High iodine concentration contrast material for noninvasive multislice computed tomography coronary angiography: iopromide 370 versus iomeprol 400.

Filippo Cademartiri1, Cecile de Monye, Francesca Pugliese, Nico R Mollet, Giuseppe Runza, Aad van der Lugt, Massimo Midiri, Pim J de Feyter, Roberto Lagalla, Gabriel P Krestin.   

Abstract

OBJECTIVE: The objective of this study was to compare intracoronary attenuation on 16-row multislice computed tomography (16-MSCT) coronary angiography using 2 contrast materials (CM) with high iodine concentration.
MATERIAL AND METHODS: Forty consecutive patients (29 male, 11 female; mean age, 61+/-11 years) with suspected coronary artery disease were randomized to 2 groups to receive 100 mL of either iopromide 370 (group 1: Ultravist 370, 370 mg iodine/mL; Schering AG, Berlin, Germany) or iomeprol 400 (group 2: Iomeron 400, 400 mg iodine/mL; Bracco Imaging SpA, Milan, Italy). Both CM were administered at a rate of 4 mL/s. All patients underwent 16-MSCT coronary angiography (Sensation 16; Siemens, Germany) with collimation 16 x 0.75 mm and rotation time 375 ms. The attenuation in Hounsfield units (HU) achieved after each CM was determined at regions of interest (ROIs) placed at the origin of coronary arteries and on the ascending aorta, descending aorta, and pulmonary artery. Differences in mean attenuation in the coronary arteries and on the ascending aorta, descending aorta, and pulmonary artery were evaluated using Student t test.
RESULTS: The mean attenuation achieved at each anatomic site was consistently greater after iomeprol 400 than after iopromide 370. At the origin of coronary arteries, the mean attenuation after iomeprol 400 (340+/-53 HU) was greater (P<0.05) than that after iopromide 370 (313+/- 42 HU). Similar findings were noted for the mean attenuation in the ascending aorta, descending aorta, and pulmonary artery.
CONCLUSION: The intravenous administration of iomeprol 400 provides higher attenuation of the coronary arteries and of the great arteries of the thorax as compared with iopromide 370 using the same injection parameters.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16481919     DOI: 10.1097/01.rli.0000191369.76521.09

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  25 in total

1.  64-Slice multidetector row CT angiography of the abdomen: comparison of low versus high concentration iodinated contrast media in a porcine model.

Authors:  N-S Holalkere; K Matthes; S P Kalva; W R Brugge; D V Sahani
Journal:  Br J Radiol       Date:  2010-11-16       Impact factor: 3.039

2.  Attenuation values of the intracranial arterial and venous vessels by bolus injection of various contrast agents: a study with a single-detector helical CT scanner.

Authors:  Morio Nagahata; Yoshinao Abe; Shuichi Ono; Hikaru Yamaguchi; Hiroyuki Miura; Takashi Ohata; Fumiyasu Tsushima; Kohei Morimoto; Hiroko Seino
Journal:  Radiat Med       Date:  2007-04-27

3.  Influence of intra-coronary enhancement on diagnostic accuracy with 64-slice CT coronary angiography.

Authors:  Filippo Cademartiri; Erica Maffei; Anselmo Alessandro Palumbo; Roberto Malagò; Ludovico La Grutta; W Bob Meiijboom; Annachiara Aldrovandi; Michele Fusaro; Luigi Vignali; Alberto Menozzi; Valerio Brambilla; Paolo Coruzzi; Massimo Midiri; Miles A Kirchin; Nico R A Mollet; Gabriel P Krestin
Journal:  Eur Radiol       Date:  2007-10-13       Impact factor: 5.315

4.  MDCT angiography of the pulmonary arteries: intravascular contrast enhancement does not depend on iodine concentration when injecting equal amounts of iodine at standardized iodine delivery rates.

Authors:  S Keil; C Plumhans; F F Behrendt; M Das; S Stanzel; G Mühlenbruch; P Seidensticker; C Knackstedt; A H Mahnken; R W Günther; J E Wildberger
Journal:  Eur Radiol       Date:  2008-04-08       Impact factor: 5.315

5.  Intra-individual comparison of different contrast media concentrations (300 mg, 370 mg and 400 mg iodine) in MDCT.

Authors:  Florian F Behrendt; Hubertus Pietsch; Gregor Jost; Martin A Sieber; Sebastian Keil; Cedric Plumhans; Peter Seidensticker; Rolf W Günther; Andreas H Mahnken
Journal:  Eur Radiol       Date:  2010-02-24       Impact factor: 5.315

6.  Indications, imaging technique, and reading of cardiac computed tomography: survey of clinical practice.

Authors:  M H Maurer; E Zimmermann; P Schlattmann; C Germershausen; B Hamm; Marc Dewey
Journal:  Eur Radiol       Date:  2011-08-16       Impact factor: 5.315

7.  Contrast agent bolus tracking with a fixed threshold or a manual fast start for coronary CT angiography.

Authors:  Fabian Stenzel; Matthias Rief; Elke Zimmermann; Johannes Greupner; Felicitas Richter; Marc Dewey
Journal:  Eur Radiol       Date:  2014-03-26       Impact factor: 5.315

8.  Contrast media volume optimization in high-pitch dual-source CT coronary angiography: feasibility study.

Authors:  Wen Jie Yang; Ke Min Chen; Bo Liu; Li Fang Pang; Huan Zhang; Zi Lai Pan; Fu Hua Yan
Journal:  Int J Cardiovasc Imaging       Date:  2012-05-23       Impact factor: 2.357

9.  Biphasic contrast medium injection in cardiac CT: moderate versus high concentration contrast material at identical iodine flux and iodine dose.

Authors:  Annemarieke Rutten; Matthijs F L Meijs; Alexander M de Vos; Peter R Seidensticker; Mathias Prokop
Journal:  Eur Radiol       Date:  2010-03-20       Impact factor: 5.315

10.  Image quality and attenuation values of multidetector CT coronary angiography using high iodine-concentration contrast material: a comparison of the use of iopromide 370 and iomeprol 400.

Authors:  Eun Young Kim; Dae Wook Yeh; Yeon Hyeon Choe; Won Jae Lee; Hyo Keun Lim
Journal:  Acta Radiol       Date:  2010-11       Impact factor: 1.990

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.