Mathematical Model analysis of Wallstent and Aneurx: dynamic responses of bare-metal endoprosthesis compared with those of stent-graft.

Suncica Canic1, K Ravi-Chandar, Zvonimir Krajcer, Dragan Mirkovic, Serguei Lapin.   

Abstract

We performed this study in order to analyze the mechanical properties of bare-metal Wallstent endoprostheses and of AneuRx stent-grafts and to compare their responses to hemodynamic forces. Mathematical modeling, numerical simulations, and experimental measurements were used to study the 2 structurally different types of endoprostheses. Our findings revealed that a single bare-metal Wallstent endoprosthesis is 10 times more flexible (elastic) than is the wall of the aneurysmal abdominal aorta. Graphs showing the changes in the diameter and length of the stent when exposed to a range of internal and external pressures were obtained. If the aorta is axially stiff and resists length change, a force as large as 1 kg can act in the axial direction on the aortic wall. If the stent is not firmly anchored, it will migrate. In contrast, a fabric-covered, fully supported, stent-graft such as the AneuRx is significantly less compliant than the aorta or the bare-metal stent. During each cardiac cycle, the stent frame tends to move due to its higher elasticity, while the fabric resists movement, which might break the sutures that join the fabric to the frame. Elevated local transmural pressure, detected along the prosthesis graft, can contribute to material fatigue.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16429893      PMCID: PMC1351820     

Source DB:  PubMed          Journal:  Tex Heart Inst J        ISSN: 0730-2347


  10 in total

1.  Noninvasive measurement of diameter changes in the distal abdominal aorta in man.

Authors:  T Länne; H Stale; H Bengtsson; D Gustafsson; D Bergqvist; B Sonesson; H Lecerof; P Dahl
Journal:  Ultrasound Med Biol       Date:  1992       Impact factor: 2.998

2.  A novel method for endoluminal treatment of abdominal aortic aneurysms. With bare-metal Wallstent endoprostheses and endovascular coils.

Authors:  A Achari; Z Krajcer
Journal:  Tex Heart Inst J       Date:  1998

3.  Time-related alterations in shape, position, and structure of self-expanding, modular aortic stent-grafts: a 4-year single-center follow-up.

Authors:  T Umscheid; W J Stelter
Journal:  J Endovasc Surg       Date:  1999-02

4.  Endoluminal exclusion of an iliac artery aneurysm by Wallstent endoprosthesis and PTFE vascular graft.

Authors:  Z Krajcer; R Khoshnevis; D R Leachman; H Herman
Journal:  Tex Heart Inst J       Date:  1997

5.  Endovascular exclusion of abdominal aortic aneurysms: initial experience with stent-grafts in cardiology practice.

Authors:  M H Howell; M Zaqqa; R P Villareal; N E Strickman; Z Krajcer
Journal:  Tex Heart Inst J       Date:  2000

Review 6.  The AneuRx stent graft: four-year results and worldwide experience 2000.

Authors:  C K Zarins; R A White; F L Moll; T Crabtree; D A Bloch; K J Hodgson; M F Fillinger; T J Fogarty
Journal:  J Vasc Surg       Date:  2001-02       Impact factor: 4.268

7.  Predictors of primary patency failure in Wallstent self-expanding endovascular prostheses for iliofemoral occlusive disease.

Authors:  S Damaraju; L Cuasay; D Le; N Strickman; Z Krajcer
Journal:  Tex Heart Inst J       Date:  1997

8.  Comparison of Wallgraft and Wallstent for treatment of complex iliac artery stenosis and occlusion. Preliminary results of a prospective randomized study.

Authors:  Z Krajcer; G Sioco; T Reynolds
Journal:  Tex Heart Inst J       Date:  1997

9.  Mechanical properties of the aneurysmal aorta.

Authors:  S T MacSweeney; G Young; R M Greenhalgh; J T Powell
Journal:  Br J Surg       Date:  1992-12       Impact factor: 6.939

10.  Explant analysis of AneuRx stent grafts: relationship between structural findings and clinical outcome.

Authors:  Christopher K Zarins; Frank R Arko; Tami Crabtree; Daniel A Bloch; Kenneth Ouriel; Robert C Allen; Rodney A White
Journal:  J Vasc Surg       Date:  2004-07       Impact factor: 4.268

  10 in total
  2 in total

1.  Intravascular laser speckle imaging catheter for the mechanical evaluation of the arterial wall.

Authors:  Zeinab Hajjarian; Jingqun Xi; Farouc A Jaffer; Guillermo J Tearney; Seemantini K Nadkarni
Journal:  J Biomed Opt       Date:  2011-02       Impact factor: 3.170

2.  Finite element modeling of a novel self-expanding endovascular stent method in treatment of aortic aneurysms.

Authors:  Mark C Arokiaraj; Igor F Palacios
Journal:  Sci Rep       Date:  2014-01-10       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.