Clinical experience with a new detection algorithm for differentiation of supraventricular from ventricular tachycardia in a dual-chamber defibrillator.

Anil-Martin Sinha1, Christoph Stellbrink, Andreas Schuchert, Bernhard Mox, Luc Jordaens, Dominique Lamaison, Jaswinder Gill, Andrew Kaplan, Bela Merkely.   

Abstract

INTRODUCTION: Inadequate therapy for supraventricular tachyarrhythmias (SVT) is a frequent problem of implantable cardioverter defibrillators (ICD). Dual-chamber ICDs have been developed to improve discrimination of SVT from ventricular tachycardia (VT). We investigated the positive predictivity, sensitivity, and specificity of a new algorithm, the SMART detection trade mark algorithm, incorporated in the Phylax AV (Biotronik) dual-chamber ICD. METHODS AND
RESULTS: Two hundred nine patients (185 men, age 64 +/- 11 years) received a Phylax AV ICD with SMART detection trade mark activated. In 138 of these patients, 1,245 sustained tachycardia episodes with a detailed electrogram were stored in the device during a follow-up period of 10 +/- 6 months. Episodes were correctly classified as ventricular fibrillation (VF, n = 178) in 52 patients, VT (n = 641) in 98 patients, and SVT (n = 385) in 48 patients by the algorithm. Forty-one true SVT episodes (3.3%) were misclassified as VT: atrial fibrillation (n = 7) and flutter (n = 1), sinus tachycardia (n = 12), and other SVT (n = 21). The positive predictivity for VF/VT was 94.5% (95% CI 92.7-95.8) uncorrected and 94.5% (95% CI 92.9-95.8%) corrected with the generalized equation estimation (GEE) method. The positive predictivity for SVT was 100%. The specificity was 88.9% (95% CI 85.6-91.6%) uncorrected and 89.0% (95% CI 85.6-91.6%) corrected with the GEE method with a sensitivity of 100%.
CONCLUSION: The SMART detection trade mark algorithm was safe and reliable for the detection of all ventricular tachycardias. Although its specificity was high, it should be improved with regard to SVT to avoid inappropriate ICD therapies.

Entities:  

Mesh:

Year:  2004        PMID: 15175058     DOI: 10.1046/j.1540-8167.2004.03290.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  11 in total

1.  Does the rate of inappropriate therapy differ in implantable cardioverter-defibrillators from different manufacturers?

Authors:  Arash Arya; Majid Haghjoo; Zahra Emkanjoo; Mohammad Reza Dehghani; Mohammad Ali Sadr-Ameli
Journal:  J Interv Card Electrophysiol       Date:  2007-01-17       Impact factor: 1.900

Review 2.  Athletes with Implantable Cardioverter Defibrillators.

Authors:  Shiva P Ponamgi; Christopher V DeSimone; Michael J Ackerman
Journal:  Clin Sports Med       Date:  2015-07       Impact factor: 2.182

Review 3.  "Two for the Price of One": A Single-Lead Implantable Cardioverter-Defibrillator System with a Floating Atrial Dipole.

Authors:  Nicole E Worden; Musab Alqasrawi; Siva M Krothapalli; Alexander Mazur
Journal:  J Atr Fibrillation       Date:  2016-04-30

4.  Reduction of the inappropriate ICD therapies by implementing a new fuzzy logic-based diagnostic algorithm.

Authors:  Michał Lewandowski; Andrzej Przybylski; Wiesław Kuźmicz; Hanna Szwed
Journal:  Ann Noninvasive Electrocardiol       Date:  2013-09       Impact factor: 1.468

5.  Sleep-disordered breathing and inappropriate defibrillator shocks in chronic heart failure.

Authors:  Thomas Bitter; Klaus-Jürgen Gutleben; Georg Nölker; Zisis Dimitriadis; Christian Prinz; Jürgen Vogt; Dieter Horstkotte; Olaf Oldenburg
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2014-07-29

Review 6.  ICD discrimination of SVT versus VT with 1:1 V-A conduction: A review of the literature.

Authors:  Rhanderson N Cardoso; Chris Healy; Juan Viles-Gonzalez; James O Coffey
Journal:  Indian Pacing Electrophysiol J       Date:  2016-02-12

Review 7.  Tachycardia detection in modern implantable cardioverter-defibrillators.

Authors:  Thomas Brüggemann; Daniel Dahlke; Amin Chebbo; Ilka Neumann
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-08-30

8.  The efficacy of the LinoxSmart DX ICD lead from a single center experience.

Authors:  Ibrahim Marai; Anat Milman; Rafael Diamante; Osnat Gurevitz; David Barlev; Igor Lipchenka; Eyal Nof; Michael Glikson; Roy Beinart
Journal:  Indian Pacing Electrophysiol J       Date:  2019-12-17

9.  Cost Saving Potential of an Early Detection of Atrial Fibrillation in Patients after ICD Implantation.

Authors:  Thomas Reinhold; Roberto Belke; Tino Hauser; Christian Grebmer; Carsten Lennerz; Verena Semmler; Christof Kolb
Journal:  Biomed Res Int       Date:  2018-08-14       Impact factor: 3.411

10.  Novel two-lead cardiac resynchronization therapy system provides equivalent CRT responses with less complications than a conventional three-lead system: Results from the QP ExCELs lead registry.

Authors:  Naushad A Shaik; Michael Drucker; Christopher Pierce; Gabor Z Duray; Shane Gillett; Crystal Miller; Camden Harrell; George Thomas
Journal:  J Cardiovasc Electrophysiol       Date:  2020-06-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.