantages to the two principal spacer materials currently used but without their limitations, at least in terms of MRI artifacting.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19468722      PMCID: PMC3085728          DOI: 10.1007/s00234-009-0537-4

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


Introduction

Spinal fusion devices such as implantable interbody spacers are well-established and routinely used by spine surgeons to keep adjacent vertebrae spaced apart while bone ingrowth and fusion take place. Such spacers also provide weight-bearing support between adjacent vertebrae. In this context, titanium alloys as well as nonmetal materials like carbon-fiber-reinforced polymers (CFRP) are commonly used [1, 3, 8, 11, 14, 17, 21]. These biomaterials have enjoyed clinical success and rapid widespread use by improving patient outcomes. However, these materials have clinical and radiological limitations. Titanium is an excellently bioinert material that exhibits high biocompatibility. Titanium spacers produce good bone ingrowth without bone grafting. However, in magnetic resonance imaging (MRI) studies, titanium-based implants tend to cause distortion of the magnetic field which may obscure normal regional anatomy. These properties pose difficulties in the postoperative MRI follow-up and evaluation of the fusion process due to the artifacting it causes [4, 6, 7, 12, 15]. The other principal material used for spacers consists of CFRP. Spacers made of this nonmetallic biomaterial are not associated with the postoperative diagnostic problems of titanium because carbon produces a very low rate of artifact reactions which allows easier evaluation of the fusion process by MRI [4]. Carbon’s modulus of elasticity affords good load bearing with sufficient hardness. But unlike titanium, carbon spacers undergo poor osteointegration because a soft tissue interface develops around the material surface that prevents direct ingrowth of bone. As a result, carbon spacers have to be filled with bone allografts to achieve long-term stability [1, 4]. CFRP implants have therefore been reviewed very critically in the literature [17]. Surgeons, over a century ago, recognized the potential of the lightweight metal magnesium as a biocompatible, osteoconductive, and degradable implant material [9]. In 1907, Lambotte was the first to introduce magnesium-based orthopedic devices; using a pure magnesium plate, he secured a bone fracture of the lower leg with gold-plated nails [9]. A half a century later, magnesium-based metals were reported to have osteoconductive bioactivity and produce a more rapid formation of hard callus when used to support fractures in humans [20, 27]. The large amount of evidence supporting the clinical advantages of magnesium has been summarized in a recent review paper [18]. None of the studies to date have yet investigated the diagnostic behavior of magnesium in MRI. This situation motivated us to determine whether magnesium is a suitable biomaterial for spinal implants by studying its MRI artifacting behavior.

Material and methods

To evaluate the behavior of spacers made with a magnesium alloy, we compared their artifacting in diagnostic MRI scans with that of spacers made of a conventional titanium alloy and of CFRP. We consecutively implanted three spacers made of each of the three biomaterials dimensioned in small, medium, and large sizes in one cadaveric spine of a Gottingen minipig (Figs. 1a–c and 2). The three spacers in group I were made of a magnesiumaluminummanganese alloy (MgAlMn50), the three in group II of a titaniumaluminum–vanadium alloy (TiAl6V4), and those in group III of a CFRP.
Fig. 1

a–c Cylindrical EST implants. a Magnesium (implant group I). b Titanium (implant group II). c CFRP (implant group III)

Fig. 2

Cadaveric porcine spine model with an implanted medium titanium test cylinder

a–c Cylindrical EST implants. a Magnesium (implant group I). b Titanium (implant group II). c CFRP (implant group III) Cadaveric porcine spine model with an implanted medium titanium test cylinder Table 1 presents the implant characteristics. A cylinder was chosen as for spacer shape because cylinders have demonstrated lowest rate of MRI artifacting behavior [4]. The spacer sizes—small, medium, and large—were dimensioned the same for each group (height in centimeter × base area in square centimeter); and their implant volume (IV) in cubic centimeter and cross-sectional area (CSA) in square centimeter were calculated for each size (Table 1). The spacer sizes were dimensioned as listed after a Newman–Keuls multiple-comparison analysis showing that the selected sizes would produce significantly different artifacting behaviors (p < 0.001). Thus, a total of nine individual spacers were implanted, scanned by MRI, and evaluated for their artifacting behavior on the scans.
Table 1

Spacer dimensions.

