Literature DB >> 16234456

Linear deposits on the surfaces of intraocular lenses implanted through a hexagonal cartridge which mimic scratches/cracks on the lenses.

G Kleinmann1, A L Marcovich, D J Apple, N Mamalis.   

Abstract

AIM: To describe unique linear deposits on the surface of posterior chamber intraocular lenses (IOLs) occurring after implantation through a hexagonal cartridge.
METHODS: Five ACR6D SE IOLs (Corneal Laboratories, Pringy, France) were injected/implanted through hexagonal cartridges. Two of these were injected into a petri dish and the remaining three were inserted into the evacuated capsular bag of cadaver eyes. In addition, three other ACR6D SE IOLs were injected into a petri dish through round cartridges. The latter served as controls.
RESULTS: All lenses that were injected/implanted through the hexagonal cartridges demonstrated linear deposits on the posterior surface of the IOL optic component. The IOLs that were injected through the round cartridges had no deposits. All the hexagonal cartridges showed signs of internal cracking. None of the round cartridges were cracked.
CONCLUSION: Implantation of at least one IOL model, the ACR6D SE IOL, through a hexagonal cartridge can result in linear deposits on the posterior optical surface of the IOL. The shape of the cartridge appears to be a significant factor in causing the depositions.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16234456      PMCID: PMC1772926          DOI: 10.1136/bjo.2005.071738

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  12 in total

1.  Cartridge cracks with different viscoelastic agents.

Authors:  R J Olson
Journal:  J Cataract Refract Surg       Date:  1999-04       Impact factor: 3.351

2.  Cartridge cracks during foldable intraocular lens insertion.

Authors:  R Rath; A D Singh; A Singh
Journal:  J Cataract Refract Surg       Date:  1999-03       Impact factor: 3.351

3.  Plastic exfoliations from the cartridge after implantation of a hydrophilic acrylic lens without viscoelastic material.

Authors:  Christoph W Faschinger
Journal:  J Cataract Refract Surg       Date:  2002-01       Impact factor: 3.351

4.  Surface abnormalities on hydrophilic acrylic intraocular lenses implanted by an injector.

Authors:  C W Faschinger
Journal:  J Cataract Refract Surg       Date:  2001-06       Impact factor: 3.351

5.  Cartridge cracks during foldable intraocular lens insertion.

Authors:  A D Singh; T Fang; R Rath
Journal:  J Cataract Refract Surg       Date:  1998-09       Impact factor: 3.351

6.  Forceps-induced scratching of a foldable acrylic intraocular lens.

Authors:  M P Vrabec; J C Syverud; C J Burgess
Journal:  Arch Ophthalmol       Date:  1996-06

7.  Cracking of acrylic intraocular lenses during capsular bag insertion.

Authors:  K H Carlson; D W Johnson
Journal:  Ophthalmic Surg Lasers       Date:  1995 Nov-Dec

8.  Scanning electron microscopic analysis of foldable acrylic and hydrogel intraocular lenses.

Authors:  T Kohnen; G Magdowski; D D Koch
Journal:  J Cataract Refract Surg       Date:  1996       Impact factor: 3.351

9.  Alterations to the AcrySof intraocular lens during folding.

Authors:  S Milazzo; P Turut; H Blin
Journal:  J Cataract Refract Surg       Date:  1996       Impact factor: 3.351

10.  Stress fractures after folding an acrylic intraocular lens.

Authors:  D R Pfister
Journal:  Am J Ophthalmol       Date:  1996-05       Impact factor: 5.258

View more
  4 in total

1.  Evaluation of a new soft tipped injector for the implantation of foldable intraocular lenses.

Authors:  Guy Kleinmann; David J Apple
Journal:  Br J Ophthalmol       Date:  2006-11-29       Impact factor: 4.638

2.  Optic surface changes in Intraocular lens scaffold: An ex vivo study.

Authors:  Dhivya Ashok Kumar; Amar Agarwal; Dhanalakshmi Panneerselvam; Josheela Sheebakalyani Gnanaprakasam
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-05-28       Impact factor: 3.117

3.  The course of surface deposits on a hydrophilic acrylic intraocular lens after implantation through a hexagonal cartridge.

Authors:  A L Marcovich; G Kleinmann; D Epstein; A Pollack
Journal:  Br J Ophthalmol       Date:  2006-07-19       Impact factor: 4.638

4.  Unusual markings on an intraocular lens postoperatively.

Authors:  Anjali Mehta
Journal:  Indian J Ophthalmol       Date:  2012 May-Jun       Impact factor: 1.848

  4 in total