Beam hardening artefacts in computed tomography with photon counting, charge integrating and energy weighting detectors: a simulation study.

Polad M Shikhaliev1.   

Abstract

Photon counting x-ray imaging provides efficient rejection of the electronics noise, no pulse height (Swank) noise, less noise due to optimal photon energy weighting and the possibility of energy resolved image acquisition. These advantages apply also to CT when projection data are acquired using a photon counting detector. However, photon counting detectors assign a weighting factor of 1 to all detected photons whereas the weighting factor of a charge integrating detector is proportional to the energy of the detected photon. Therefore, data collected by photon counting and charge integrating detectors represent the 'hardening' of the photon beam passed through the object differently. This affects the beam hardening artefacts in the reconstructed CT images. This work represents the first comparative evaluation of the effect of photon counting, charge integrating and energy weighting photon detectors on beam hardening artefacts in CT. Beam hardening artefacts in CT images were evaluated for 20 cm and 14 cm diameter water cylinders with bone and low contrast inserts, at 120 kVp and 90 kVp x-ray tube voltages, respectively. It was shown that charge integrating results in 1.8% less beam hardening artefacts from bone inserts (i.e., CT numbers in the 'shadow' of the bone are less by 1.8% as compared to CT numbers over the periphery of the image), as compared to photon counting. However, optimal photon energy weighting, which provides highest SNR, results in 7.7% higher beam hardening artefacts from bone inserts as compared to photon counting. The magnitude of the 'cupping' artefacts was lower by 1% for charge integrating and higher by 6.1% for energy weighting acquisitions as compared to photon counting. Only the photon counting systems provide an accurate representation of the beam hardening effect due to its flat energy weighting. Because of their energy dependent weighting factors, the charge integrating and energy weighting systems do not provide accurate representation of the beam hardening effect.

Entities:  

Mesh:

Year:  2005        PMID: 16333157     DOI: 10.1088/0031-9155/50/24/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  24 in total

1.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

2.  Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.

Authors:  Han Chen; Cheng Xu; Mats Persson; Mats Danielsson
Journal:  J Med Imaging (Bellingham)       Date:  2015-11-06

3.  X-ray scatter correction method for dedicated breast computed tomography.

Authors:  Ioannis Sechopoulos
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

Review 4.  Vision 20/20: Single photon counting x-ray detectors in medical imaging.

Authors:  Katsuyuki Taguchi; Jan S Iwanczyk
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

5.  Characterization of the homogeneous tissue mixture approximation in breast imaging dosimetry.

Authors:  Ioannis Sechopoulos; Kristina Bliznakova; Xulei Qin; Baowei Fei; Steve Si Jia Feng
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

6.  Energy dispersive CdTe and CdZnTe detectors for spectral clinical CT and NDT applications.

Authors:  W C Barber; J C Wessel; E Nygard; J S Iwanczyk
Journal:  Nucl Instrum Methods Phys Res A       Date:  2015-06-01       Impact factor: 1.455

7.  Material separation in x-ray CT with energy resolved photon-counting detectors.

Authors:  Xiaolan Wang; Dirk Meier; Katsuyuki Taguchi; Douglas J Wagenaar; Bradley E Patt; Eric C Frey
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

8.  Photon-counting CT for simultaneous imaging of multiple contrast agents in the abdomen: An in vivo study.

Authors:  Rolf Symons; Bernhard Krauss; Pooyan Sahbaee; Tyler E Cork; Manu N Lakshmanan; David A Bluemke; Amir Pourmorteza
Journal:  Med Phys       Date:  2017-08-20       Impact factor: 4.071

9.  Cupping artifact correction and automated classification for high-resolution dedicated breast CT images.

Authors:  Xiaofeng Yang; Shengyong Wu; Ioannis Sechopoulos; Baowei Fei
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

10.  Medipix-based Spectral Micro-CT.

Authors:  Hengyong Yu; Qiong Xu; Peng He; James Bennett; Raja Amir; Bruce Dobbs; Xuanqin Mou; Biao Wei; Anthony Butler; Phillip Butler; Ge Wang
Journal:  CT Li Lun Yu Ying Yong Yan Jiu       Date:  2012-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.