Effect of implant angulation upon retention of overdenture attachments.

Michael P Gulizio1, John R Agar, J Robert Kelly, Thomas D Taylor.   

Abstract

INTRODUCTION: Overdentures supported and retained by endosteal implants depend upon mechanical components to provide retention. Ball attachments are frequently described because of simplicity and low cost, but retentive capacity of these components may be altered by a lack of implant parallelism.
PURPOSE: The aim of this in vitro study was to investigate the retention of gold and titanium overdenture attachments when placed on ball abutments positioned off-axis. METHODS AND MATERIALS: Four ball abutments were hand-tightened onto ITI dental implants and placed in an aluminum fixture that allowed positioning of the implants at 0 degrees, 10 degrees, 20 degrees, and 30 degrees from a vertical reference axis. Gold and titanium matrices were then coupled to the ball abutments at various angles and then subjected to pull tests at a rate of 2 mm/second; the peak loads of release (maximum dislodging forces) were recorded and subjected to statistical analyses. A balanced and randomized factorial experimental design testing procedure was implemented.
RESULTS: Statistically significant differences in retention of gold matrices were noted when ball abutments were positioned at 20 degrees and 30 degrees , but not at 0 degrees and 10 degrees. Statistically significant differences were noted among the titanium matrices employed for the testing procedure, as well as for the 4 ball abutments tested. Angle was not a factor affecting retention for titanium matrices.
CONCLUSIONS: (1) The gold matrices employed for the testing procedures exhibited consistent values in retention compared to titanium matrices, which exhibited large variability in retention. (2) Angle had an effect upon the retention of gold matrices, but not for titanium matrices. Copyright 2005 by The American College of Prosthodontists

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15733129     DOI: 10.1111/j.1532-849X.2005.00005.x

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  10 in total

1.  Stud attachments for the mandibular implant-retained overdentures: Prosthetic complications. A literature review.

Authors:  Elie E Daou
Journal:  Saudi Dent J       Date:  2013-02-06

2.  A finite element analysis of stress distribution in the bone, around the implant supporting a mandibular overdenture with ball/o ring and magnetic attachment.

Authors:  Jins John; V Rangarajan; Ravindra C Savadi; K S Satheesh Kumar; Preeti Satheesh Kumar
Journal:  J Indian Prosthodont Soc       Date:  2012-03-09

3.  The influences of implant angulations in one and two directions on the retentive properties of overdenture attachments: an in vitro study.

Authors:  Parsa Atashrazm; Mohammad Hossein Dashti; Leila Zamani Alavijeh; Samira Azarmaeh; Sayena Mahdizadeh
Journal:  J Indian Prosthodont Soc       Date:  2013-04-12

Review 4.  Angulated implants: an alternative to bone augmentation and sinus lift procedure: systematic review.

Authors:  Nikhil Asawa; Nilesh Bulbule; Dilip Kakade; Riddhi Shah
Journal:  J Clin Diagn Res       Date:  2015-03-01

5.  The influence of saliva pH value on the retention and durability of bar-clip attachments.

Authors:  Antonio Sergio Silva; Carlos Aroso; Raul Ustrell; Ana Cristina Braga; Jose Manuel Mendes; Tomas Escuin
Journal:  J Adv Prosthodont       Date:  2015-02-17       Impact factor: 1.904

6.  The effect of inter-implant distance on retention and resistance to dislodging forces for mandibular implant-tissue-supported overdentures.

Authors:  Farhad Tabatabaian; Abolfazl Saboury; Zahra Sadat Sobhani; Vicki C Petropoulos
Journal:  J Dent (Tehran)       Date:  2014-09-30

7.  Influence of Number of Implants and Attachment Type on Stress Distribution in Mandibular Implant-Retained Overdentures: Finite Element Analysis.

Authors:  Mohamed I El-Anwar; Eman A El-Taftazany; Hamdy A Hamed; Mohamed A Abd ElHay
Journal:  Open Access Maced J Med Sci       Date:  2017-03-22

8.  Stress analysis of mandibular implant overdentures retained with one, two, or four ball attachments: A finite element study.

Authors:  Mohamed M El-Zawahry; Eman M Ibraheem; Mohammad Zakaria Nassani; Sahar A Ghorab; Mohamed I El-Anwar
Journal:  Dent Res J (Isfahan)       Date:  2018 Nov-Dec

Review 9.  Mandibular implant-supported overdenture: A systematic review and meta-analysis for optimum selection of attachment system.

Authors:  Priyanka Vaibhav Sutariya; Hitendra Mohanlal Shah; Surbhi Dipakbhai Patel; Hemil Hitesh Upadhyay; Mansoorkhan Rafikahmed Pathan; Rutu Paresh Shah
Journal:  J Indian Prosthodont Soc       Date:  2021 Oct-Dec

10.  Strains around distally inclined implants retaining mandibular overdentures with Locator attachments: an in vitro study.

Authors:  Moustafa Abdou Elsyad; Fathi Abo Setta; Ahmed Samir Khirallah
Journal:  J Adv Prosthodont       Date:  2016-04-21       Impact factor: 1.904

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.