Literature DB >> 9113539

Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris.

P A Devane1, J G Horne, K Martin, G Coldham, B Krause.   

Abstract

Between 1985 and 1990, 108 consecutive Harris-Galante I (Zimmer, Warsaw, IN) total hip arthroplasties were performed by four surgeons at a single hospital. At the time of follow-up evaluation, 80 hips were available for review. The mean rate of linear wear was 0.15 mm/y, mean rate of three-dimensional femoral head displacement was 0.21 mm/y, and mean rate of volumetric wear was 121 mm3/y. Eight hips (10%) in this series had radiologic osteolysis around either the femoral or acetabular component. A significantly greater volumetric wear rate was found in patients who were younger, those with a higher activity level, those who received a 32-mm-diameter femoral head, and those with vertical orientation of their acetabular component. No relationship could be made with patient weight, gender, Harris hip score, or cup diameter.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9113539     DOI: 10.1016/s0883-5403(97)90021-8

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  50 in total

1.  Wear analysis of chamfered elongated acetabular cup liners.

Authors:  Hsiao-Che Lin; Tzuo-Liang Luo; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2011-12-22       Impact factor: 2.602

2.  Dual mobility cup: dislocation rate and survivorship at ten years of follow-up.

Authors:  Jean-Louis Prudhon; André Ferreira; Régis Verdier
Journal:  Int Orthop       Date:  2013-09-13       Impact factor: 3.075

3.  Does impact sport activity influence total hip arthroplasty durability?

Authors:  Matthieu Ollivier; Solenne Frey; Sebastien Parratte; Xavier Flecher; Jean-Noël Argenson
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

4.  Correlation between groin pain and cup design of hip-resurfacing implants: a prospective study.

Authors:  Julien Girard; Erwan Pansard; Reda Ouahes; Henri Migaud; Cyril Delay; Laurent Vasseur
Journal:  Int Orthop       Date:  2013-12-10       Impact factor: 3.075

5.  Cementless Total Hip Arthroplasty With Metasul Bearings Provides Good Results in Active Young Patients: A Concise Followup.

Authors:  Christian P Delaunay; Sophie Putman; Benjamin Puliéro; Matthieu Bégin; Henri Migaud; François Bonnomet
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

6.  Fifteen-year survival of the Cedior™ total knee prosthesis.

Authors:  Roger Erivan; Edouard Fadlallah; Guillaume Villatte; Aurélien Mulliez; Stéphane Descamps; Stéphane Boisgard
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-07-06

7.  Cementless total knee arthroplasty in chronic inflammatory rheumatism.

Authors:  Jonathan Buchheit; Antoine Serre; Xavier Bouilloux; Marc Puyraveau; Laurent Jeunet; Patrick Garbuio
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-09-17

8.  Polyethylene in total hip arthroplasty: half a century in the limelight.

Authors:  Eduardo García-Rey; Eduardo García-Cimbrelo
Journal:  J Orthop Traumatol       Date:  2010-05-27

9.  A stature-specific concept for uncemented, primary total hip arthroplasty.

Authors:  Georg W Omlor; Hannah Ullrich; Knut Krahmer; Alexander Jung; Günther Aldinger; Peter Aldinger
Journal:  Acta Orthop       Date:  2010-02       Impact factor: 3.717

10.  Custom stems for femoral deformity in patients less than 40 years of age: 70 hips followed for an average of 14 years.

Authors:  Michael Akbar; Guenther Aldinger; Knut Krahmer; Thomas Bruckner; Peter R Aldinger
Journal:  Acta Orthop       Date:  2009-08       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.