High-resolution black-blood MRI of the carotid vessel wall using phased-array coils at 1.5 and 3 Tesla.

Seshasailaja Anumula1, Hee Kwon Song, Alexander C Wright, Felix W Wehrli.   

Abstract

RATIONALE AND
OBJECTIVES: The aim of this report is to investigate the magnetic field dependence of the signal-to-noise ratio (SNR) for carotid vessel wall magnetic resonance imaging using phased-array (PA) surface coils by comparing images obtained at 1.5 and 3 Tesla (T) and determine the extent to which the improved SNR at the higher field can be traded for improved spatial resolution.
MATERIALS AND METHODS: Two pairs of dual-element PA coils were constructed for operation at the two field strengths. The individual elements of each PA were matched to 50 Omega impedance on the neck and tuned at the respective frequencies. The coils were evaluated on a cylindrical phantom positioned with its axis parallel to the main field and the coils placed on either side of the phantom parallel to the sagittal plane. In vivo magnetic resonance images of the carotid arteries were obtained in five subjects at both field strengths with a fast spin-echo double-inversion black-blood pulse sequence with fat saturation. SNR was measured at both field strengths by using standard techniques.
RESULTS: At a depth corresponding to the average location of the carotid arteries in the study subjects, mean phantom SNR for the two coils was higher at 3 T by a factor of 2.5. The greater than linear increase is caused by only partial coil loading of these relatively small coils. The practically achievable average SNR gain in vivo was 2.1. The lower in vivo SNR gain is attributed to a reduction in T2 and prolongation of T1 at the higher field strength and, to a lesser extent, the requirement for a decreased refocusing pulse flip angle to operate within specific absorption rate limits. The superior SNR at 3 T appears to provide considerably improved vessel-wall delineation.
CONCLUSIONS: Carotid artery vessel-wall magnetic resonance imaging using PA surface coils provides a considerable increase in SNR when field strength is increased from 1.5 to 3 T. This increase can be traded for enhanced in-plane resolution.

Entities:  

Mesh:

Year:  2005        PMID: 16321740      PMCID: PMC1343477          DOI: 10.1016/j.acra.2005.08.009

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  16 in total

1.  7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images.

Authors:  J T Vaughan; M Garwood; C M Collins; W Liu; L DelaBarre; G Adriany; P Andersen; H Merkle; R Goebel; M B Smith; K Ugurbil
Journal:  Magn Reson Med       Date:  2001-07       Impact factor: 4.668

2.  Geometrical optimization of a phased array coil for high-resolution MR imaging of the carotid arteries.

Authors:  A Liffers; H H Quick; C U Herborn; H Ermert; M E Ladd
Journal:  Magn Reson Med       Date:  2003-08       Impact factor: 4.668

Review 3.  MR imaging at high magnetic fields.

Authors:  Masaya Takahashi; Hidemasa Uematsu; Hiroto Hatabu
Journal:  Eur J Radiol       Date:  2003-04       Impact factor: 3.528

4.  Nuclear relaxation of human brain gray and white matter: analysis of field dependence and implications for MRI.

Authors:  H W Fischer; P A Rinck; Y Van Haverbeke; R N Muller
Journal:  Magn Reson Med       Date:  1990-11       Impact factor: 4.668

5.  Fast selective black blood MR imaging.

Authors:  R R Edelman; D Chien; D Kim
Journal:  Radiology       Date:  1991-12       Impact factor: 11.105

6.  Surface coil phased arrays for high-resolution imaging of the carotid arteries.

Authors:  C E Hayes; C M Mathis; C Yuan
Journal:  J Magn Reson Imaging       Date:  1996 Jan-Feb       Impact factor: 4.813

7.  Measurement of signal intensities in the presence of noise in MR images.

Authors:  R M Henkelman
Journal:  Med Phys       Date:  1985 Mar-Apr       Impact factor: 4.071

8.  A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1-100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age.

Authors:  P A Bottomley; T H Foster; R E Argersinger; L M Pfeifer
Journal:  Med Phys       Date:  1984 Jul-Aug       Impact factor: 4.071

