Literature DB >> 24203851

Heating and safety of a new MR-compatible guidewire prototype versus a standard nitinol guidewire.

Malgorzata Wolska-Krawczyk1, Martin A Rube, Erwin Immel, Andreas Melzer, Arno Buecker.   

Abstract

Our purpose in this study was to examine heating of nitinol and polyetheretherketone (PEEK) guidewires during near-real-time MR imaging in an artificial vascular model an "aorta phantom". The first 100 cm of the nitinol- and PEEK-based guidewires both 145 × 0.08 cm were immersed in a saline-filled aorta phantom. The probes of a fiber-optic thermometer were positioned at the tips of both wires. Balanced steady-state free precession (bSSFP) [TE 1.6 ms; TR 3.5 ms; flip angle (FA) 60°; field of view (FOV) 40 cm; matrix 256 × 256; specific absorption rate (SAR); 1.15 Watt (W)/kg] and spoiled gradient-echo (SPGR) (TE 1.8 ms; TR 60 ms; FA 60°; FOV 40 cm; matrix 256 × 256; SAR 1.15 W/kg) pulse sequences were acquired in a 1.5-T MR scanner with use of an 8-channel array coil. Temperatures were recorded while the phantom was placed centrally in the bore of a MR scanner and in an off-center position (x = 24 cm, y = -5 cm, z = -10/10 cm). The temperature of the nitinol guidewire increased by 0.3 °C (center) and 1.1 °C (off-center position) with use of the bSSFP and by 9.6 and 13 °C (off-center position) with use of the SPGR sequence. Only minor temperature changes up to a maximum of 0.4 °C were observed with the MR-compatible PEEK guidewire when any position or sequence was applied. The PEEK guidewire showed substantially lower heating as compared to the nitinol guidewire in near-real-time imaging sequences in a phantom.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24203851     DOI: 10.1007/s12194-013-0236-z

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  22 in total

1.  Heating around intravascular guidewires by resonating RF waves.

Authors:  M K Konings; L W Bartels; H F Smits; C J Bakker
Journal:  J Magn Reson Imaging       Date:  2000-07       Impact factor: 4.813

Review 2.  Safety of MRI-guided vascular interventions.

Authors:  Arno Buecker
Journal:  Minim Invasive Ther Allied Technol       Date:  2006       Impact factor: 2.442

3.  Use of a nonmetallic guide wire for magnetic resonance-guided coronary artery catheterization.

Authors:  Arno Buecker; Elmar Spuentrup; Thomas Schmitz-Rode; Sylvia Kinzel; Jochen Pfeffer; Christian Hohl; Joop J van Vaals; Rolf W Günther
Journal:  Invest Radiol       Date:  2004-11       Impact factor: 6.016

Review 4.  Radiofrequency safety for interventional MRI procedures.

Authors:  Ergin Atalar
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

5.  MRI safety of medical devices and procedures.

Authors:  Andreas Melzer
Journal:  Minim Invasive Ther Allied Technol       Date:  2006       Impact factor: 2.442

6.  An MR guidewire based on micropultruded fiber-reinforced material.

Authors:  Sascha Krueger; Sebastian Schmitz; Steffen Weiss; Daniel Wirtz; Marita Linssen; Heinz Schade; Nils Kraemer; Elmar Spuentrup; Gabriele Krombach; Arno Buecker
Journal:  Magn Reson Med       Date:  2008-11       Impact factor: 4.668

7.  Thermoregulatory consequences of cardiovascular impairment during NMR imaging in warm/humid environments.

Authors:  E R Adair; L G Berglund
Journal:  Magn Reson Imaging       Date:  1989 Jan-Feb       Impact factor: 2.546

8.  Experimental and numerical assessment of MRI-induced temperature change and SAR distributions in phantoms and in vivo.

Authors:  Sukhoon Oh; Andrew G Webb; Thomas Neuberger; BuSik Park; Christopher M Collins
Journal:  Magn Reson Med       Date:  2010-01       Impact factor: 4.668

9.  Feasibility of real-time magnetic resonance-guided angioplasty and stenting of renal arteries in vitro and in Swine, using a new polyetheretherketone-based magnetic resonance-compatible guidewire.

Authors:  Sebastian Kos; Rolf Huegli; Eugen Hofmann; Harald H Quick; Hilmar Kuehl; Stephanie Aker; Gernot M Kaiser; Paul J Borm; Augustinus L Jacob; Deniz Bilecen
Journal:  Invest Radiol       Date:  2009-04       Impact factor: 6.016

10.  Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations.

Authors:  Eugenio Mattei; Michele Triventi; Giovanni Calcagnini; Federica Censi; Wolfgang Kainz; Gonzalo Mendoza; Howard I Bassen; Pietro Bartolini
Journal:  Biomed Eng Online       Date:  2008-03-03       Impact factor: 2.819

View more
  4 in total

1.  Radial MRI with variable echo times: reducing the orientation dependency of susceptibility artifacts of an MR-safe guidewire.

Authors:  Katharina E Schleicher; Michael Bock; Klaus Düring; Stefan Kroboth; Axel J Krafft
Journal:  MAGMA       Date:  2017-08-02       Impact factor: 2.310

2.  Preclinical feasibility of a technology framework for MRI-guided iliac angioplasty.

Authors:  Martin A Rube; Fabiola Fernandez-Gutierrez; Benjamin F Cox; Andrew B Holbrook; J Graeme Houston; Richard D White; Helen McLeod; Mahsa Fatahi; Andreas Melzer
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-08-08       Impact factor: 2.924

3.  Segmented nitinol guidewires with stiffness-matched connectors for cardiovascular magnetic resonance catheterization: preserved mechanical performance and freedom from heating.

Authors:  Burcu Basar; Toby Rogers; Kanishka Ratnayaka; Adrienne E Campbell-Washburn; Jonathan R Mazal; William H Schenke; Merdim Sonmez; Anthony Z Faranesh; Robert J Lederman; Ozgur Kocaturk
Journal:  J Cardiovasc Magn Reson       Date:  2015-11-30       Impact factor: 5.364

4.  Invasive cardiovascular magnetic resonance (iCMR) for diagnostic right and left heart catheterization using an MR-conditional guidewire and passive visualization in congenital heart disease.

Authors:  Surendranath R Veeram Reddy; Yousef Arar; Riad Abou Zahr; Vasu Gooty; Jennifer Hernandez; Amanda Potersnak; Phillip Douglas; Zachary Blair; Joshua S Greer; Sébastien Roujol; Mari Nieves Velasco Forte; Gerald Greil; Alan W Nugent; Tarique Hussain
Journal:  J Cardiovasc Magn Reson       Date:  2020-03-26       Impact factor: 5.364

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.