Literature DB >> 17191225

Augmented generalized SENSE reconstruction to correct for rigid body motion.

Roland Bammer1, Murat Aksoy, Chunlei Liu.   

Abstract

The correction of motion artifacts continues to be a significant problem in MRI. In the case of uncooperative patients, such as children, or patients who are unable to remain stationary, the accurate determination and correction of motion artifacts becomes a very important prerequisite for achieving good image quality. The application of conventional motion-correction strategies often produces inconsistencies in k-space data. As a result, significant residual artifacts can persist. In this work a formalism is introduced for parallel imaging in the presence of motion. The proposed method can improve overall image quality because it diminishes k-space inconsistencies by exploiting the complementary image encoding capacity of individual receiver coils. Specifically, an augmented version of an iterative SENSE reconstruction is used as a means of synthesizing the missing data in k-space. Motion is determined from low-resolution navigator images that are coregistered by an automatic registration routine. Navigator data can be derived from self-navigating k-space trajectories or in combination with other navigation schemes that estimate patient motion. This correction method is demonstrated by interleaved spiral images collected from volunteers. Conventional spiral scans and scans corrected with proposed techniques are shown, and the results illustrate the capacity of this new correction approach.

Entities:  

Mesh:

Year:  2007        PMID: 17191225      PMCID: PMC3985846          DOI: 10.1002/mrm.21106

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  32 in total

1.  Resampling of data between arbitrary grids using convolution interpolation.

Authors:  V Rasche; R Proksa; R Sinkus; P Börnert; H Eggers
Journal:  IEEE Trans Med Imaging       Date:  1999-05       Impact factor: 10.048

2.  Detection and elimination of motion artifacts by regeneration of k-space.

Authors:  M Bydder; D J Larkman; J V Hajnal
Journal:  Magn Reson Med       Date:  2002-04       Impact factor: 4.668

3.  SMASH navigators.

Authors:  M Bydder; D Atkinson; D J Larkman; D L G Hill; J V Hajnal
Journal:  Magn Reson Med       Date:  2003-03       Impact factor: 4.668

4.  Generalized autocalibrating partially parallel acquisitions (GRAPPA).

Authors:  Mark A Griswold; Peter M Jakob; Robin M Heidemann; Mathias Nittka; Vladimir Jellus; Jianmin Wang; Berthold Kiefer; Axel Haase
Journal:  Magn Reson Med       Date:  2002-06       Impact factor: 4.668

5.  Simultaneous phase correction and SENSE reconstruction for navigated multi-shot DWI with non-cartesian k-space sampling.

Authors:  Chunlei Liu; Michael E Moseley; Roland Bammer
Journal:  Magn Reson Med       Date:  2005-12       Impact factor: 4.668

6.  Selection of a convolution function for Fourier inversion using gridding [computerised tomography application].

Authors:  J I Jackson; C H Meyer; D G Nishimura; A Macovski
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

7.  Multi-modal volume registration by maximization of mutual information.

Authors:  W M Wells; P Viola; H Atsumi; S Nakajima; R Kikinis
Journal:  Med Image Anal       Date:  1996-03       Impact factor: 8.545

8.  Self-navigated spiral fMRI: interleaved versus single-shot.

Authors:  G H Glover; S Lai
Journal:  Magn Reson Med       Date:  1998-03       Impact factor: 4.668

9.  Matrix description of general motion correction applied to multishot images.

Authors:  P G Batchelor; D Atkinson; P Irarrazaval; D L G Hill; J Hajnal; D Larkman
Journal:  Magn Reson Med       Date:  2005-11       Impact factor: 4.668

Review 10.  Artifact reduction using parallel imaging methods.

Authors:  David J Larkman; David Atkinson; Jo V Hajnal
Journal:  Top Magn Reson Imaging       Date:  2004-08
View more
  36 in total

1.  Correction of B 0-induced geometric distortion variations in prospective motion correction for 7T MRI.

Authors:  Uten Yarach; Chaiya Luengviriya; Daniel Stucht; Frank Godenschweger; Peter Schulze; Oliver Speck
Journal:  MAGMA       Date:  2016-02-09       Impact factor: 2.310

Review 2.  Advances in magnetic resonance neuroimaging.

Authors:  Michael E Moseley; Chunlei Liu; Sandra Rodriguez; Thomas Brosnan
Journal:  Neurol Clin       Date:  2009-02       Impact factor: 3.806

3.  On the significance of motion degradation in high-resolution 3D μMRI of trabecular bone.

Authors:  Yusuf A Bhagat; Chamith S Rajapakse; Jeremy F Magland; Michael J Wald; Hee Kwon Song; Mary B Leonard; Felix W Wehrli
Journal:  Acad Radiol       Date:  2011-08-04       Impact factor: 3.173

4.  Homogeneous coordinates in motion correction.

Authors:  Benjamin Zahneisen; Thomas Ernst
Journal:  Magn Reson Med       Date:  2015-02-03       Impact factor: 4.668

5.  Motion correction for diffusion weighted SMS imaging.

Authors:  M Herbst; B A Poser; A Singh; W Deng; B Knowles; M Zaitsev; V A Stenger; T Ernst
Journal:  Magn Reson Imaging       Date:  2016-12-15       Impact factor: 2.546

6.  Prospective cardiac motion self-gating.

Authors:  Fei Han; Stanislas Rapacchi; Peng Hu
Journal:  Quant Imaging Med Surg       Date:  2017-04

Review 7.  Motion artifacts in MRI: A complex problem with many partial solutions.

Authors:  Maxim Zaitsev; Julian Maclaren; Michael Herbst
Journal:  J Magn Reson Imaging       Date:  2015-01-28       Impact factor: 4.813

8.  Self-encoded marker for optical prospective head motion correction in MRI.

Authors:  Christoph Forman; Murat Aksoy; Joachim Hornegger; Roland Bammer
Journal:  Med Image Anal       Date:  2011-06-13       Impact factor: 8.545

9.  Model for the correction of motion-induced phase errors in multishot diffusion-weighted-MRI of the head: are cardiac-motion-induced phase errors reproducible from beat-to-beat?

Authors:  R L O'Halloran; S Holdsworth; M Aksoy; R Bammer
Journal:  Magn Reson Med       Date:  2011-12-28       Impact factor: 4.668

10.  Motion immune diffusion imaging using augmented MUSE for high-resolution multi-shot EPI.

Authors:  Shayan Guhaniyogi; Mei-Lan Chu; Hing-Chiu Chang; Allen W Song; Nan-Kuei Chen
Journal:  Magn Reson Med       Date:  2015-03-11       Impact factor: 4.668

View more