BACKGROUND: Use of the Harmonic Scalpel transduces a lower amount of energy to tissues, thereby limiting the potential for lateral thermal damage and deep penetration because only low temperatures are reached. The working principle of the Harmonic Scalpel is the transformation of electric power into the longitudinal mechanical movement of the instrument tip. This study aimed to determine the effects from varying durations of Harmonic Scalpel application on the experimental model of rat abdominal wall without skin.
METHODS: After the rats had been anesthetized, and a laparotomy was performed. The Harmonic Scalpel shears were used on the muscular part of the abdominal wall without skin. Different durations of output power 3 were applied: a single 5-s application, a single 10-s application, and a regimen of two sequential 5-s applications. Each animal in each group of 10 received five individual activations, after which the animals were killed. Tissue samples were fixed and embedded in paraffin before sections were taken and stained. Using light microscopy and morphometric imaging analysis, the width of tissue lateral thermal damage was measured from the point of Harmonic Scalpel incision.
RESULTS: The rat abdominal wall showed lateral thermal damage over a mean width of 0.0522 +/- 0.0097 mm after a 5-s Harmonic Scalpel application, a damage width of 0.1544 +/- 0.0419 mm after a 10-s application, and a damage width of 0.1020 +/- 0.0430 mm after a 5-s application followed by 5 s of inactivity and another 5 s of activity. These differences in thermal damage width between all the groups are statistically significant.
CONCLUSIONS: The findings lead to the conclusion that tissue lateral thermal damage after Harmonic Scalpel application at standard output power is greater when a longer sustained period of application is used. Lateral thermal damage also is greater if the Harmonic Scalpel application time is continuous rather than of the same total duration with a brief midpoint interruption.

Entities:  

Mesh:

Year:  2005        PMID: 16333532     DOI: 10.1007/s00464-005-0089-6

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  9 in total

1.  [Use of the harmonic scalpel (Ultracision) in laparoscopic antireflux surgery].

Authors:  G Bischof; J Zacherl; M Imhof; R Jakesz; R Függer
Journal:  Zentralbl Chir       Date:  1999       Impact factor: 0.942

2.  UltraCision or high-frequency knife in transanal endoscopic microsurgery (TEM)? Advantages of a new procedure.

Authors:  C Langer; P Markus; T Liersch; L Füzesi; H Becker
Journal:  Surg Endosc       Date:  2001-03-13       Impact factor: 4.584

3.  How safe is high-power ultrasonic dissection?

Authors:  Tarek A Emam; Alfred Cuschieri
Journal:  Ann Surg       Date:  2003-02       Impact factor: 12.969

4.  Chemical composition of smoke produced by high-frequency electrosurgery in a closed gaseous environment. An in vitro study.

Authors:  C Hensman; D Baty; R G Willis; A Cuschieri
Journal:  Surg Endosc       Date:  1998-08       Impact factor: 4.584

5.  [Laparoscopic fenestration of non-parasitic liver cysts].

Authors:  F Köckerling; C Schneider; T Reck; H Scheuerlein; W Hohenberger
Journal:  Zentralbl Chir       Date:  1997       Impact factor: 0.942

6.  Aerosol exposure from an ultrasonically activated (Harmonic) device.

Authors:  D E Ott; E Moss; K Martinez
Journal:  J Am Assoc Gynecol Laparosc       Date:  1998-02

7.  [Harmonic scalpel in laparoscopic surgery].

Authors:  Zdravko Perko; Zeljko Mimica; Nikica Druzijanić; Damir Kraljević; Ante Petricević; Arsen Depolo; Zoran Cala; Harry Grbas
Journal:  Lijec Vjesn       Date:  2004 Sep-Oct

8.  Ultrasonic energy vs monopolar electrosurgery in laparoscopic cholecystectomy: influence on the postoperative systemic immune response.

Authors:  C Sietses; B M von Blomberg; M A Cuesta
Journal:  Surg Endosc       Date:  2001-01       Impact factor: 4.584

9.  Laparoscopic cholecystectomy in 200 consecutive patients using an ultrasonically activated scalpel.

Authors:  J F Amaral
Journal:  Surg Laparosc Endosc       Date:  1995-08
  9 in total
  14 in total

1.  Heat production, nerve function, and morphology following nerve close dissection with surgical instruments.

Authors:  Johan Carlander; C Koch; L Brudin; C Nordborg; O Gimm; K Johansson
Journal:  World J Surg       Date:  2012-06       Impact factor: 3.352

Review 2.  Surgical Efficacy Among Laparoscopic Ultrasonic Dissectors: Are We Advancing Safely? A Review of Literature.

Authors:  Rajesh Devassy; Sreelatha Gopalakrishnan; Rudy Leon De Wilde
Journal:  J Obstet Gynaecol India       Date:  2015-09-14

3.  Ultrasonic thermal damage during robotic hysterectomy.

Authors:  Jason Massengill; Eric Lombardini; Jonathan Oliva; Jerome Buller; Daniel Gruber
Journal:  J Robot Surg       Date:  2014-04-27

4.  Usefulness of T-Shaped Gauze for Precise Dissection of Supra-Pancreatic Lymph Nodes and for Reduced Postoperative Pancreatic Fistula in Patients Undergoing Laparoscopic Gastrectomy for Gastric Cancer.

Authors:  Tomoyuki Matsunaga; Hiroaki Saito; Yuki Murakami; Hirohiko Kuroda; Yoji Fukumoto; Tomohiro Osaki
Journal:  Yonago Acta Med       Date:  2016-09-12       Impact factor: 1.641

5.  Laparoscopic transabdominal preperitoneal approach for inguinal hernia repair: a five-year experience at a single center.

Authors:  Zdravko Perko; Mislav Rakić; Zenon Pogorelić; Nikica Družijanić; Jasenka Kraljević
Journal:  Surg Today       Date:  2011-01-26       Impact factor: 2.549

6.  A new approach to laparoscopic appendectomy in children-clipless/sutureless Harmonic scalpel laparoscopic appendectomy.

Authors:  Zenon Pogorelić; Vedran Beara; Miro Jukić; Hayssam Rashwan; Tomislav Šušnjar
Journal:  Langenbecks Arch Surg       Date:  2021-11-29       Impact factor: 3.445

7.  Could Harmonic Scalpel (Ultracision®) be considered the best device in surgical treatment of vulvar cancer of patients with implanted pace-maker? Proposal and rationale.

Authors:  Salvatore Gizzo; Alessandra Andrisani; Emanuele Ancona; Michela Quaranta; Amerigo Vitagliano; Marco Noventa; Giovanni Battista Nardelli; Guido Ambrosini
Journal:  Int J Clin Exp Med       Date:  2015-06-15

8.  Hybrid NOTES transvaginal cholecystectomy: operative and long-term results after 18 cases.

Authors:  Raffaele Pugliese; Antonello Forgione; Fabio Sansonna; Giovanni Carlo Ferrari; Stefano Di Lernia; Carmelo Magistro
Journal:  Langenbecks Arch Surg       Date:  2009-07-09       Impact factor: 3.445

9.  Comparison of lateral thermal damage of the human peritoneum using monopolar diathermy, Harmonic scalpel and LigaSure.

Authors:  Nikica Družijanić; Zenon Pogorelić; Zdravko Perko; Ivana Mrklić; Snježana Tomić
Journal:  Can J Surg       Date:  2012-10       Impact factor: 2.089

10.  Ureteric Injury due to the Use of LigaSure.

Authors:  Muazzam Tahir; William Gilkison
Journal:  Case Rep Urol       Date:  2013-09-19
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.