Calibration of an amorphous-silicon flat panel portal imager for exit-beam dosimetry.

Josephine Chen1, Cynthia F Chuang, Olivier Morin, Michèle Aubin, Jean Pouliot.   

Abstract

Amorphous-silicon flat panel detectors are currently used to acquire digital portal images with excellent image quality for patient alignment before external beam radiation therapy. As a first step towards interpreting portal images acquired during treatment in terms of the actual dose delivered to the patient, a calibration method is developed to convert flat panel portal images to the equivalent water dose deposited in the detector plane and at a depth of 1.5 cm. The method is based on empirical convolution models of dose deposition in the flat panel detector and in water. A series of calibration experiments comparing the response of the flat panel imager and ion chamber measurements of dose in water determines the model parameters. Kernels derived from field size measurements account for the differences in the production and detection of scattered radiation in the two systems. The dissimilar response as a function of beam energy spectrum is characterized from measurements performed at various off-axis positions and for increasing attenuator thickness in the beam. The flat panel pixel inhomogeneity is corrected by comparing a large open field image with profiles measured in water. To verify the accuracy of the calibration method, calibrated flat panel profiles were compared with measured dose profiles for fields delivered through solid water slabs, a solid water phantom containing an air cavity, and an anthropomorphic head phantom. Open rectangular fields of various sizes and locations as well as a multileaf collimator-shaped field were delivered. For all but the smallest field centered about the central axis, the calibrated flat panel profiles matched the measured dose profiles with little or no systematic deviation and approximately 3% (two standard deviations) accuracy for the in-field region. The calibrated flat panel profiles for fields located off the central axis showed a small -1.7% systematic deviation from the measured profiles for the in-field region. Out of the field, the differences between the calibrated flat panel and measured profiles continued to be small, approximately 0%-2% of the mean in-field dose. Further refinement of the calibration model should increase the accuracy of the procedure. This calibration method for flat panel portal imagers may be used as part of a validation scheme to verify the dose delivered to the patient during treatment.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16878562     DOI: 10.1118/1.2168294

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules.

Authors:  Janett Liebich; Jörg Licher; Christian Scherf; Eugen Kara; Nadine Koch; Claus Rödel; Ulla Ramm
Journal:  Strahlenther Onkol       Date:  2011-04-26       Impact factor: 3.621

2.  Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy.

Authors:  Aiping Ding; Lei Xing; Bin Han
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

3.  In patient dose reconstruction using a cine acquisition for dynamic arc radiation therapy.

Authors:  Angelo Piermattei; Andrea Fidanzio; Luigi Azario; Francesca Greco; Alessandra Mameli; Savino Cilla; Luca Grimaldi; Guido D'Onofrio; Boris Giuseppe Augelli; Gerardina Stimato; Diego Gaudino; Sara Ramella; Rolando D'Angelillo; Francesco Cellini; Lucio Trodella
Journal:  Med Biol Eng Comput       Date:  2009-02-17       Impact factor: 2.602

4.  Transit dosimetry in dynamic IMRT with an a-Si EPID.

Authors:  Mahsheed Sabet; Pejman Rowshanfarzad; Fred W Menk; Peter B Greer
Journal:  Med Biol Eng Comput       Date:  2014-05-31       Impact factor: 2.602

5.  Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software.

Authors:  Mareike Held; Joey Cheung; Angelica Perez Andujar; François Husson; Olivier Morin
Journal:  Cureus       Date:  2018-02-02

6.  Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry.

Authors:  Heeteak Chung; Jonathan Li; Sanjiv Samant
Journal:  J Appl Clin Med Phys       Date:  2011-04-08       Impact factor: 2.102

7.  A new approach for the pixel map sensitivity (PMS) evaluation of an electronic portal imaging device (EPID).

Authors:  Alberto Boriano; Francesco Lucio; Elisa Calamia; Elvio Russi; Flavio Marchetto
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

8.  Calculation of exit dose for conformal and dynamically-wedged fields, based on water-equivalent path length measured with an amorphous silicon electronic portal imaging device.

Authors:  Awusi Kavuma; Martin Glegg; Mohamed Metwaly; Garry Currie; Alex Elliott
Journal:  J Appl Clin Med Phys       Date:  2011-03-03       Impact factor: 2.102

9.  An empirical model of electronic portal imager response implemented within a commercial treatment planning system for verification of intensity-modulated radiation therapy fields.

Authors:  Rao F H Khan; Orest Z Ostapiak; Joe J Szabo
Journal:  J Appl Clin Med Phys       Date:  2008-11-11       Impact factor: 2.102

10.  An Empirical Transmitted EPID Dosimetry Method using a Back-Projection Algorithm.

Authors:  Hashemi S M; Bahreyni M H; Mohammadi M; Nasseri S; Bayani S; Gholamhosseinian H; Salek R; Shahedi F; Momennezhad M
Journal:  J Biomed Phys Eng       Date:  2019-10-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.