Mohamed I El-Anwar1, Eman A El-Taftazany2, Hamdy A Hamed3, Mohamed A Abd ElHay2.
Abstract
AIM: This study aimed to compare the stresses generated by using two or four root form dental implants supporting mandibular overdentures that were retained with ball and locator attachments.Entities:
Keywords: Attachment; Finite element analysis; Implant; Overdenture
Year: 2017 PMID: 28507636 PMCID: PMC5420782 DOI: 10.3889/oamjms.2017.047
Source DB: PubMed Journal: Open Access Maced J Med Sci ISSN: 1857-9655
Figure 1Implant complex with (A) Ball, and (B) locator models on Autodesk Inventor screen
Material properties of used in the finite element model
| Material | Young’s modules [GPa] | Posison’s ratio |
|---|---|---|
| Overdenture [ | 2.70 | 0.35 |
| Mucosa [ | 0.01 | 0.40 |
| nylon ring (Cap) [ | 0.35 | 0.40 |
| Implant complex [ | 110.00 | 0.35 |
| Cortical bone [ | 13.70 | 0.30 |
| Cancellous bone [ | 1.37 | 0.30 |
Number of nodes and elements in all meshed components
| 2x Ball Attachment | 4x Ball Attachment | 2x Locator Attachment | 4x Locator Attachment | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Component | Elements | Nodes | Elements | Nodes | Elements | Nodes | Elements | Nodes | |
| Overdenture | 8.219. | 2.272. | 6.228. | 1.724. | 9.658. | 2.796. | 7.287. | 2.056. | |
| Mucosa | 13.234. | 40324. | 7.326. | 2.556. | 13.574. | 4.410. | 7.470. | 2.594. | |
| nylon ring (Cap) | 40.427. | 9.391. | 27.031. | 5.825. | 2.603. | 4.788. | 10.264. | 2.388. | |
| Implant complex | 493.369. | 94.650. | 244.131. | 46.893. | 516.713. | 96.216. | 317.994. | 56.92. | |
| Cortical bone | 5.870. | 2.177. | 4.897. | 1.773. | 5.844. | 2.177. | 4.892. | 1.769. | |
| Spongy bone | 59.199. | 16.428. | 29.451. | 8.313. | 58.867. | 16.356. | 29.486. | 8.307. | |
Figure 2Meshed components of the used model(s)
Figure 3Comparison between using two and four ball & socket attachments
Figure 4Comparison between using two and four locator attachments
Figure 5Overdenture total deformation under vertical load in cases of using (a) two and (b) four ball & socket attachment
Figure 6Mucosa maximum compressive stress under oblique load in cases of using (a) two ball & socket and (b) two locator attachments
Figure 7Ball & Socket attachment Von Mises stress and lateral deformation “Ux” under oblique load in case of using four ball & socket attachments
Figure 8Locator attachment Von Mises stresses under vertical load in case of using four implants