Literature DB >> 17226087

Does the rate of inappropriate therapy differ in implantable cardioverter-defibrillators from different manufacturers?

Arash Arya1, Majid Haghjoo, Zahra Emkanjoo, Mohammad Reza Dehghani, Mohammad Ali Sadr-Ameli.   

Abstract

OBJECTIVE: We conducted this study to compare the rate of >/=1 inappropriate therapy between ICDs from two manufacturers which use different discriminatory protocols.
METHOD: One hundred sixty two patients (mean age 58 +/- 13 years, 126 male) who received ICDs between January 2001 and 2005 were included in the study. Clinical, electrocardiographic, and ICD stored data and electrograms were collected and analyzed. Immediately after implantation all the detection and discrimination criteria were activated with the nominal values in order to compare the two discriminatory protocols under the default manufacturer's settings.
RESULTS: During the follow up period of 14.3 +/- 10 months, 49 (30%) patients received >/=1 inappropriate ICD therapy. The rate of >/=1 inappropriate ICD therapy in manufacturer A and B ICDs was 26% (n = 29) and 41% (n = 20), respectively. Comparing the rate of >/=1 inappropriate ICD therapy between the two groups by Kaplan-Meier analysis and the log rank test resulted in P = 0.04.
CONCLUSION: Having all discriminatory variables activated with the nominal values, discriminatory performance differs between the two manufacturers. Further larger-scale studies are warranted to prospectively compare the performance of various available ICDs' discriminatory protocols, and define the optimum combination of discriminators in each ICD to decrease the rate of inappropriate therapy.

Entities:  

Mesh:

Year:  2007        PMID: 17226087     DOI: 10.1007/s10840-006-9046-0

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  14 in total

1.  Clinical variables predicting inappropriate use of implantable cardioverter-defibrillator in patients with coronary heart disease or nonischemic dilated cardiomyopathy.

Authors:  Dominic A M J Theuns; A Peter J Klootwijk; Maarten L Simoons; Luc J Jordaens
Journal:  Am J Cardiol       Date:  2005-01-15       Impact factor: 2.778

2.  Do current dual chamber cardioverter defibrillators have advantages over conventional single chamber cardioverter defibrillators in reducing inappropriate therapies? A randomized, prospective study.

Authors:  I Deisenhofer; C Kolb; G Ndrepepa; J Schreieck; M Karch; S Schmieder; B Zrenner; C Schmitt
Journal:  J Cardiovasc Electrophysiol       Date:  2001-02

3.  The Rhythm ID Going Head to Head Trial (RIGHT): design of a randomized trial comparing competitive rhythm discrimination algorithms in implantable cardioverter defibrillators.

Authors:  Ronald D Berger; Darin R Lerew; Joseph M Smith; Chris Pulling; Michael R Gold
Journal:  J Cardiovasc Electrophysiol       Date:  2006-07

4.  Discrimination of ventricular tachycardia from supraventricular tachycardia by a downloaded wavelet-transform morphology algorithm: a paradigm for development of implantable cardioverter defibrillator detection algorithms.

Authors:  Charles D Swerdlow; Mark L Brown; Keith Lurie; Jianxin Zhang; Nicole M Wood; Walter H Olson; Jeffrey M Gillberg
Journal:  J Cardiovasc Electrophysiol       Date:  2002-05

5.  Importance of the atrial channel for ventricular arrhythmia therapy in the dual chamber implantable cardioverter defibrillator.

Authors:  B Dijkman; H J Wellens
Journal:  J Cardiovasc Electrophysiol       Date:  2000-12

6.  Optimal combination of discriminators for differentiating ventricular from supraventricular tachycardia by dual-chamber defibrillators.

Authors:  Michael Glikson; Charles D Swerdlow; Osnat T Gurevitz; Emile Daoud; Kalyanam Shivkumar; Bruce Wilkoff; Tamara Shipman; Paul A Friedman
Journal:  J Cardiovasc Electrophysiol       Date:  2005-07

Review 7.  Supraventricular tachycardia-ventricular tachycardia discrimination algorithms in implantable cardioverter defibrillators: state-of-the-art review.

Authors:  C D Swerdlow
Journal:  J Cardiovasc Electrophysiol       Date:  2001-05

8.  Clinical experience with a new detection algorithm for differentiation of supraventricular from ventricular tachycardia in a dual-chamber defibrillator.

Authors:  Anil-Martin Sinha; Christoph Stellbrink; Andreas Schuchert; Bernhard Mox; Luc Jordaens; Dominique Lamaison; Jaswinder Gill; Andrew Kaplan; Bela Merkely
Journal:  J Cardiovasc Electrophysiol       Date:  2004-06

9.  Is inappropriate implantable defibrillator shock therapy predictable?

Authors:  Kumaraswamy Nanthakumar; Paul Dorian; Miney Paquette; Mary Greene; Janet Edwards; Denis Heng; James Noble; David Newman
Journal:  J Interv Card Electrophysiol       Date:  2003-06       Impact factor: 1.900

10.  Enhanced specificity of a dual chamber ICD arrhythmia detection algorithm by rate stability criteria.

Authors:  Ralph Mletzko; Frederic Anselme; Didier Klug; Wolfgang Schoels; Robert Bowes; Nicolas Iscolo; Rémi Nitzsché; Nicolas Sadoul
Journal:  Pacing Clin Electrophysiol       Date:  2004-08       Impact factor: 1.976

View more