Literature DB >> 15920679

A preliminary prospective study of the usefulness of a magnetic endoscope locating device during colonoscopy.

S Ambardar1, T D Arnell, R L Whelan, A Nihalani, K A Forde.   

Abstract

BACKGROUND: Although magnetic endoscope imaging of the colonoscope via the Endoscope Positioning Detecting Unit (EPDU) has been studied to some extent in Europe, its application in the United States has been limited. The purposes of this study were to determine whether the technique enabled for accurate localization of the lesion and to determine if and how the device facilitated scope insertion and completion of the colonoscopic exam.
METHODS: Outpatient colonoscopies using the EPDU were performed by three experienced surgical endoscopists over a 5-month period. A specialized scope with electromagnetic coils or a regular scope with a magnetic probe insert in the instrument channel was used for the duration of the examination to identify loops and localize pathology.
RESULTS: A total of 80 colonoscopies were performed with the device. In two patients, the probe insert was removed prior to completion of the procedure; thus, the total number of examinations included in the study was 78. The EPDU was used in conjunction with transillumination to estimate the location of polyps or cancers in the 33 patients (42%) in whom such lesions were found. In the four patients who subsequently underwent operation, the lesion's location as estimated by EPDU was verified. In regard to the usefulness of the device during insertion, the EPDU led to the discovery of loops and to the application of pressure that resulted in prompt completion of the examination in 28% of cases (deemed most useful). In 33% of cases, the device identified loops and led to the application of abdominal wall pressure and early position changes, thus facilitating the examination; however it did not lead to its immediate or rapid completion. In 39% of cases, the device was not required or used for insertion due to the simple nature of the examination.
CONCLUSIONS: The EPDU was accurate in estimating lesion location. The device also holds promise as an aid in the completion of difficult exams (about 30% of cases in this study).

Entities:  

Mesh:

Year:  2005        PMID: 15920679     DOI: 10.1007/s00464-004-8948-0

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  16 in total

1.  Effect of magnetic endoscope imaging on colonoscopy performance: a randomised controlled trial.

Authors:  S G Shah; J C Brooker; C B Williams; C Thapar; B P Saunders
Journal:  Lancet       Date:  2000-11-18       Impact factor: 79.321

2.  Use of video and magnetic endoscope imaging for rating competence at colonoscopy: validation of a measurement tool.

Authors:  Syed G Shah; Siwan Thomas-Gibson; Jim C Brooker; Noriko Suzuki; Christopher B Williams; Catherine Thapar; Brian P Saunders
Journal:  Gastrointest Endosc       Date:  2002-10       Impact factor: 9.427

3.  Real-time magnetic three-dimensional imaging of flexible endoscopy.

Authors:  C B Williams; B P Saunders; G D Bell; J S Bladen; D F Gillies; C M Guy
Journal:  Gastrointest Endosc Clin N Am       Date:  1997-07

4.  Air contrast fluoroscopy as a diagnostic aid during colonoscopy.

Authors:  W Dickey; K G Porter
Journal:  Endoscopy       Date:  1996-10       Impact factor: 10.093

5.  Fluoroscopy. A valuable ally during difficult colonoscopy.

Authors:  W C Cirocco; L C Rusin
Journal:  Surg Endosc       Date:  1996-11       Impact factor: 4.584

6.  Magnetic endoscope imaging: a new technique for localizing colonic lesions.

Authors:  S G Shah; H J Pearson; S Moss; E Kweka; P K Jalal; B P Saunders
Journal:  Endoscopy       Date:  2002-11       Impact factor: 10.093

7.  Magnetic imaging of colonoscopy: an audit of looping, accuracy and ancillary maneuvers.

Authors:  S G Shah; B P Saunders; J C Brooker; C B Williams
Journal:  Gastrointest Endosc       Date:  2000-07       Impact factor: 9.427

8.  Patient pain during colonoscopy: an analysis using real-time magnetic endoscope imaging.

Authors:  S G Shah; J C Brooker; C Thapar; C B Williams; B P Saunders
Journal:  Endoscopy       Date:  2002-06       Impact factor: 10.093

9.  First clinical results with a real time, electronic imager as an aid to colonoscopy.

Authors:  B P Saunders; G D Bell; C B Williams; J S Bladen; A P Anderson
Journal:  Gut       Date:  1995-06       Impact factor: 23.059

10.  The variable stiffness colonoscope: assessment of efficacy by magnetic endoscope imaging.

Authors:  Syed G Shah; Jim C Brooker; Christopher B Williams; Catherine Thapar; Noriko Suzuki; Brian P Saunders
Journal:  Gastrointest Endosc       Date:  2002-08       Impact factor: 9.427

View more
  7 in total

1.  Determining scope position during colonoscopy without use of ionizing radiation or magnetic imaging: the enhanced mapping ability of the NeoGuide Endoscopy System.

Authors:  Johannes Striegel; Ralf Jakobs; Jacques Van Dam; Uwe Weickert; Jürgen F Riemann; Axel Eickhoff
Journal:  Surg Endosc       Date:  2010-08-21       Impact factor: 4.584

2.  Early use of magnetic endoscopic imaging by novice colonoscopists: improved performance without increase in workload.

Authors:  Sylvain Coderre; John Anderson; Remy Rikers; Paul Dunckley; Karen Holbrook; Kevin McLaughlin
Journal:  Can J Gastroenterol       Date:  2010-12       Impact factor: 3.522

Review 3.  Mechanical analysis of insertion problems and pain during colonoscopy: why highly skill-dependent colonoscopy routines are necessary in the first place... and how they may be avoided.

Authors:  Arjo J Loeve; Paul Fockens; Paul Breedveld
Journal:  Can J Gastroenterol       Date:  2013       Impact factor: 3.522

4.  Image Registered Gastroscopic Ultrasound (IRGUS) in human subjects: a pilot study to assess feasibility.

Authors:  K L Obstein; R S Estépar; J Jayender; V D Patil; I S Spofford; M B Ryan; B I Lengyel; R Shams; K G Vosburgh; C C Thompson
Journal:  Endoscopy       Date:  2011-03-21       Impact factor: 10.093

5.  Evaluation of magnetic scope navigation in screening endoscopic examination of colorectal cancer.

Authors:  Miroslaw Szura; Krzysztof Bucki; Andrzej Matyja; Jan Kulig
Journal:  Surg Endosc       Date:  2011-09-30       Impact factor: 4.584

6.  Use of an electromagnetic colonoscope to assess maneuvers associated with cecal intubation.

Authors:  Russell I Heigh; John K DiBaise; James A Prechel; Billie J Horn; Sarah San Miguel; Evelyn G Heigh; Jonathan A Leighton; Cynthia J Edgelow; David E Fleischer
Journal:  BMC Gastroenterol       Date:  2009-04-09       Impact factor: 3.067

7.  Accuracy of preoperative tumor localization in large bowel using 3D magnetic endoscopic imaging: randomized clinical trial.

Authors:  Miroslaw Szura; Artur Pasternak; Rafal Solecki; Maciej Matyja; Antoni Szczepanik; Andrzej Matyja
Journal:  Surg Endosc       Date:  2016-08-29       Impact factor: 4.584

  7 in total