Sizes for all groupsDimensions height × base area, cm × cm2Cross-sectional area (CSA), cm2Implant volume (IV), cm³
Small1.5 × 0.781.51.2
Medium2.0 × 1.132.42.3
Large2.5 × 1.543.83.5
Spacer dimensions.

Spacer implantation

For each serial MRI study, the cylindrical implant was placed exactly between two adjacent vertebrae of the cadaveric porcine spine. The spine with implant was then completely packed in a soft tissue mass and placed in a plastic container [5]. To create comparable trial conditions, markings were drawn on the container wall to demarcate the vertebrae and implant positions. These demarcations were used to define the median artifact area (MAA). The container with the spine implanted with each spacer was examined by serial MRI.

Magnetic resonance imaging

MRI was performed with a 1.5-T MRI (Magnetom Symphony, Siemens AG Medical Solutions, Erlangen, Germany). The T1w-TSE sequences were used to acquire a slice thickness of 3 mm (Fig. 3a–c) which included a first sequence (TR 600; TE 14; flip angle 15; band width 150) and a second sequence (TR 2,260, TE 14, flip angle 15, band width 150). We selected a matrix of 512 × 512 pixels combined with a field of view of 500 mm. The T1w-TSE sequence has been established to produce best imaging results for implants and the least amount of intrinsic artifacting [4, 7, 10, 12, 14, 16, 25].
Fig. 3

a–c Median MRI artifact range depicted in a selection of three large test implants

a–c Median MRI artifact range depicted in a selection of three large test implants Using a current version of DICOM reader software, one author (TE) measured the artifact area on the scan of each of the nine implants six times, i.e., a total of 54 individual tracings were recorded and analyzed. The measurements started with the slice with the first artifacting reaction and ended with the last slice exhibiting an artifact reaction. Corresponding to the respective implant CSA, the middle slice of all slices exhibiting artifact reactions was defined as the MAA for each implant. To calculate the total artifact volume (TAV) for each spacer, all artifact areas measured for that spacer were added and multiplied by the slice thickness of 3 mm according to the multisection slice technique described by Debatin [2]. The ratio of CSA to MAA and the ratio of IV to TAV were calculated and presented in tables (Table 1).

Statistical analysis

Newman–Keuls multiple comparisons were used to calculate intragroup differences in TAV and MAA (Table 2). t test correlations were performed to determine any intergroup differences regarding the implant materials (Table 2). A p value < 0.05 indicated a significant difference between the means of any two groups.
Table 2

Intragroup comparison of target variables.

Spacer materialSizeMAAa, cm2 (mean ± SD)Ratio CSA to MAATAVa, cm2 (mean ± SD)Ratio IV to TAV
Group I MgAlMn50 (n = 3)Small1.91 ± 0.041:1.31.83 ± 0.091:1.5
Medium3.26 ± 0.061:1.44.17 ± 0.091:1.8
Large4.06 ± 0.071:1.25.08 ± 0.151:1.3
Group II TiAl6V4 (n = 3)Small3.26 ± 0.041:2.25.71 ± 0.091:4.8
Medium4.61 ± 0.231:1.99.32 ± 0.101:4.1
Large5.54 ± 0.041:1.610.84 ± 0.131:2.9
Group III CFRP (n = 3)Small1.89 ± 0.071:1.31.81 ± 0.071:1.5
Medium3.18 ± 0.061:1.34.09 ± 0.111:1.7
Large4.06 ± 0.131:1.25.08 ± 0.131:1.3

CSA cross-sectional area, MAA median artifact area, IV implant volume, TAV total artifact volume, SD standard deviation

aNewman–Keuls multiple-comparison analysis p < 0.001

Intragroup comparison of target variables. CSA cross-sectional area, MAA median artifact area, IV implant volume, TAV total artifact volume, SD standard deviation aNewman–Keuls multiple-comparison analysis p < 0.001

Results

Table 1 presents the spacer dimensions. Table 2 shows the intragroup comparisons of target variables. Table 3 lists the results of the intergroup t test correlations between TAV and MAA in relation to spacer material. Mean artifacting behavior increased with spacer size. When magnesium was compared with titanium, there were significant differences in both MAA and TAV. When magnesium was compared with carbon, the differences were not significant. In fact, magnesium produces an artifacting behavior very similar to that of CFRP.
Table 3

Intergroup comparisons of artifacting behavior by t test correlation.

