Literature DB >> 16696485

A bench-top megavoltage fan-beam CT using CdWO4-photodiode detectors. I. System description and detector characterization.

S Rathee1, D Tu, T T Monajemi, D W Rickey, B G Fallone.   

Abstract

We describe the components of a bench-top megavoltage computed tomography (MVCT) scanner that uses an 80-element detector array consisting of CdWO4 scintillators coupled to photodiodes. Each CdWO4 crystal is 2.75 x 8 x 10 mm3. The detailed design of the detector array, timing control, and multiplexer are presented. The detectors show a linear response to dose (dose rate was varied by changing the source to detector distance) with a correlation coefficient (R2) nearly unity with the standard deviation of signal at each dose being less than 0.25%. The attenuation of a 6 MV beam by solid water measured by this detector array indicates a small, yet significant spectral hardening that needs to be corrected before image reconstruction. The presampled modulation transfer function is strongly affected by the detector's large pitch and a large improvement can be obtained by reducing the detector pitch. The measured detective quantum efficiency at zero spatial frequency is 18.8% for 6 MV photons which will reduce the dose to the patient in MVCT applications. The detector shows a less than a 2% reduction in response for a dose of 24.5 Gy accumulated in 2 h; however, the lost response is recovered on the following day. A complete recovery can be assumed within the experimental uncertainty (standard deviation <0.5%); however, any smaller permanent damage could not be assessed.

Entities:  

Mesh:

Year:  2006        PMID: 16696485     DOI: 10.1118/1.2181290

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Monte Carlo investigations of megavoltage cone-beam CT using thick, segmented scintillating detectors for soft tissue visualization.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Amit Sawant; Hong Du
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

2.  High-DQE EPIDs based on thick, segmented BGO and CsI:Tl scintillators: performance evaluation at extremely low dose.

Authors:  Yi Wang; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Louis Perna
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

3.  A Monte Carlo investigation of Swank noise for thick, segmented, crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

4.  Monte Carlo investigations of the effect of beam divergence on thick, segmented crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Youcef El-Mohri; Larry E Antonuk; Qihua Zhao
Journal:  Phys Med Biol       Date:  2010-06-04       Impact factor: 3.609

5.  Optimization of the performance of segmented scintillators for radiotherapy imaging through novel binning techniques.

Authors:  Youcef El-Mohri; Larry E Antonuk; Richard B Choroszucha; Qihua Zhao; Hao Jiang; Langechuan Liu
Journal:  Phys Med Biol       Date:  2014-02-03       Impact factor: 3.609

6.  Low-dose megavoltage cone-beam CT imaging using thick, segmented scintillators.

Authors:  Youcef El-Mohri; Larry E Antonuk; Qihua Zhao; Richard B Choroszucha; Hao Jiang; Langechuan Liu
Journal:  Phys Med Biol       Date:  2011-02-16       Impact factor: 3.609

7.  A piecewise-focused high DQE detector for MV imaging.

Authors:  Josh Star-Lack; Daniel Shedlock; Dennis Swahn; Dave Humber; Adam Wang; Hayley Hirsh; George Zentai; Daren Sawkey; Isaac Kruger; Mingshan Sun; Eric Abel; Gary Virshup; Mihye Shin; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

8.  Countering beam divergence effects with focused segmented scintillators for high DQE megavoltage active matrix imagers.

Authors:  Langechuan Liu; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Hao Jiang
Journal:  Phys Med Biol       Date:  2012-08-01       Impact factor: 3.609

9.  Development of a novel high quantum efficiency MV x-ray detector for image-guided radiotherapy: A feasibility study.

Authors:  Jian Liu; Yuan Xu; Aram Teymurazyan; Zisis Papandreou; Geordi Pang
Journal:  Med Phys       Date:  2019-11-04       Impact factor: 4.071

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.