| Literature DB >> 16645101 |
M J Oddy1, M J Jones, C J Pendegrass, J R Pilling, J A Wimhurst.
Abstract
In 20 patients undergoing hybrid total hip arthroplasty, the reproducibility and accuracy of templating using digital radiographs were assessed. Digital images were manipulated using either a ten-pence coin as a marker to scale for magnification, or two digital-line methods using computer software. On-screen images were templated with standard acetate templates and compared with templating performed on hard-copy digital prints. The digital-line methods were the least reliable and accuracy of sizing compared with the inserted prostheses varied between -1.6% and +10.2%. The hard-copy radiographs showed better reproducibility than the ten-pence coin method, but were less accurate with 3.7% undersizing. The ten-pence coin method was the most accurate, with no significant differences for offset or acetabulum, and undersizing of only 0.9%. On-screen templating of digital radiographs with standard acetate templates is accurate and reproducible if a radiopaque marker such as a ten-pence coin is included when the original radiograph is taken.Entities:
Mesh:
Year: 2006 PMID: 16645101 DOI: 10.1302/0301-620X.88B5.17184
Source DB: PubMed Journal: J Bone Joint Surg Br ISSN: 0301-620X