Spacer materialSizeMAA P valueaTAV P valuea
Group I vs. group IISmall≤0.001≤0.001
Medium≤0.001≤0.001
Large≤0.001≤0.001
Group I vs. group IIISmall0.590.61
Medium0.090.26
Large1.00.96

MAA median artifact area, TAV total artifact volume

aSignificance level p < 0.05

Intergroup comparisons of artifacting behavior by t test correlation. MAA median artifact area, TAV total artifact volume aSignificance level p < 0.05

Discussion

Spinal surgeons have not stopped searching for the optimum spacer material that combines high biocompatibility with artifact-free MRI imaging behavior in the implant environment. This study was conducted to determine whether cylindrical spacers made of the biomaterial magnesium are suitable as spinal implants by comparing their MRI artifacting with that of identically dimensioned spacers made of a titanium alloy and a carbon-fiber-reinforced polymer. In radiological spinal diagnostics, MRI is highly effective for clarifying postfusion questions regarding osseus and soft tissue structures in relation to implant position. A comparative in vitro study shows that MRI has a higher sensitivity than CT in detecting osseus changes in the implant’s direct surroundings [24]. Moreover, MRI is well suited to demonstrate myelopathies, inflammatory and infectious processes, and any neurodegenerative changes. The MRI imaging behavior of metallic spinal implants is obviously well documented in the literature [6, 10, 13–16, 19, 22, 23, 25]. However, the aims of the published studies differed in that most focused on determining sequence-related artifact size. In a comparative cadaveric artifact studies, Wang et al. [26] described the MRI behavior of intervertebral spacers made of titanium and tantalum. The authors concluded that T1- and T2-weighted spin echo images were suitable for both metals to visualize the neural structures of the spine with comparable amounts of artifact. The artifact rate of the titanium spacer as well as the tantalum spacer was primarily limited to the implant’s direct surroundings and anatomic neighboring structures were clearly distinguishable. In a phantom study by Rudisch et al. [15], the relevance of metallic artifacts and implant-related characteristics, such as implant material and position, was demonstrated in addition to effects caused by the selected MRI sequence. In materials with a higher magnetizability like titanium alloys, implant shape additionally has an effect on the range of MRI artifacts [5]. The results of this comparative study showed that implant material and volume both affected the MRI artifacting behavior of our cylindrical test spacers. It was also noted that the smaller the implant size, the smaller was the range of susceptibility artifacts produced. The ratios calculated in Table 2 prove that the magnesium metal alloy exhibited behavior artifacting that was more like a nonmetal. In this context, the magnesium test spacer produced significantly less artifact compared to the titanium implant. Given its osseoconductive potential as a metal [9, 18], implant alloys made with magnesium would combine the advantages to the two principal spacer materials currently used, but without their limitations, at least in terms of MRI artifacting
  24 in total

Review 1.  The postoperative spine.

Authors:  C B Henk; W Brodner; S Grampp; M Breitenseher; M Thurnher; G H Mostbeck; H Imhof
Journal:  Top Magn Reson Imaging       Date:  1999-08

2.  [The posterior lumbar interbody fusion with cages (PLIF) and transpedicular stabilization].

Authors:  O Diedrich; C N Kraft; L Perlick; O Schmitt
Journal:  Zentralbl Neurochir       Date:  2001

Review 3.  Review article: MRI of the postoperative lumbar spine.

Authors:  J W M Van Goethem; P M Parizel; J R Jinkins
Journal:  Neuroradiology       Date:  2002-08-10       Impact factor: 2.804

4.  Postoperative magnetic resonance imaging with titanium implants of the thoracic and lumbar spine.

Authors:  O Ortiz; T G Pait; P McAllister; K Sauter
Journal:  Neurosurgery       Date:  1996-04       Impact factor: 4.654

5.  Single-stage treatment of pyogenic spinal infection with titanium mesh cages.

Authors:  Timothy R Kuklo; Benjamin K Potter; Randy S Bell; Ross R Moquin; Michael K Rosner
Journal:  J Spinal Disord Tech       Date:  2006-07

6.  Optimizing imaging parameters for MR evaluation of the spine with titanium pedicle screws.

Authors:  C A Petersilge; J S Lewin; J L Duerk; J U Yoo; A J Ghaneyem
Journal:  AJR Am J Roentgenol       Date:  1996-05       Impact factor: 3.959

7.  Metallic spinal artifacts in magnetic resonance imaging.

Authors:  A R Vaccaro; R M Chesnut; G Scuderi; J F Healy; J B Massie; S R Garfin
Journal:  Spine (Phila Pa 1976)       Date:  1994-06-01       Impact factor: 3.468

Review 8.  Magnesium and its alloys as orthopedic biomaterials: a review.