9.  RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imaging.

Authors:  P A Bottomley; E R Andrew
Journal:  Phys Med Biol       Date:  1978-07       Impact factor: 3.609

10.  Noninvasive in vivo human coronary artery lumen and wall imaging using black-blood magnetic resonance imaging.

Authors:  Z A Fayad; V Fuster; J T Fallon; T Jayasundera; S G Worthley; G Helft; J G Aguinaldo; J J Badimon; S K Sharma
Journal:  Circulation       Date:  2000-08-01       Impact factor: 29.690

View more
  10 in total

1.  Variability of carotid artery measurements on 3-Tesla MRI and its impact on sample size calculation for clinical research.

Authors:  Mushabbar A Syed; John N Oshinski; Charles Kitchen; Arshad Ali; Richard J Charnigo; Arshed A Quyyumi
Journal:  Int J Cardiovasc Imaging       Date:  2009-05-21       Impact factor: 2.357

2.  Carotid intima-media thickness and distensibility measured by MRI at 3 T versus high-resolution ultrasound.

Authors:  Andreas Harloff; Timo Zech; Alex Frydrychowicz; Martin Schumacher; Joachim Schöllhorn; Jürgen Hennig; Cornelius Weiller; Michael Markl
Journal:  Eur Radiol       Date:  2009-02-13       Impact factor: 5.315

3.  Signal-to-noise ratio increase in carotid atheroma MRI: a comparison of 1.5 and 3 T.

Authors:  V E Young; A J Patterson; E M Tunnicliffe; U Sadat; M J Graves; T Y Tang; A N Priest; P J Kirkpatrick; J H Gillard
Journal:  Br J Radiol       Date:  2012-01-31       Impact factor: 3.039

4.  High-field atherosclerotic plaque magnetic resonance imaging.

Authors:  Chun Yuan; Jinnan Wang; Niranjan Balu
Journal:  Neuroimaging Clin N Am       Date:  2012-05       Impact factor: 2.264

Review 5.  High-resolution magnetic resonance imaging: an emerging tool for evaluating intracranial arterial disease.

Authors:  Jeffrey D Bodle; Edward Feldmann; Richard H Swartz; Zoran Rumboldt; Truman Brown; Tanya N Turan
Journal:  Stroke       Date:  2012-11-29       Impact factor: 7.914

6.  Vessel-wall imaging and quantification of flow-mediated dilation using water-selective 3D SSFP-echo.

Authors:  Michael C Langham; Cheng Li; Erin K Englund; Erica N Chirico; Emile R Mohler; Thomas F Floyd; Felix W Wehrli
Journal:  J Cardiovasc Magn Reson       Date:  2013-10-30       Impact factor: 5.364

7.  Carotid Plaque Imaging: Strategies beyond Stenosis.

Authors:  Darshana Sanghvi; Manish Shrivastava
Journal:  Ann Indian Acad Neurol       Date:  2021-08-23       Impact factor: 1.383

Review 8.  Cardiovascular magnetic resonance in carotid atherosclerotic disease.

Authors:  Li Dong; William S Kerwin; Marina S Ferguson; Rui Li; Jinnan Wang; Huijun Chen; Gador Canton; Thomas S Hatsukami; Chun Yuan
Journal:  J Cardiovasc Magn Reson       Date:  2009-12-15       Impact factor: 5.364

9.  Accelerated fast spin-echo magnetic resonance imaging of the heart using a self-calibrated split-echo approach.

Authors:  Sabrina Klix; Fabian Hezel; Katharina Fuchs; Jan Ruff; Matthias A Dieringer; Thoralf Niendorf
Journal:  PLoS One       Date:  2014-04-11       Impact factor: 3.240

10.  Advantage in Bright-blood and Black-blood Magnetic Resonance Imaging with High-resolution for Analysis of Carotid Atherosclerotic Plaques.

Authors:  Mei Li; Wei-Jie Le; Xiao-Feng Tao; Ming-Hua Li; Yue-Hua Li; Nan Qu
Journal:  Chin Med J (Engl)       Date:  2015-09-20       Impact factor: 2.628

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.