Reproducibility of self-measured intraocular pressure with the phosphene tonometer in patients with ocular hypertension and early to advanced glaucoma.

Luca Brigatti1, Srigouri Maguluri.   

Abstract

PURPOSE: To evaluate the reproducibility of self-measured intraocular pressure with the Proview tonometer in ocular hypertensives and early to advanced glaucoma patients. PATIENTS AND METHODS: Thirty-three patients with glaucoma and three with ocular hypertension (72 eyes) were prospectively enrolled. History of cornea surgery was the only exclusion criterion. A trainer took a set of three readings first. Subsequently, the patient (after successful training) took one set of three readings and a physician took two sets of three Goldmann readings. One hour later this was repeated.
RESULTS: 72% of patients could see the phosphene in both eyes, 8% could see it in one eye, and 19% could not see it in either eye. Patients who could not see the Phosphene had an average mean deviation of -10.9 compared with -3.4 for those who could see it (P = 0.01). Eyes that could not see the phosphene were excluded from further analysis. The average absolute difference between two Proview measurements one hour apart was 1.76 +/- 1.76 mm Hg (r = 0.71). The measurement error (difference between Goldmann and Proview readings) tended to remain constant between sets of measurements taken one hour apart (average error was 2.4 +/- 2.1 mm Hg, r = 0.85). No adverse event was recorded.
CONCLUSIONS: Repeated self-tonometry measurements showed very good reproducibility. The systematic, reproducible error between Proview and Goldmann readings may be clinically acceptable and suggests that a baseline "calibration" should be obtained for each patient before the Proview values are used clinically. Self-administered tonometry can play a role in the clinical care of ocular hypertensive and glaucoma patients with the possible exception of very advanced cases.

Entities:  

Mesh:

Year:  2005        PMID: 15650602     DOI: 10.1097/01.ijg.0000146374.59119.42

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  3 in total

1.  A clinical evaluation of proview pressure phosphene tonometry in children.

Authors:  Ida Chung; Amelia Bartolone; William H Swanson; Andrea P Thau
Journal:  Optom Vis Sci       Date:  2006-11       Impact factor: 1.973

2.  Effects of topical travoprost 0.004% on intraocular pressure and corneal biomechanical properties in an animal model.

Authors:  Gabriel Lazcano-Gomez; David Ancona-Lezama; Felix Gil-Carrasco; Jesus Jimenez-Roman
Journal:  Digit J Ophthalmol       Date:  2016-02-11

Review 3.  Ageing and ocular surface immunity.

Authors:  Alireza Mashaghi; Jiaxu Hong; Sunil K Chauhan; Reza Dana
Journal:  Br J Ophthalmol       Date:  2016-07-04       Impact factor: 4.638

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.