Authors:  Mark P Staiger; Alexis M Pietak; Jerawala Huadmai; George Dias
Journal:  Biomaterials       Date:  2005-10-24       Impact factor: 12.479

9.  Comparison of CT, MRI, and radiographs in assessing pelvic osteolysis: a cadaveric study.

Authors:  Tim A Walde; Daniel E Weiland; Serena B Leung; Nobuto Kitamura; Christi J Sychterz; C Anderson Engh; Alexandra M Claus; Hollis G Potter; Charles A Engh
Journal:  Clin Orthop Relat Res       Date:  2005-08       Impact factor: 4.176

10.  A carbon fiber implant to aid interbody lumbar fusion. Two-year clinical results in the first 26 patients.

Authors:  J W Brantigan; A D Steffee
Journal:  Spine (Phila Pa 1976)       Date:  1993-10-15       Impact factor: 3.468

View more
  9 in total

Review 1.  Zinc-based alloys for degradable vascular stent applications.

Authors:  Ehsan Mostaed; Malgorzata Sikora-Jasinska; Jaroslaw W Drelich; Maurizio Vedani
Journal:  Acta Biomater       Date:  2018-03-10       Impact factor: 8.947

2.  Metal-induced artifacts in computed tomography and magnetic resonance imaging: comparison of a biodegradable magnesium alloy versus titanium and stainless steel controls.

Authors:  Lukas Filli; Roger Luechinger; Thomas Frauenfelder; Stefan Beck; Roman Guggenberger; Nadja Farshad-Amacker; Gustav Andreisek
Journal:  Skeletal Radiol       Date:  2014-11-23       Impact factor: 2.199

3.  MR-compatible antibiotic interlocked nail fabrication for the management of long bone infections: first case report of a new technique.

Authors:  Cyril Mauffrey; George W Chaus; Nathan Butler; Heather Young
Journal:  Patient Saf Surg       Date:  2014-03-17

4.  Is the radiographic subsidence of stand-alone cages associated with adverse clinical outcomes after cervical spine fusion? An observational cohort study with 2-year follow-up outcome scoring.

Authors:  Dirk Zajonz; Anne-Catherine Franke; Nicolas von der Höh; Anna Voelker; Michael Moche; Jens Gulow; Christoph-Eckhard Heyde
Journal:  Patient Saf Surg       Date:  2014-11-07

Review 5.  Artifacts in magnetic resonance imaging.

Authors:  Katarzyna Krupa; Monika Bekiesińska-Figatowska
Journal:  Pol J Radiol       Date:  2015-02-23

6.  Biodegradable magnesium Herbert screw - image quality and artifacts with radiography, CT and MRI.

Authors:  Lena Sonnow; Sören Könneker; Peter M Vogt; Frank Wacker; Christian von Falck
Journal:  BMC Med Imaging       Date:  2017-02-14       Impact factor: 1.930

7.  Comparison of MRI Visualization Following Minimally Invasive and Open TLIF: A Retrospective Single-Center Study.

Authors:  Vadim A Byvaltsev; Andrei A Kalinin; Morgan B Giers; Valerii V Shepelev; Yurii Ya Pestryakov; Mikhail Yu Biryuchkov
Journal:  Diagnostics (Basel)       Date:  2021-05-19

8.  Interactions between magnetic resonance imaging and dental material.

Authors:  Chalakuzhiyl Abraham Mathew; Sudhakara Maller
Journal:  J Pharm Bioallied Sci       Date:  2013-06

9.  Design and development of novel MRI compatible zirconium- ruthenium alloys with ultralow magnetic susceptibility.

Authors:  H F Li; F Y Zhou; L Li; Y F Zheng
Journal:  Sci Rep       Date:  2016-04-19       Impact factor: 4.379

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.