Automated ultrasound scanning on a dual-modality breast imaging system: coverage and motion issues and solutions.

Sumedha P Sinha1, Mitchell M Goodsitt, Marilyn A Roubidoux, Rebecca C Booi, Gerald L LeCarpentier, Christine R Lashbrook, Kai E Thomenius, Carl L Chalek, Paul L Carson.   

Abstract

OBJECTIVE: We are developing an automated ultrasound imaging-mammography system wherein a digital mammography unit has been augmented with a motorized ultrasound transducer carriage above a special compression paddle. Challenges of this system are acquiring complete coverage of the breast and minimizing motion. We assessed these problems and investigated methods to increase coverage and stabilize the compressed breast.
METHODS: Visual tracings of the breast-to-paddle contact area and breast periphery were made for 10 patients to estimate coverage area. Various motion artifacts were evaluated in 6 patients. Nine materials were tested for coupling the paddle to the breast. Fourteen substances were tested for coupling the transducer to the paddle in lateral-to-medial and medial-to-lateral views and filling the gap between the peripheral breast and paddle. In-house image registration software was used to register adjacent ultrasound sweeps.
RESULTS: The average breast contact area was 56%. The average percentage of the peripheral air gap filled with ultrasound gel was 61%. Shallow patient breathing proved equivalent to breath holding, whereas speech and sudden breathing caused unacceptable artifacts. An adhesive spray that preserves image quality was found to be best for coupling the breast to the paddle and minimizing motion. A highly viscous ultrasound gel proved most effective for coupling the transducer to the paddle for lateral-to-medial and medial-to-lateral views and for edge fill-in.
CONCLUSIONS: The challenges of automated ultrasound scanning in a multimodality breast imaging system have been addressed by developing methods to fill in peripheral gaps, minimize patient motion, and register and reconstruct multisweep ultrasound image volumes.

Entities:  

Mesh:

Year:  2007        PMID: 17460006     DOI: 10.7863/jum.2007.26.5.645

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  12 in total

1.  Digital Breast Tomosynthesis: State of the Art.

Authors:  Srinivasan Vedantham; Andrew Karellas; Gopal R Vijayaraghavan; Daniel B Kopans
Journal:  Radiology       Date:  2015-12       Impact factor: 11.105

2.  Full-field acoustomammography using an acousto-optic sensor.

Authors:  J S Sandhu; R A Schmidt; P J La Rivière
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

Review 3.  Digital mammography imaging: breast tomosynthesis and advanced applications.

Authors:  Mark A Helvie
Journal:  Radiol Clin North Am       Date:  2010-09       Impact factor: 2.303

4.  Effect of a gel retainment dam on automated ultrasound coverage in a dual-modality breast imaging system.

Authors:  Jie Li; Mitchell M Goodsitt; Frederic Padilla; J Brian Fowlkes; Fong Ming Hooi; Christine R Lashbrook; Kai E Thomenius; Paul L Carson
Journal:  J Ultrasound Med       Date:  2010-07       Impact factor: 2.153

Review 5.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

6.  Spatial registration of temporally separated whole breast 3D ultrasound images.

Authors:  Ganesh Narayanasamy; Gerald L LeCarpentier; Marilyn Roubidoux; J Brian Fowlkes; Anne F Schott; Paul L Carson
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

7.  Automated Breast Ultrasound: Dual-Sided Compared with Single-Sided Imaging.

Authors:  Eric D Larson; Won-Mean Lee; Marilyn A Roubidoux; Mitchel M Goodsitt; Chris Lashbrook; Fouzaan Zafar; Oliver D Kripfgans; Kai Thomenius; Paul L Carson
Journal:  Ultrasound Med Biol       Date:  2016-06-03       Impact factor: 2.998

8.  Characterization of cysts using differential correlation coefficient values from two dimensional breast elastography: preliminary study.

Authors:  Rebecca C Booi; Paul L Carson; Matthew O'Donnell; Marilyn A Roubidoux; Anne L Hall; Jonathan M Rubin
Journal:  Ultrasound Med Biol       Date:  2007-09-27       Impact factor: 2.998

9.  Breast mass characterization using 3-dimensional automated ultrasound as an adjunct to digital breast tomosynthesis: a pilot study.

Authors:  Frederic Padilla; Marilyn A Roubidoux; Chintana Paramagul; Sumedha P Sinha; Mitchell M Goodsitt; Gerald L Le Carpentier; Heang-Ping Chan; Lubomir M Hadjiiski; J Brian Fowlkes; Annette D Joe; Katherine A Klein; Alexis V Nees; Mitra Noroozian; Stephanie K Patterson; Renee W Pinsky; Fong Ming Hooi; Paul L Carson
Journal:  J Ultrasound Med       Date:  2013-01       Impact factor: 2.153

10.  Image registration for detection and quantification of change on digital tomosynthesis mammographic volumes.

Authors:  Sumedha P Sinha; Ramkrishnan Narayanan; Bing Ma; Marilyn A Roubidoux; He Liu; Paul L Carson
Journal:  AJR Am J Roentgenol       Date:  2009-02       Impact factor: 3.959

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.