Literature DB >> 15777330

Implant design and intraosseous stability of immediately placed implants: a human cadaver study.

Murat Akkocaoglu1, Serdar Uysal, Ibrahim Tekdemir, Kivanc Akca, Murat Cavit Cehreli.   

Abstract

OBJECTIVE: The objective of this study was to explore effects of implant macrodesign and diameter on initial intraosseous stability and interface mechanical properties of immediately placed implants. MATERIAL AND
METHOD: Mandibular premolars of four fresh-frozen human cadavers were extracted. Ø 4.1/4.8 mm ITI TE, Ø 4.1 and 4.8 mm solid screw synOcta ITI implants were placed into freshly prepared extraction sockets. Resonance frequency analysis was conducted to quantify primary implant stability quotient (ISQ). Installation torque value (ITV) and removal torque value (RTV) of the implants were measured using a custom-made strain-gauged torque wrench connected to a data acquisition system at a sample rate of 10,000 Hz. The vertical defect depth around the collar of each implant was measured directly by an endodontic spreader. The bone-implant contact was determined in digitalized images of periapical radiographs and expressed as percentage bone contact.
RESULTS: The ISQ values of the TE implant was higher than the Ø 4.1 mm implant (P<0.01), and comparable with the Ø 4.8 mm implants (P>0.05). ITVs and RTVs of TE and Ø 4.8 mm implants were higher than the Ø 4.1 mm implant, although the differences between groups were statistically insignificant (P>0.05). The vertical defect depths around all types of implants were similar. In the radiographic analyses, percentage bone-implant contact of the TE and Ø 4.8 mm implants were comparable at the marginal bone region and both were higher than that of the Ø 4.1 mm ITI implant. Nonparametric correlations between groups revealed a significant correlation between ITV and RTV (r=0.838; P<0.001), but not between ISQ values and ITVs and RTVs (P>0.05).
CONCLUSION: Immediately placed ITI TE implant leads to initial intraosseous stability and interface mechanical properties comparable with a wide diameter implant.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15777330     DOI: 10.1111/j.1600-0501.2004.01099.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  21 in total

1.  Relation between insertion torque and bone-implant contact percentage: an artificial bone study.

Authors:  Cheng Liu; Ming-Tzu Tsai; Heng-Li Huang; Michael Yuan-Chien Chen; Jui-Ting Hsu; Kuo-Chih Su; Chih-Han Chang; Aaron Yu-Jen Wu
Journal:  Clin Oral Investig       Date:  2012-01-10       Impact factor: 3.573

2.  The effect of implant shape and bone preparation on primary stability.

Authors:  Sang-Hyun Moon; Heung-Sik Um; Jae-Kwan Lee; Beom-Seok Chang; Min-Ku Lee
Journal:  J Periodontal Implant Sci       Date:  2010-10-31       Impact factor: 2.614

3.  Towards the limit of quantifying low-amplitude strains on bone and in coagulum around immediately loaded oral implants in extraction sockets.

Authors:  Murat Cavit Cehreli; Ayhan Comert; Murat Akkocaoglu; Ibrahim Tekdemir; Kivanc Akca
Journal:  Med Biol Eng Comput       Date:  2006-03       Impact factor: 2.602

4.  Clinical study on the primary stability of two dental implant systems with resonance frequency analysis.

Authors:  Annette Rabel; Steffen Gerhard Köhler; Andrea Maria Schmidt-Westhausen
Journal:  Clin Oral Investig       Date:  2007-03-31       Impact factor: 3.573

5.  Bone strains around apically free versus grafted implants in the posterior maxilla of human cadavers.

Authors:  Murat Cavit Cehreli; Murat Akkocaoglu; Ayhan Comert; Ibrahim Tekdemir; Kivanc Akca
Journal:  Med Biol Eng Comput       Date:  2007-03-06       Impact factor: 2.602

6.  Biomechanical analysis of alveolar bone stress around implants with different thread designs and pitches in the mandibular molar area.

Authors:  Ting-Hsun Lan; Je-Kang Du; Chin-Yun Pan; Huey-Er Lee; Wei-Hao Chung
Journal:  Clin Oral Investig       Date:  2011-02-08       Impact factor: 3.573

7.  Template-guided vs. non-guided drilling in site preparation of dental implants.

Authors:  Uta Scherer; Marcus Stoetzer; Martin Ruecker; Nils-Claudius Gellrich; Constantin von See
Journal:  Clin Oral Investig       Date:  2014-10-30       Impact factor: 3.573

8.  3D full-field strain in bone-implant and bone-tooth constructs and their morphological influential factors.

Authors:  Yuxiao Zhou; Chujie Gong; Mehran Hossaini-Zadeh; Jing Du
Journal:  J Mech Behav Biomed Mater       Date:  2020-05-19

9.  Biomechanics and strain mapping in bone as related to immediately-loaded dental implants.

Authors:  Jing Du; Ji-Hyun Lee; Andrew T Jang; Allen Gu; Mehran Hossaini-Zadeh; Richard Prevost; Donald A Curtis; Sunita P Ho
Journal:  J Biomech       Date:  2015-06-19       Impact factor: 2.712

10.  Numerical simulation of in vivo intraosseous torsional failure of a hollow-screw oral implant.

Authors:  Murat Cehreli; Murat Akkocaoglu; Kivanc Akca
Journal:  Head Face Med       Date:  2006-11-04       Impact factor: 2.151

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.