Use of a computer simulator to investigate optimized tube voltage for chest imaging of average patients with a digital radiography (DR) imaging system.

Craig Steven Moore1,2, Tim Wood1,2, Ged Avery3, Steve Balcam3, Liam Needler3, Hiten Joshi3, Najeeb Ahmed3, John Saunderson1,2, Andrew Beavis1,2,4.   

Abstract

OBJECTIVE: The aim of this study was to investigate via computer simulation a proposed improvement to clinical practice by deriving an optimized tube voltage (kVp) range for digital radiography (DR) chest imaging.
METHODS: A digitally reconstructed radiograph algorithm was used which was capable of simulating DR chest radiographs containing clinically relevant anatomy. Five experienced image evaluators graded clinical image criteria, i.e. overall quality, rib, lung, hilar, spine, diaphragm and lung nodule in images of 20 patients at tube voltages across the diagnostic energy range. These criteria were scored against corresponding images of the same patient reconstructed at a specific reference kVp. Evaluators were blinded to kVp. Evaluator score for each criterion was modelled with a linear mixed effects algorithm and compared with the score for the reference image.
RESULTS: Score was dependent on tube voltage and image criteria in a statistically significant manner for both. Overall quality, hilar, diaphragm and spine criteria performed poorly at low and high tube voltages, peaking at 80-100 kVp. Lung and lung nodule demonstrated little variation. Rib demonstrated superiority at low kVp.
CONCLUSION: A virtual clinical trial has been performed with simulated chest DR images. Results indicate mid-range tube voltages of 80-100 kVp are optimum for average adults. ADVANCES IN KNOWLEDGE: There are currently no specific recommendations for optimized tube voltage parameters for DR chest imaging. This study, validated with images containing realistic anatomical noise, has investigated and recommended an optimal tube voltage range.

Entities:  

Mesh:

Year:  2019        PMID: 31559857      PMCID: PMC6913367          DOI: 10.1259/bjr.20190470

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  32 in total

1.  Schemes for the optimization of chest radiography using a computer model of the patient and x-ray imaging system.

Authors:  M Sandborg; G McVey; D R Dance; G Alm Carlsson
Journal:  Med Phys       Date:  2001-10       Impact factor: 4.071

2.  Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography.

Authors:  Patrik Sund; Magnus Båth; Susanne Kheddache; Lars Gunnar Månsson
Journal:  Eur Radiol       Date:  2003-10-16       Impact factor: 5.315

3.  Chest radiography: optimization of X-ray spectrum for cesium iodide-amorphous silicon flat-panel detector.

Authors:  James T Dobbins; Ehsan Samei; Harrell G Chotas; Richard J Warp; Alan H Baydush; Carey E Floyd; Carl E Ravin
Journal:  Radiology       Date:  2003-01       Impact factor: 11.105

4.  Validation of a large-scale audit technique for CT dose optimisation.

Authors:  T J Wood; A W Davis; C S Moore; A W Beavis; J R Saunderson
Journal:  Radiat Prot Dosimetry       Date:  2011-11-16       Impact factor: 0.972

5.  Nodule detection in digital chest radiography: effect of anatomical noise.

Authors:  Magnus Båth; Markus Håkansson; Sara Börjesson; Christoph Hoeschen; Oleg Tischenko; Susanne Kheddache; Jenny Vikgren; Lars Gunnar Månsson
Journal:  Radiat Prot Dosimetry       Date:  2005       Impact factor: 0.972

6.  Nodule detection in digital chest radiography: summary of the RADIUS chest trial.

Authors:  Markus Håkansson; Magnus Båth; Sara Börjesson; Susanne Kheddache; Anna Grahn; Mark Ruschin; Anders Tingberg; Sören Mattsson; Lars Gunnar Månsson
Journal:  Radiat Prot Dosimetry       Date:  2005       Impact factor: 0.972

7.  Comparison of clinical and physical measures of image quality in chest and pelvis computed radiography at different tube voltages.

Authors:  Michael Sandborg; Anders Tingberg; Gustaf Ullman; David R Dance; Gudrun Alm Carlsson
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

8.  Optimisation of radiological protocols for chest imaging using computed radiography and flat-panel X-ray detectors.

Authors:  G Compagnone; M Casadio Baleni; E Di Nicola; M Valentino; M Benati; L F Calzolaio; N Oberhofer; E Fabbri; S Domenichelli; L Barozzi
Journal:  Radiol Med       Date:  2012-10-22       Impact factor: 3.469

9.  Computerized three-dimensional segmented human anatomy.

Authors:  I G Zubal; C R Harrell; E O Smith; Z Rattner; G Gindi; P B Hoffer
Journal:  Med Phys       Date:  1994-02       Impact factor: 4.071

10.  Towards image quality, beam energy and effective dose optimisation in digital thoracic radiography.

Authors:  J H Launders; A R Cowen; R F Bury; P Hawkridge
Journal:  Eur Radiol       Date:  2001       Impact factor: 5.315

View more
  1 in total

1.  Optimisation of tube voltage range (kVp) for AP abdomen, pelvis and spine imaging of average patients with a digital radiography (DR) imaging system using a computer simulator.

Authors:  Craig Steven Moore; Tim Wood; Stephen Balcam; Liam Needler; Tim Guest; Wee Ping Ngu; Lee Wun Chong; John Saunderson; Andrew Beavis
Journal:  Br J Radiol       Date:  2020-08-12       Impact factor: 3.039

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.