A clinical evaluation of proview pressure phosphene tonometry in children.

Ida Chung1, Amelia Bartolone, William H Swanson, Andrea P Thau.   

Abstract

PURPOSE: The Proview tonometer measures intraocular pressure by inducing a pressure phosphene through the eyelid and, if reliable and valid, may offer a quick, nonthreatening and noninvasive alternative method of obtaining intraocular pressures (IOPs) without the use of eye drops. This study compares the IOP measurements obtained in children using Proview pressure phosphene tonometry (PPPT) and Goldmann tonometry (GT).
METHODS: One hundred four 5- to 12-year-old patients of the University Optometric Center/SUNY College of Optometry participated in the study. Subjects were randomized to receive, by different investigators, either PPPT or GT first. Two measurements with each instrument were attempted on each eye of all subjects. A subgroup of 41 subjects was asked which of the two methods was preferred.
RESULTS: Seven percent of the subjects did not report a pressure phosphene response compared with 12% of the subjects on whom the investigators were unable to perform GT. The remaining 85 subjects completed the subject protocol. Of the 41 subjects asked, 56% preferred PPPT, 24% had no preference, and 20% preferred GT. The coefficient of repeatability between the two readings was higher for PPPT (3-4 mm Hg) than for GT (1 mm Hg). Mean IOP was 4 mm Hg higher for PPPT than GT with the difference in readings between the two instruments increasing with higher IOPs (r >19%, p < 0.005).
CONCLUSIONS: In our study of healthy young subjects, PPPT measurements of IOP appear to be repeatable within a few millimeters of mercury in most children, but for some children, variability in repeat measurements can be substantial. Our data showed a mean difference in readings of 4 mm Hg with a 95% confidence interval that the PPPT reading was between 12 mm Hg above GT and 4 mm Hg below GT. This wide range of values indicates that PPPT is not comparable to GT. However, because our study found that children can appreciate pressure phosphenes and most prefer PPPT over GT, the Proview monitor may have value as a noninvasive, portable screener in pediatric patients. To fully evaluate this potential, further studies are needed that include patients with high IOPs.

Entities:  

Mesh:

Year:  2006        PMID: 17106408      PMCID: PMC3124284          DOI: 10.1097/01.opx.0000239101.75107.3c

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  14 in total

1.  The site of electrical excitation of the human eye.

Authors:  G S BRINDLEY
Journal:  J Physiol       Date:  1955-01-28       Impact factor: 5.182

2.  The x-ray and radium phosphenes.

Authors:  L E LIPETZ
Journal:  Br J Ophthalmol       Date:  1955-10       Impact factor: 4.638

3.  The pressure phosphene tonometer--a clinical evaluation.

Authors:  G S M Chew; G F Sanderson; A C B Molteno
Journal:  Eye (Lond)       Date:  2005-06       Impact factor: 3.775

4.  The Proview Eye Pressure Monitor: influence of clinical factors on accuracy and agreement with the Goldmann tonometer.

Authors:  Peter Herse; Avleen Hans; Julia Hall; Jessica Langejans; Maria Markoulli
Journal:  Ophthalmic Physiol Opt       Date:  2005-09       Impact factor: 3.117

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Reproducibility of self-measured intraocular pressure with the phosphene tonometer in patients with ocular hypertension and early to advanced glaucoma.

Authors:  Luca Brigatti; Srigouri Maguluri
Journal:  J Glaucoma       Date:  2005-02       Impact factor: 2.503

7.  Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients.

Authors:  Dennis S C Lam; Dexter Y L Leung; Thomas Y H Chiu; Dorothy S P Fan; Eva Y Y Cheung; Tien-Yin Wong; Jimmy S M Lai; Clement C Y Tham
Journal:  Invest Ophthalmol Vis Sci       Date:  2004-09       Impact factor: 4.799

8.  The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice.

Authors:  Tara L Alvarez; Stephen A Gollance; Gordon A Thomas; Richard J Greene; Peter M Marchetto; Eugene J Moore; Tony Realini; Jeffrey M Liebmann; Robert Ritch; Paul Lama; Robert D Fechtner
Journal:  Ophthalmology       Date:  2004-06       Impact factor: 12.079

9.  Comparison of the Proview pressure phosphene tonometer performed by the patient and examiner with the Goldmann applanation tonometer.

Authors:  Helen V Danesh-Meyer; Rachel Niederer; Brent J Gaskin; Greg Gamble
Journal:  Clin Exp Ophthalmol       Date:  2004-02       Impact factor: 4.207

10.  A new tonometer--the pressure phosphene tonometer: clinical comparison with Goldman tonometry.

Authors:  B B Fresco
Journal:  Ophthalmology       Date:  1998-11       Impact factor: 12.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.