B J Conway1, J L McCrohan, R G Antonsen, F G Rueter, R J Slayton, O H Suleiman.
Abstract
In 1990, as part of the Nationwide Evaluation of X-ray Trends (NEXT) program, 252 computed tomographic (CT) systems were evaluated to measure radiation doses associated with standard head CT in adults. The multiple-scan average dose (MSAD) was used as the dose descriptor. For most of the systems, the MSAD at the midpoint on the central axis of a standard dosimetry phantom was between 34 and 55 mGy. Doses were as high as 140 mGy, and dose sometimes varied by a factor of two or more for identical CT units. This range indicates that dose can potentially be reduced by careful selection of standard CT techniques. Users of CT systems should be aware of radiation dose delivered with CT, dose ranges associated with different systems, and doses delivered with their particular unit, which requires that dose performance of CT systems be assessed by means of a protocol that allows comparison of data collected for identical and/or different units.Mesh:
Year: 1992 PMID: 1609069 DOI: 10.1148/radiology.184.1.1609069
Source DB: PubMed Journal: Radiology ISSN: 0033-8419 Impact factor: 11.105