Target-specific optimization of four-dimensional cone beam computed tomography.

Moiz Ahmad1, Tinsu Pan.   

Abstract

PURPOSE: Under-sampling artifacts are a major problem in four-dimensional cone-beam computed tomography (4D-CBCT) and may compromise evaluation of target motion. Several studies have addressed scan parameter selection for minimizing under-sampling artifacts; however, the role of the target characteristics in scan parameter selection has not been investigated. In this work, the authors evaluated 4D-CBCT performance by assessing the accuracy of target motion measurements for various target sizes and motions. The results may serve as patient-specific guidelines for scan parameters selection in 4D-CBCT.
METHODS: The authors acquired 4D-CBCT scans of a moving phantom consisting of six water-filled sphere targets of sizes 10-37 mm, using various scan times ranging from 30 s to 3 min., setting the motion to 3-s and 6-s periods. The authors used automatic image registration to extract the target motion trajectories and evaluated these measurements for various target sizes and motions over various combinations of scan parameters including scan time, detector configuration, number of respiration phases, and reconstruction filters.
RESULTS: The most important object parameter to 4D-CBCT performance was the period of motion. Measurements for the 6-s motion were always systematically less accurate than measurements for the 3-s motion for 34 of 36 objects of various sizes and periods of motion. The 6-s motion required a greater scan time than the 3-s motion for equivalent measurement accuracy. The second most influential parameter was the target size. For the 3-s period of motion, objects larger than 13 mm were tracked with sub-millimeter accuracy with a 1-min scan time. For the 6-s period of motion, objects larger than 22-mm were tracked with sub-millimeter accuracy with a 1.5-min scan time. For all sizes and motions, temporal blurring was observed when the number of phases was fewer than 8. Offset detector configuration provided the same performance as centered detector except for small targets (20 mm.) at short scan times (≤1 min). Finally, reconstruction filtering and number of respiratory phases did not affect performance.
CONCLUSIONS: Scan time should be set according to target size and motion. The authors have provided figures that provide the minimum scan time needed to achieve the particular motion measurement accuracy for a particular size and motion period.

Entities:  

Mesh:

Year:  2012        PMID: 22957634      PMCID: PMC4032412          DOI: 10.1118/1.4747609

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  27 in total

1.  X-ray micro-CT with a displaced detector array.

Authors:  Ge Wang
Journal:  Med Phys       Date:  2002-07       Impact factor: 4.071

2.  Extraction of tumor motion trajectories using PICCS-4DCBCT: a validation study.

Authors:  Zhihua Qi; Guang-Hong Chen
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

3.  An investigation of 4D cone-beam CT algorithms for slowly rotating scanners.

Authors:  Frank Bergner; Timo Berkus; Markus Oelhafen; Patrik Kunz; Tinsu Pa; Rainer Grimmer; Ludwig Ritschl; Marc Kachelriess
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

4.  Compressed sensing based cone-beam computed tomography reconstruction with a first-order method.

Authors:  Kihwan Choi; Jing Wang; Lei Zhu; Tae-Suk Suh; Stephen Boyd; Lei Xing
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

5.  Autoadaptive phase-correlated (AAPC) reconstruction for 4D CBCT.

Authors:  Frank Bergner; Timo Berkus; Markus Oelhafen; Patrik Kunz; Tinsu Pan; Marc Kachelriess
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

6.  Performance studies of four-dimensional cone beam computed tomography.

Authors:  Zhihua Qi; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2011-09-30       Impact factor: 3.609

7.  Evaluation of sparse-view reconstruction from flat-panel-detector cone-beam CT.

Authors:  Junguo Bian; Jeffrey H Siewerdsen; Xiao Han; Emil Y Sidky; Jerry L Prince; Charles A Pelizzari; Xiaochuan Pan
Journal:  Phys Med Biol       Date:  2010-10-20       Impact factor: 3.609

8.  Correction of motion artifacts in cone-beam CT using a patient-specific respiratory motion model.

Authors:  Qinghui Zhang; Yu-Chi Hu; Fenghong Liu; Karyn Goodman; Kenneth E Rosenzweig; Gig S Mageras
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

9.  High temporal resolution and streak-free four-dimensional cone-beam computed tomography.

Authors:  Shuai Leng; Jie Tang; Joseph Zambelli; Brian Nett; Ranjini Tolakanahalli; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2008-09-24       Impact factor: 3.609

10.  Streaking artifacts reduction in four-dimensional cone-beam computed tomography.

Authors:  Shuai Leng; Joseph Zambelli; Ranjini Tolakanahalli; Brian Nett; Peter Munro; Joshua Star-Lack; Bhudatt Paliwal; Guang-Hong Chen
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

View more
  5 in total

1.  A hybrid reconstruction algorithm for fast and accurate 4D cone-beam CT imaging.

Authors:  Hao Yan; Xin Zhen; Michael Folkerts; Yongbao Li; Tinsu Pan; Laura Cervino; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

2.  Optimization of acquisition parameters and accuracy of target motion trajectory for four-dimensional cone-beam computed tomography with a dynamic thorax phantom.

Authors:  Yoshinobu Shimohigashi; Fujio Araki; Masato Maruyama; Yuji Nakaguchi; Kengo Nakato; Nozomu Nagasue; Yudai Kai
Journal:  Radiol Phys Technol       Date:  2014-10-07

3.  SPARE: Sparse-view reconstruction challenge for 4D cone-beam CT from a 1-min scan.

Authors:  Chun-Chien Shieh; Yesenia Gonzalez; Bin Li; Xun Jia; Simon Rit; Cyril Mory; Matthew Riblett; Geoffrey Hugo; Yawei Zhang; Zhuoran Jiang; Xiaoning Liu; Lei Ren; Paul Keall
Journal:  Med Phys       Date:  2019-07-19       Impact factor: 4.071

4.  Quantitative evaluation of 4D Cone beam CT scans with reduced scan time in lung cancer patients.

Authors:  Abigail Bryce-Atkinson; Thomas Marchant; John Rodgers; Geoff Budgell; Alan McWilliam; Corinne Faivre-Finn; Gillian Whitfield; Marcel van Herk
Journal:  Radiother Oncol       Date:  2019-04-11       Impact factor: 6.280

5.  Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study.

Authors:  Jihun Kim; Ki Chang Keum; Ho Lee; Chae-Seon Hong; Kwangwoo Park; Jin Sung Kim
Journal:  Radiat Oncol       Date:  2020-09-25       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.