Literature DB >> 2913112

Determination of cardiac output in critically ill patients by dual beam Doppler echocardiography.

D S Looyenga1, P R Liebson, R C Bone, R A Balk, J V Messer.   

Abstract

Recent technology in Doppler echocardiography has produced a dual beam Doppler instrument that is capable of insonating the total cross-sectional area of the ascending aorta. The purpose of this study was to evaluate the accuracy of this instrument in measuring cardiac output in critically ill patients by comparing results with those of the thermodilution-derived cardiac output. A technically adequate Doppler cardiac output measurement was attained in 71 (91%) of 78 patients. The range of thermodilution-derived cardiac output measurements was from 1.58 to 11.70 liters/min. To maximize thermodilution cardiac output reliability, several measurements were made for each patient. Those patients in whom the difference between the highest and lowest measurement varied by less than 10% from the averaged results were accepted into the 50 patient study. There was significant correlation between dual beam Doppler- and thermodilution-derived cardiac output (r = 0.96, SEE = 0.55 liters/min, p less than 0.0001). This study demonstrates that dual beam Doppler ultrasound is a promising noninvasive method of measuring cardiac output in the critically ill patient.

Entities:  

Mesh:

Year:  1989        PMID: 2913112     DOI: 10.1016/0735-1097(89)90509-3

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  7 in total

1.  Effect of meal induced splanchnic arterial vasodilatation on renal arterial haemodynamics in normal subjects and patients with cirrhosis.

Authors:  T Iwao; K Oho; R Nakano; M Yamawaki; T Sakai; M Sato; Y Miyamoto; A Toyonaga; K Tanikawa
Journal:  Gut       Date:  1998-12       Impact factor: 23.059

2.  Cardiac output variations in supine resting subjects during head-out cold water immersion.

Authors:  P Vogelaere; G Deklunder; J Lecroart
Journal:  Int J Biometeorol       Date:  1995-08       Impact factor: 3.787

3.  Influence of nebivolol on the cardiovascular hemodynamics during postural changes and isometric exercise.

Authors:  D Duprez; R Lefebvre; T De Backer; P De Sutter; J Trouerbach; D L Clement
Journal:  Cardiovasc Drugs Ther       Date:  1991-08       Impact factor: 3.727

4.  Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants.

Authors:  C F Wippermann; D Schranz; R Huth; F Zepp; H Oelert; B K Jüngst
Journal:  Br Heart J       Date:  1992-02

5.  The interaction of the calcium antagonist RO 40-5967 with digoxin.

Authors:  M Siepmann; C Kleinbloesem; W Kirch
Journal:  Br J Clin Pharmacol       Date:  1995-05       Impact factor: 4.335

6.  Automated flow quantification in valvular heart disease based on backscattered Doppler power analysis: implementation on matrix-array ultrasound imaging systems.

Authors:  Thomas Buck; Shawn M Hwang; Björn Plicht; Ronald A Mucci; Peter Hunold; Raimund Erbel; Robert A Levine
Journal:  Int J Cardiovasc Imaging       Date:  2008-03-02       Impact factor: 2.357

Review 7.  Cardiac output monitoring by echocardiography: should we pass on Swan-Ganz catheters?

Authors:  A C Perrino
Journal:  Yale J Biol Med       Date:  1993 Sep-Oct
  7 in total