Comparison of the Proview pressure phosphene tonometer performed by the patient and examiner with the Goldmann applanation tonometer.

Helen V Danesh-Meyer1, Rachel Niederer, Brent J Gaskin, Greg Gamble.   

Abstract

OBJECTIVE: To compare the results of Proview pressure phosphene tonometry (PPPT) performed by the patient and an examiner with Goldmann applanation tonometry (GAT).
METHODS: A comparative case series of 96 (192 eyes) consecutive patients from a glaucoma clinic was conducted. Intraocular pressure (IOP) was measured with GAT by one examiner, PPPT by another examiner, and PPPT by the patient. All examiners were masked to the results from any preceding IOP measurement.
RESULTS: The coefficient of repeatability for repeated measurements for the GAT was +/-0.48 mmHg. The limits of agreement for self-measurement of IOP with the PPPT and examiner measured IOP with PPPT were 6.3 mmHg and 4.8 mmHg, respectively. The limits of agreement between GAT and self assessed PPPT were +/-11.8 mmHg (mean difference of 0.63 mmHg). When the same comparison was made between GAT and examiner assessed PPPT, the results were limits of agreement of +/-10 mmHg (mean difference of 2.86 mmHg). No significant difference was identified in the agreement of the GAT and the PPPT when subanalysed for age of patient or diagnosis (P > 0.05). The limits of agreement between self-assessed IOP with the PPPT and the GAT were +/-8.2 for those with IOP <20 mmHg and +/-14.9 mmHg for those with IOP>20 mmHg.
CONCLUSIONS: Poor agreement exists between IOP measured by GAT and PPPT measured by an examiner or by the patient.

Entities:  

Mesh:

Year:  2004        PMID: 14746587     DOI: 10.1046/j.1442-9071.2004.00753.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  4 in total

1.  [Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry].

Authors:  V Witte; Ä Glass; R Beck; R Guthoff
Journal:  Ophthalmologe       Date:  2012-10       Impact factor: 1.059

2.  A clinical evaluation of proview pressure phosphene tonometry in children.

Authors:  Ida Chung; Amelia Bartolone; William H Swanson; Andrea P Thau
Journal:  Optom Vis Sci       Date:  2006-11       Impact factor: 1.973

3.  Effects of topical travoprost 0.004% on intraocular pressure and corneal biomechanical properties in an animal model.

Authors:  Gabriel Lazcano-Gomez; David Ancona-Lezama; Felix Gil-Carrasco; Jesus Jimenez-Roman
Journal:  Digit J Ophthalmol       Date:  2016-02-11

Review 4.  An overview of home tonometry and telemetry for intraocular pressure monitoring in humans.

Authors:  Edward Yung; Valerie Trubnik; L Jay Katz
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-29       Impact factor: 3.117

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.