Influence of phantom material and phantom size on radiographic film response in therapy photon beams.

Asa Palm1, Thomas LoSasso.   

Abstract

The energy dependence of radiographic film can introduce dosimetric errors when evaluating photon beams. The variation of the film response, which is attributed to the changing photon spectrum with depth and field size, has been the subject of numerous publications in recent years. However, these data show large unexplained differences in the magnitude of this variation among independent studies. To try to resolve this inconsistency, this study assesses the dependence of radiographic film response on phantom material and phantom size using film measurements and Monte Carlo calculations. The relative dose measured with film exposed to 6 MV x rays in various phantoms (polystyrene, acrylic, Solid Water, and water; the lateral phantom dimensions vary from 25 to 50 cm square; backscatter thickness varies from 10 to 30 cm) is compared with ion chamber measurements in water. Ranges of field size (5 x 5 to 40 x 40 cm2) and depth (dmax to 20 cm) are studied. For similar phantom and beam configurations, Monte Carlo techniques generate photon fluence spectra from which the relative film response is known from an earlier study. Results from film response measurements agree with those derived from Monte Carlo calculations within 3%. For small fields (< or = 10 x 10 cm2) and shallow depths (< or = 10 cm) the film response variation is small, less than 4%, for all phantoms. However, for larger field sizes and depths, the phantom material and phantom size have a greater influence on the magnitude of the film response. The variation of film response, over the ranges of field sizes and depths studied, is 50% in polystyrene compared with 30% in water. Film responses in Solid Water and water phantoms are similar; acrylic is between water and polystyrene. In polystyrene the variation of film response for a 50 cm square phantom is nearly twice that observed in a 25 cm square phantom. This study shows that differences in the configuration of the phantoms used for film dosimetry can explain much of the inconsistency for film response reported in the literature.

Entities:  

Mesh:

Year:  2005        PMID: 16193772     DOI: 10.1118/1.1949747

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Clinical implementation of volumetric intensity-modulated arc therapy (VMAT) with ERGO++.

Authors:  Dirk Wolff; Florian Stieler; Brigitte Hermann; Katharina Heim; Sven Clausen; Jens Fleckenstein; Martin Polednik; Volker Steil; Frederik Wenz; Frank Lohr
Journal:  Strahlenther Onkol       Date:  2010-04-26       Impact factor: 3.621

2.  Theoretical and empirical investigations of KCl : Eu2+ for nearly water-equivalent radiotherapy dosimetry.

Authors:  Yuanshui Zheng; Zhaohui Han; Joseph P Driewer; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

3.  Quasi-IMAT study with conventional equipment to show high plan quality with a single gantry arc.

Authors:  Judith Alvarez Moret; Oliver Kölbl; Ludwig Bogner
Journal:  Strahlenther Onkol       Date:  2009-02-18       Impact factor: 3.621

4.  Two-dimensional high spatial-resolution dosimeter using europium doped potassium chloride: a feasibility study.

Authors:  H Harold Li; Joseph P Driewer; Zhaohui Han; Daniel A Low; Deshan Yang; Zhiyan Xiao
Journal:  Phys Med Biol       Date:  2014-03-20       Impact factor: 3.609

5.  Quantitative megavoltage radiation therapy dosimetry using the storage phosphor KCl: Eu2+.

Authors:  Zhaohui Han; Joseph P Driewer; Yuanshui Zheng; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

6.  Analysis of variation in calibration curves for Kodak XV radiographic film using model-based parameters.

Authors:  Shu-Hui Hsu; Ravi Kulasekere; Peter L Roberson
Journal:  J Appl Clin Med Phys       Date:  2010-08-05       Impact factor: 2.102

7.  Soft tissue and water substitutes for megavoltage photon beams: An EGSnrc-based evaluation.

Authors:  Ambreen Aslam; Muhammad Basim Kakakhel; Shaukat Ali Shahid; Lubna Younas; Sobia Zareen
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.