The course of surface deposits on a hydrophilic acrylic intraocular lens after implantation through a hexagonal cartridge.

A L Marcovich1, G Kleinmann, D Epstein, A Pollack.   

Abstract

AIM: To evaluate the outcome of surface deposits that occurred during implantation of hydrophylic acrylic intraocular lenses (IOLs) through a hexagonal cartridge.
METHODS: Surface deposits were observed on the posterior surface of the ACR6D SE IOLs that were injected through a hexagonal cartridge filled with sodium hyaluronate 1%. All the patients were examined 1 day, 1 week, 1 month, 6 months and 1 year postoperatively. The location of the deposits was recorded and photographed. The patients were questioned about blurred vision, glare or halos.
RESULTS: Linear or curly deposits were detected on the posterior surface of the IOL in six patients. In four patients, the deposits were peripheral and were observed 1 week postoperatively. In two patients, the deposits were noticed immediately after implantation. In one eye, they were misinterpreted as a crack in the IOL's optic and were left in the eye. In the second patient the deposits were removed immediately after implantation with forceps. The deposits that were left after implantation (five eyes) did not resolve during 1 year of follow-up. None of the eyes developed abnormal inflammatory reaction. In three eyes the best-corrected visual acuity (BCVA) was 6/6. In the other three eyes the BCVA was 6/12. None of the patients experienced any visual disturbance.
CONCLUSIONS: Implantation of the ACR6D SE IOL through a hexagonal cartridge can lead to the formation of deposits on the posterior surface of the lens. The deposits do not resolve and may resemble a crack in the IOL. The deposits left on the IOL had no clinical relevance in our patients.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16854825      PMCID: PMC1857429          DOI: 10.1136/bjo.2006.097857

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  8 in total

1.  Comparison of the effects of viscoelastic agents on clinical properties of the Unfolder lens injection system.

Authors:  T D Myers; R J Olson
Journal:  J Cataract Refract Surg       Date:  1999-07       Impact factor: 3.351

2.  Cartridge cracks with different viscoelastic agents.

Authors:  H B Dick; O Schwenn; E Fabian; T Neuhann; D Eisenmann
Journal:  J Cataract Refract Surg       Date:  1999-04       Impact factor: 3.351

3.  Plastic exfoliations from the cartridge after implantation of a hydrophilic acrylic lens without viscoelastic material.

Authors:  Christoph W Faschinger
Journal:  J Cataract Refract Surg       Date:  2002-01       Impact factor: 3.351

4.  Surface abnormalities on hydrophilic acrylic intraocular lenses implanted by an injector.

Authors:  C W Faschinger
Journal:  J Cataract Refract Surg       Date:  2001-06       Impact factor: 3.351

5.  Linear deposits on the surfaces of intraocular lenses implanted through a hexagonal cartridge which mimic scratches/cracks on the lenses.

Authors:  G Kleinmann; A L Marcovich; D J Apple; N Mamalis
Journal:  Br J Ophthalmol       Date:  2005-11       Impact factor: 4.638

6.  Cartridge cracks during foldable intraocular lens insertion.

Authors:  A D Singh; T Fang; R Rath
Journal:  J Cataract Refract Surg       Date:  1998-09       Impact factor: 3.351

7.  Intraoperative crystallization on the intraocular lens surface.

Authors:  R J Olson; K D Caldwell; A S Crandall; M K Jensen; S C Huang
Journal:  Am J Ophthalmol       Date:  1998-08       Impact factor: 5.258

8.  Crystallization on intraocular lens surfaces associated with the use of Healon GV.

Authors:  M K Jensen; A S Crandall; N Mamalis; R J Olson
Journal:  Arch Ophthalmol       Date:  1994-08
  8 in total
  1 in total

1.  [Refractive predictability and stability of three-piece versus single-piece intraocular lenses in patients with high axial myopia].

Authors:  A Frings; I Steinbrugger; E Wissiak; B Vidic; Y El-Shabrawi; N Ardjomand
Journal:  Ophthalmologe       Date:  2012-06       Impact factor: 1.059

  1 in total