Literature DB >> 17159973

Comparison of the ICare rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists.

L M Abraham1, N C R Epasinghe, D Selva, R Casson.   

Abstract

PURPOSE: To assess the agreement between ICare rebound tonometer and Goldmann applanation tonometer in the hands of experienced and inexperienced tonometrists. PATIENTS AND METHODS: Two tonometrists, experienced with both Goldmann applanation tonometry (GAT) and ICare Tonometry (ICT) measured intraocular pressure (IOP), in a masked fashion, in 100 patients. In another series of 58 patients, ICT was performed by an inexperienced tonometrist and GAT by an experienced tonometrist.
RESULTS: In approximately 80% of patients, the difference in IOP between GAT and ICT was <or=2 mmHg in group 1 and <or=3 mmHg in group 2. The 95% limits of agreement were -4.0-4.4 mmHg in group 1 and -6.0-5.0 mmHg in group 2.
CONCLUSION: ICT compares reasonably with GAT, in both experienced and inexperienced hands. Its ease of use, portability, and sterility make it an attractive tonometer. Its degree of accuracy in inexperienced hands would make it a useful instrument for health care workers with limited ophthalmic experience.

Entities:  

Mesh:

Year:  2006        PMID: 17159973     DOI: 10.1038/sj.eye.6702669

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  28 in total

1.  Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects.

Authors:  Ioannis Halkiadakis; Aimilianos Stratos; George Stergiopoulos; Eleni Patsea; Sotiris Skouriotis; Panagiotis Mitropoulos; Dimitrios Papaconstantinou; Gerasimos Georgopoulos
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08       Impact factor: 3.117

2.  Age variations in intraocular pressure in a cohort of healthy Austrian school children.

Authors:  W A Dusek; B K Pierscionek; J F McClelland
Journal:  Eye (Lond)       Date:  2012-03-23       Impact factor: 3.775

3.  Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties.

Authors:  J Moreno-Montañés; I Gosende; J Caire; M García-Granero
Journal:  Eye (Lond)       Date:  2010-10-01       Impact factor: 3.775

4.  The intraocular pressure response to dehydration: a pilot study.

Authors:  Andrew P Hunt; Beatrix Feigl; Ian B Stewart
Journal:  Eur J Appl Physiol       Date:  2011-08-30       Impact factor: 3.078

5.  [Rebound tonometry after vitreoretinal surgery].

Authors:  A Rosentreter; R Hoerster; T Schick; N Eter; T S Dietlein; S Fauser
Journal:  Ophthalmologe       Date:  2015-11       Impact factor: 1.059

6.  Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study.

Authors:  Senay Asik Nacaroglu; Emine Seker Un; Mehmet Giray Ersoz; Yelda Tasci
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

7.  Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses.

Authors:  Alexandra Anton; Matthias Neuburger; Daniel Böhringer; Jens F Jordan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-03-28       Impact factor: 3.117

8.  Effectiveness of the ICare rebound tonometer in patients with overestimated intraocular pressure due to tight orbit syndrome.

Authors:  You Kyung Lee; Y K Lee; Ji Young Lee; J Y Lee; Jung Il Moon; J I Moon; Myoung Hee Park; M H Park
Journal:  Jpn J Ophthalmol       Date:  2014-08-29       Impact factor: 2.447

9.  [Rebound tonometry and applanation tonometry during narcosis investigation of pediatric glaucoma].

Authors:  A Rosentreter; A M Schild; A Lappas; G K Krieglstein; T S Dietlein
Journal:  Ophthalmologe       Date:  2011-04       Impact factor: 1.059

10.  Intraocular pressure variation during colorectal laparoscopic surgery: standard pneumoperitoneum leads to reversible elevation in intraocular pressure.

Authors:  Andrea Grosso; Gitana Scozzari; Fabrizio Bert; Maria Antonietta Mabilia; Roberta Siliquini; Mario Morino
Journal:  Surg Endosc       Date:  2013-04-03       Impact factor: 4.584