Literature DB >> 8610804

Stress fractures after folding an acrylic intraocular lens.

D R Pfister1.   

Abstract

PURPOSE: To manage the complication of stress fractures in foldable acrylic intraocular lenses.
METHODS: During an implantation of a foldable acrylic intraocular lens (AcrySof, Alcon Laboratories, Fort Worth, Texas), the optic was noted to have developed paracentral stress fractures along the fold approximately 50% deep.
RESULTS: The lens was explanted and replaced without difficulty.
CONCLUSIONS: The observed complication can be avoided by recognizing that it might occur, wetting the lens before slowly folding it, and inspecting it carefully before insertion.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8610804     DOI: 10.1016/s0002-9394(14)75435-2

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  6 in total

1.  Evaluation of a new soft tipped injector for the implantation of foldable intraocular lenses.

Authors:  Guy Kleinmann; David J Apple
Journal:  Br J Ophthalmol       Date:  2006-11-29       Impact factor: 4.638

2.  Linear deposits on the surfaces of intraocular lenses implanted through a hexagonal cartridge which mimic scratches/cracks on the lenses.

Authors:  G Kleinmann; A L Marcovich; D J Apple; N Mamalis
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

3.  [Conspicuous intraocular lens].

Authors:  I Neuhann; B Lege; B Sommer; T Neuhann
Journal:  Ophthalmologe       Date:  2008-12       Impact factor: 1.059

4.  A practical and easy surgical technique for recovery of an incarcerated haptic.

Authors:  Yusuf Koçluk; Emine Alyamaç Sukgen; Burcu Kasım; Oğuzhan Saygılı
Journal:  Int Ophthalmol       Date:  2018-11-22       Impact factor: 2.031

5.  Optic surface changes in Intraocular lens scaffold: An ex vivo study.

Authors:  Dhivya Ashok Kumar; Amar Agarwal; Dhanalakshmi Panneerselvam; Josheela Sheebakalyani Gnanaprakasam
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-05-28       Impact factor: 3.117

6.  Delayed presentation of retained acrylic intraocular lens (IOL) fragment after uncomplicated cataract surgery.

Authors:  Cara E Capitena; Kevin Gamett; Mina B Pantcheva
Journal:  Am J Ophthalmol Case Rep       Date:  2016-05-13
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.