16-MDCT angiography in living kidney donors at various tube potentials: impact on image quality and radiation dose.

Dushyant V Sahani1, Sanjeeva P Kalva, Peter F Hahn, Sanjay Saini.   

Abstract

OBJECTIVE: The objective of our study was to compare the performance of 16-MDCT angiography at various peak kilovoltage (kVp) settings and the impact of the different settings on image quality and on radiation dose in adult kidney donors.
MATERIALS AND METHODS: Sixty-two renal donors (32 men, 30 women) who underwent 16-MDCT were divided into three groups: 18 subjects were studied at 140 kVp (group A); 20, at 120 kVp (group B); and 24, at 100 kVp (group C). Other constant scanning parameters were as follows: detector collimation, 0.625 mm; table feed, 9.375 mm/rotation; gantry rotation time, 500 milliseconds; and automatic current tube modulation (ATCM) using a noise index of 15. A total of 135-140 mL of iodinated contrast material (300 mg I/mL) was administered at 5 mL/s via an 18-gauge cannula, and arterial phase scanning was initiated using a bolus-tracking technique. Two observers evaluated image quality of the axial and 3D images and the visibility of branch order in the superior mesenteric artery (SMA) and renal arteries. Attenuation (in Hounsfield units [H]) in the aorta, SMA, and main renal artery was also measured by placing a region of interest. Radiation dose measurements were based on the scanner-generated CT dose index volume (CTDI(vol)). Each parameter tested was compared among the three groups using a nonparametric analysis of variance test, and a p value of 0.05 was considered significant.
RESULTS: Differences in the quality of the axial images existed between groups A and C (p < 0.001) and between groups B and C (p < 0.01); the image quality of the 3D images and the visibility of branch order in the SMA and renal arteries were comparable for all groups. The difference in mean attenuation of the aorta, SMA, and renal arteries was significant between groups A and C (p < 0.001) and between groups B and C (p < 0.01). All groups had 100% diagnostic accuracy in identifying the number of renal arteries on the side of nephrectomy. The mean radiation dose in CTDI(vol) was 25 +/- 3 mGy at 140 kVp, 17 +/- 4 mGy at 120 kVp, and 12 +/- 3 mGy at 100 kVp (p < 0.001).
CONCLUSION: Our initial observations suggest that the image quality of 16-MDCT angiography performed at 120 kVp is similar to that of CT angiography (CTA) performed at 140 kVp in adult kidney donors but with a significant radiation dose reduction. CTA at 100 kVp results in higher image noise but provides diagnostically acceptable images with significant radiation dose reduction compared with CTA at 120 or 140 kVp.

Entities:  

Mesh:

Year:  2007        PMID: 17179353     DOI: 10.2214/ajr.05.0583

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  17 in total

1.  Current status of low dose multi-detector CT in the urinary tract.

Authors:  Mi Kim Sung; Sarabjeet Singh; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2011-11-28

2.  Dose levels at coronary CT angiography--a comparison of Dual Energy-, Dual Source- and 16-slice CT.

Authors:  J Matthias Kerl; Ralf W Bauer; Tobias B Maurer; Rene Aschenbach; Huedayi Korkusuz; Thomas Lehnert; Simon Deseive; Hanns Ackermann; Thomas J Vogl
Journal:  Eur Radiol       Date:  2010-09-23       Impact factor: 5.315

3.  Feasibility of low-concentration iodinated contrast medium with lower-tube-voltage dual-source CT aortography using iterative reconstruction: comparison with automatic exposure control CT aortography.

Authors:  Hee Jeong Shin; Song Soo Kim; Jae-Hwan Lee; Jae-Hyeong Park; Jin-Ok Jeong; Seon Ah Jin; Byung Seok Shin; Kyung-Sook Shin; Moonsang Ahn
Journal:  Int J Cardiovasc Imaging       Date:  2015-11-30       Impact factor: 2.357

4.  Cerebral computed tomography angiography using a 70 kVp protocol: improved vascular enhancement with a reduced volume of contrast medium and radiation dose.

Authors:  Eun-Suk Cho; Tae-Sub Chung; Sung Jun Ahn; KyoungHoon Chong; Jang Hun Baek; Sang Hyun Suh
Journal:  Eur Radiol       Date:  2014-12-16       Impact factor: 5.315

5.  A prospective evaluation of contrast and radiation dose and image quality in cardiac CT in children with complex congenital heart disease using low-concentration iodinated contrast agent and low tube voltage and current.

Authors:  Qiao-Ru Hou; Wei Gao; Ai-Min Sun; Qian Wang; Hai-Sheng Qiu; Fang Wang; Li-Wei Hu; Jian-Ying Li; Yu-Min Zhong
Journal:  Br J Radiol       Date:  2016-12-07       Impact factor: 3.039

6.  "Sweet spot" for endoleak detection: optimizing contrast to noise using low keV reconstructions from fast-switch kVp dual-energy CT.

Authors:  Katherine E Maturen; Ravi K Kaza; Peter S Liu; Leslie E Quint; Shokoufeh H Khalatbari; Joel F Platt
Journal:  J Comput Assist Tomogr       Date:  2012 Jan-Feb       Impact factor: 1.826

7.  Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose.

Authors:  Matthias S May; Manuel R Kramer; Achim Eller; Wolfgang Wuest; Michael Scharf; Michael Brand; Marc Saake; Bernhard Schmidt; Michael Uder; Michael M Lell
Journal:  Neuroradiology       Date:  2014-06-25       Impact factor: 2.804

8.  Non-contrast MR angiography using three-dimensional balanced steady-state free-precession imaging for evaluation of stenosis in the celiac trunk and superior mesenteric artery: a preliminary comparative study with computed tomography angiography.

Authors:  Patricia P Cardia; Thiago J Penachim; Adilson Prando; Ulysses S Torres; Giuseppe D'Ippólito
Journal:  Br J Radiol       Date:  2017-06-07       Impact factor: 3.039

9.  Dual-energy-CT of hypervascular liver lesions in patients with HCC: investigation of image quality and sensitivity.

Authors:  Jens Altenbernd; Till A Heusner; Adrian Ringelstein; Susanne C Ladd; Michael Forsting; Gerald Antoch
Journal:  Eur Radiol       Date:  2010-10-10       Impact factor: 5.315

10.  Pediatric renovascular hypertension in Thailand: CT angiographic findings.

Authors:  Pannee Visrutaratna; Tanop Srisuwan; Chusak Sirivanichai
Journal:  Pediatr Radiol       Date:  2009-08-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.