Measurement accuracy in acetabular cup migration. A comparison of four radiologic methods versus roentgen stereophotogrammetric analysis.

T Ilchmann1, H Franzén, B Mjöberg, H Wingstrand.   

Abstract

Four different methods of radiologic evaluation of the acetabular component migration following total hip arthroplasty have been compared with roentgen stereophotogrammetry, a proven highly accurate method for studying early migration. In the Sutherland and Wetherell method the implant's position is measured with a pencil and a ruler from an ordinary pelvis radiograph. New reference lines of the Wetherell method are thought to be more accurate. The Sulzer and EBRA methods are computerized. In the Sulzer method prominent bone markers are digitized and used as reference points. In the EBRA method a system of tangents on prominent pelvis structure is digitized and used to detect radiographs with similar projection. The implant's position is calculated as the mean position of similar radiographs. The Sutherland, Wetherell, and Sulzer methods had almost the same accuracy, whereas the EBRA method was more accurate and could be used for pro- and retrospective studies in a large number of patients.

Entities:  

Mesh:

Year:  1992        PMID: 1613517     DOI: 10.1016/0883-5403(92)90004-a

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  11 in total

1.  Migration of the Duraloc cup after 5 years.

Authors:  Christoph Stihsen; Christof Pabinger; Roman Radl; Peter Rehak; Reinhard Windhager
Journal:  Int Orthop       Date:  2007-07-03       Impact factor: 3.075

2.  Accuracy of model-based RSA contour reduction in a typical clinical application.

Authors:  Christof Hurschler; Frank Seehaus; Judith Emmerich; Bart L Kaptein; Henning Windhagen
Journal:  Clin Orthop Relat Res       Date:  2008-05-29       Impact factor: 4.176

3.  A new tilt on pelvic radiographs: a pilot study.

Authors:  P J Richards; J M Pattison; J Belcher; R W Decann; Suzanne Anderson; C Wynn-Jones
Journal:  Skeletal Radiol       Date:  2008-06-20       Impact factor: 2.199

4.  Is Plain Anterior-Posterior Radiograph of the Pelvis Adequate for Assessment of Radiographic Implant Migration Evaluation in Total Hip Arthroplasty?

Authors:  Alexander Cm Chong; Lisa N MacFadden; Bruce E Piatt; Benjamin C Noonan
Journal:  Iowa Orthop J       Date:  2020

5.  A prospective study of the migration of two acetabular components. PCA versus RM cups.

Authors:  M Krismer; M Fischer; P Mayrhofer; F Stöckl; C Bittner; C Trojer; B Stöckl
Journal:  Int Orthop       Date:  1994-02       Impact factor: 3.075

6.  Clinical evaluation of a novel press-fit acetabular cup using "Ein-Bild-Roentgen-Analysis" (EBRA): A positive short-term prognosis.

Authors:  B A Ishaque; J Wieczorek; C A Fonseca Ulloa; J B Seeger; G A Ahmed; M Rickert; A Jahnke
Journal:  J Orthop       Date:  2020-03-30

7.  The final follow-up plain radiograph is sufficient for clinical evaluation of polyethylene wear in total hip arthroplasty. A study of validity and reliability.

Authors:  Maiken Stilling; Kristian Larsen; Niels T Andersen; Kjeld Søballe; Søren Kold; Ole Rahbek
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

8.  Good mid-term results with the trident peripheral self-locking cup: a clinical evaluation and migration measurement with EBRA.

Authors:  Dietmar Dammerer; Philipp Blum; David Putzer; Andreas Tscholl; Michael C Liebensteiner; Martin Thaler
Journal:  Arch Orthop Trauma Surg       Date:  2020-11-09       Impact factor: 3.067

9.  Biomechanical Evaluation of the Accuracy in Radiographic Assessment of Femoral Component Migration Measurement after Total Hip Arthroplasty.

Authors:  Mikayla Forness; Zachary J Podoll; Benjamin C Noonan; Alexander C M Chong
Journal:  Kans J Med       Date:  2020-04-17

10.  Two-year radiologic assessment of the Pinnacle cup-a migration analysis with EBRA.

Authors:  Dietmar Dammerer; Alexander Ruzicka; Philipp Blum; David Putzer; Maximilian Liebsch; Julian Lair; Martin Thaler
Journal:  Arch Orthop Trauma Surg       Date:  2020-10-30       Impact factor: 3.067